Abstract
A single HPIC method was developed and validated for the analysis of both [11C]Choline and [13N]Ammonia with the same setup. The HPIC system suitability tests were performed and [11C]Choline and [13N]Ammonia were used to verify their performance on this HPIC method. The HPIC setup and method provides qualitative and quantitative analysis information of [11C]Choline and [13N]Ammonia. The data suggested this HPIC method is validated for determining radiochemical/chemical purity and radiochemical identity of [11C]Choline and [13N]Ammonia products in CGMP compliant manufacturing process.
Original language | English (US) |
---|---|
Article number | 109560 |
Journal | Applied Radiation and Isotopes |
Volume | 168 |
DOIs | |
State | Published - Feb 2021 |
Externally published | Yes |
Keywords
- HPIC
- Method validation
- PET
- Quality control
- [C]Choline
- [N]Ammonia
ASJC Scopus subject areas
- Radiation