A new scheme for real-time high-contrast imaging in lung cancer radiotherapy

A proof-of-concept study

Hao Yan, Zhen Tian, Yiping Shao, Steve B. Jiang, Xun Jia

Research output: Contribution to journalArticle

8 Citations (Scopus)

Abstract

Visualization of anatomy in real time is of critical importance for motion management in lung cancer radiotherapy. To achieve real-time, and high-contrast in-treatment imaging, we propose a novel scheme based on the measurement of Compton scatter photons. In our method, a slit x-ray beam along the superior-inferior direction is directed to the patient, (intersecting the lung region at a 2D plane) containing most of the tumor motion trajectory. X-ray photons are scattered off this plane primarily due to the Compton interaction. An imager with a pinhole or a slat collimator is placed at one side of the plane to capture the scattered photons. The resulting image, after correcting for incoming fluence inhomogeneity, x-ray attenuation, scatter angle variation, and outgoing beam geometry, represents the linear attenuation coefficient of Compton scattering. This allows the visualization of the anatomy on this plane. We performed Monte Carlo simulation studies both on a phantom and a patient for proof-of-principle purposes. In the phantom case, a small tumor-like structure could be clearly visualized. The contrast-resolution calculated using tumor/lung as foreground/background for kV fluoroscopy, cone beam computed tomography (CBCT), and scattering image were 0.037, 0.70, and 0.54, respectively. In the patient case, tumor motion could be clearly observed in the scatter images. Imaging dose to the voxels directly exposed by the slit beam was ∼0.4 times of that under a single CBCT projection. These studies demonstrated the potential feasibility of the proposed imaging scheme to capture the instantaneous anatomy of a patient on a 2D plane with a high image contrast. Clear visualization of the tumor motion may facilitate marker-less tumor tracking.

Original languageEnglish (US)
Pages (from-to)2372-2388
Number of pages17
JournalPhysics in Medicine and Biology
Volume61
Issue number6
DOIs
StatePublished - Mar 4 2016

Fingerprint

Lung Neoplasms
Radiotherapy
Photons
Anatomy
Cone-Beam Computed Tomography
X-Rays
Neoplasms
Lung
Fluoroscopy
Tumor Biomarkers
Therapeutics

Keywords

  • radiotherapy
  • scatter imaging
  • x-ray

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

A new scheme for real-time high-contrast imaging in lung cancer radiotherapy : A proof-of-concept study. / Yan, Hao; Tian, Zhen; Shao, Yiping; Jiang, Steve B.; Jia, Xun.

In: Physics in Medicine and Biology, Vol. 61, No. 6, 04.03.2016, p. 2372-2388.

Research output: Contribution to journalArticle

@article{9e820f99f6a7410c98be214ff41ecbd3,
title = "A new scheme for real-time high-contrast imaging in lung cancer radiotherapy: A proof-of-concept study",
abstract = "Visualization of anatomy in real time is of critical importance for motion management in lung cancer radiotherapy. To achieve real-time, and high-contrast in-treatment imaging, we propose a novel scheme based on the measurement of Compton scatter photons. In our method, a slit x-ray beam along the superior-inferior direction is directed to the patient, (intersecting the lung region at a 2D plane) containing most of the tumor motion trajectory. X-ray photons are scattered off this plane primarily due to the Compton interaction. An imager with a pinhole or a slat collimator is placed at one side of the plane to capture the scattered photons. The resulting image, after correcting for incoming fluence inhomogeneity, x-ray attenuation, scatter angle variation, and outgoing beam geometry, represents the linear attenuation coefficient of Compton scattering. This allows the visualization of the anatomy on this plane. We performed Monte Carlo simulation studies both on a phantom and a patient for proof-of-principle purposes. In the phantom case, a small tumor-like structure could be clearly visualized. The contrast-resolution calculated using tumor/lung as foreground/background for kV fluoroscopy, cone beam computed tomography (CBCT), and scattering image were 0.037, 0.70, and 0.54, respectively. In the patient case, tumor motion could be clearly observed in the scatter images. Imaging dose to the voxels directly exposed by the slit beam was ∼0.4 times of that under a single CBCT projection. These studies demonstrated the potential feasibility of the proposed imaging scheme to capture the instantaneous anatomy of a patient on a 2D plane with a high image contrast. Clear visualization of the tumor motion may facilitate marker-less tumor tracking.",
keywords = "radiotherapy, scatter imaging, x-ray",
author = "Hao Yan and Zhen Tian and Yiping Shao and Jiang, {Steve B.} and Xun Jia",
year = "2016",
month = "3",
day = "4",
doi = "10.1088/0031-9155/61/6/2372",
language = "English (US)",
volume = "61",
pages = "2372--2388",
journal = "Physics in Medicine and Biology",
issn = "0031-9155",
publisher = "IOP Publishing Ltd.",
number = "6",

