Abstract
Objective: The objective was to report on the possibility of allergic reaction to the components of a spinal cord stimulator. Design: We describe a severe allergic reaction after the insertion of a spinal cord stimulator in a patient with complex regional pain syndrome type 1. Setting: The patient was being followed in an office-based pain management practice. Patient: The patient is a 41-year-old woman with complex regional pain syndrome type 1, posttrauma. Intervention: Insertion of a cervical and lumbar spinal cord stimulator. Outcome Measures: The outcome measures were a numerical scale of pain intensity and the ability to perform the activities of daily living. Results: Adequate pain control complicated by allergic reaction. Conclusions: There exists a possibility that a patient may experience an allergic reaction to spinal cord stimulator components. Recognition of such contact sensitivity is important for physicians implanting such devices. Patients may be misdiagnosed as having infections, which can delay appropriate management; definitive diagnosis can be confirmed with a patch test. Treatment consists of removal of such devices.
Original language | English (US) |
---|---|
Pages (from-to) | 178-180 |
Number of pages | 3 |
Journal | Clinical Journal of Pain |
Volume | 16 |
Issue number | 2 |
DOIs | |
State | Published - Jul 3 2000 |
Keywords
- Biomaterials
- Contact sensitivity
- Dorsal column spinal cord stimulator
ASJC Scopus subject areas
- Clinical Neurology
- Anesthesiology and Pain Medicine