}

TY - JOUR

T1 - A new scheme for real-time high-contrast imaging in lung cancer radiotherapy

T2 - A proof-of-concept study

AU - Yan, Hao

AU - Tian, Zhen

AU - Shao, Yiping

AU - Jiang, Steve B.

AU - Jia, Xun

PY - 2016/3/4

Y1 - 2016/3/4

N2 - Visualization of anatomy in real time is of critical importance for motion management in lung cancer radiotherapy. To achieve real-time, and high-contrast in-treatment imaging, we propose a novel scheme based on the measurement of Compton scatter photons. In our method, a slit x-ray beam along the superior-inferior direction is directed to the patient, (intersecting the lung region at a 2D plane) containing most of the tumor motion trajectory. X-ray photons are scattered off this plane primarily due to the Compton interaction. An imager with a pinhole or a slat collimator is placed at one side of the plane to capture the scattered photons. The resulting image, after correcting for incoming fluence inhomogeneity, x-ray attenuation, scatter angle variation, and outgoing beam geometry, represents the linear attenuation coefficient of Compton scattering. This allows the visualization of the anatomy on this plane. We performed Monte Carlo simulation studies both on a phantom and a patient for proof-of-principle purposes. In the phantom case, a small tumor-like structure could be clearly visualized. The contrast-resolution calculated using tumor/lung as foreground/background for kV fluoroscopy, cone beam computed tomography (CBCT), and scattering image were 0.037, 0.70, and 0.54, respectively. In the patient case, tumor motion could be clearly observed in the scatter images. Imaging dose to the voxels directly exposed by the slit beam was ∼0.4 times of that under a single CBCT projection. These studies demonstrated the potential feasibility of the proposed imaging scheme to capture the instantaneous anatomy of a patient on a 2D plane with a high image contrast. Clear visualization of the tumor motion may facilitate marker-less tumor tracking.

AB - Visualization of anatomy in real time is of critical importance for motion management in lung cancer radiotherapy. To achieve real-time, and high-contrast in-treatment imaging, we propose a novel scheme based on the measurement of Compton scatter photons. In our method, a slit x-ray beam along the superior-inferior direction is directed to the patient, (intersecting the lung region at a 2D plane) containing most of the tumor motion trajectory. X-ray photons are scattered off this plane primarily due to the Compton interaction. An imager with a pinhole or a slat collimator is placed at one side of the plane to capture the scattered photons. The resulting image, after correcting for incoming fluence inhomogeneity, x-ray attenuation, scatter angle variation, and outgoing beam geometry, represents the linear attenuation coefficient of Compton scattering. This allows the visualization of the anatomy on this plane. We performed Monte Carlo simulation studies both on a phantom and a patient for proof-of-principle purposes. In the phantom case, a small tumor-like structure could be clearly visualized. The contrast-resolution calculated using tumor/lung as foreground/background for kV fluoroscopy, cone beam computed tomography (CBCT), and scattering image were 0.037, 0.70, and 0.54, respectively. In the patient case, tumor motion could be clearly observed in the scatter images. Imaging dose to the voxels directly exposed by the slit beam was ∼0.4 times of that under a single CBCT projection. These studies demonstrated the potential feasibility of the proposed imaging scheme to capture the instantaneous anatomy of a patient on a 2D plane with a high image contrast. Clear visualization of the tumor motion may facilitate marker-less tumor tracking.

KW - radiotherapy

KW - scatter imaging

KW - x-ray

UR - http://www.scopus.com/inward/record.url?scp=84960976476&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84960976476&partnerID=8YFLogxK

U2 - 10.1088/0031-9155/61/6/2372

DO - 10.1088/0031-9155/61/6/2372

M3 - Article

VL - 61

SP - 2372

EP - 2388

JO - Physics in Medicine and Biology

JF - Physics in Medicine and Biology

SN - 0031-9155

IS - 6

ER -