Benchmarking techniques for stereotactic body radiotherapy for early-stage glottic laryngeal cancer: LINAC-based non-coplanar VMAT vs. Cyberknife planning

Research output: Contribution to journalArticle

Abstract

Introduction: Stereotactic body radiation therapy (SBRT) was found effective in treating laryngeal cancer with only five treatment fractions by a recent clinical trial (NCT01984502, ClinicalTrials.gov). Nevertheless, this trial used the Cyberknife system, which is not widely accessible enough to benefit all patients affected by laryngeal cancer. Our study investigates the feasibility of larynx SBRT treatment planning on a conventional gantry-based LINAC and compares its plan quality with that from the Cyberknife. Materials & methods: Ten larynx SBRT cases were originally treated by Cyberknife using fixed cones in our institution, with plans created and optimized using the Monte-Carlo algorithm in the MultiPlan treatment planning system. These cases were retrospectively planned in the Eclipse planning system for a LINAC with the same prescription dose. We used volumetric modulated arc therapy (VMAT) for larynx SBRT planning in Eclipse and incorporated non-coplanar arcs to approach the Cyberknife's large solid angle delivery space. We used both anisotropic analytical algorithm (AAA) and Acuros XB (AXB) algorithm for dose calculation and compared their accuracy by measurements on an in-house larynx phantom. We compared the LINAC VMAT plans (VMAT-AAA and VMAT-AXB) with the original Cyberknife plans using dosimetric endpoints such as the conformity index, gradient indices (R50, R20), OAR maximum/mean doses, and the monitor units. Results: Phantom measurement showed that both the AAA and the AXB algorithms provided adequate dose calculation accuracy (94.7% gamma pass rate on 2%/2 mm criteria for AAA vs. 97.3% for AXB), though AXB provided better accuracy in the air cavity. The LINAC-based VMAT plans achieved similar dosimetric endpoints as the Cyberknife planning, and all plans met the larynx SBRT dosimetric constraints. Cyberknife plans achieved an average conformity index of 1.13, compared to 1.20 of VMAT-AXB and 1.19 of VMAT-AAA. The VMAT plans spared the thyroid gland better with average Dmean of 2.4 Gy (VMAT-AXB) and 2.7 Gy (VMAT-AAA), as compared to 4.3 Gy for Cyberknife plans. The VMAT-AAA plans had a slightly lower contralateral arytenoid Dmax (average: 15.2 Gy) than Cyberknife plans (average: 17.9 Gy) with statistical significance, while the contralateral arytenoid Dmax was similar between VMAT-AXB and Cyberknife plans with no statistically significant difference. Cyberknife plans offered slightly better R50 (average: 5.0) than VMAT-AXB (5.9) and VMAT-AAA (5.7) plans. The VMAT plans substantially reduced the plan MUs to less than 1/3 of the Cyberknife plans, and the differences were statistically significant. The other metrics were similar between VMAT and Cyberknife plans with no statistically significant differences. Conclusions: Gantry-based LINACs can achieve similar plan quality to Cyberknife systems. Treatment outcome with both methods remains to be investigated.

Original languageEnglish (US)
Article number193
JournalRadiation Oncology
Volume14
Issue number1
DOIs
StatePublished - Nov 4 2019

Fingerprint

Tongue Neoplasms
Benchmarking
Intensity-Modulated Radiotherapy
Laryngeal Neoplasms
Radiosurgery
Larynx
Radiotherapy

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging

Cite this

@article{909acd5bca0449649693a52aa1ca2102,
title = "Benchmarking techniques for stereotactic body radiotherapy for early-stage glottic laryngeal cancer: LINAC-based non-coplanar VMAT vs. Cyberknife planning",
abstract = "Introduction: Stereotactic body radiation therapy (SBRT) was found effective in treating laryngeal cancer with only five treatment fractions by a recent clinical trial (NCT01984502, ClinicalTrials.gov). Nevertheless, this trial used the Cyberknife system, which is not widely accessible enough to benefit all patients affected by laryngeal cancer. Our study investigates the feasibility of larynx SBRT treatment planning on a conventional gantry-based LINAC and compares its plan quality with that from the Cyberknife. Materials & methods: Ten larynx SBRT cases were originally treated by Cyberknife using fixed cones in our institution, with plans created and optimized using the Monte-Carlo algorithm in the MultiPlan treatment planning system. These cases were retrospectively planned in the Eclipse planning system for a LINAC with the same prescription dose. We used volumetric modulated arc therapy (VMAT) for larynx SBRT planning in Eclipse and incorporated non-coplanar arcs to approach the Cyberknife's large solid angle delivery space. We used both anisotropic analytical algorithm (AAA) and Acuros XB (AXB) algorithm for dose calculation and compared their accuracy by measurements on an in-house larynx phantom. We compared the LINAC VMAT plans (VMAT-AAA and VMAT-AXB) with the original Cyberknife plans using dosimetric endpoints such as the conformity index, gradient indices (R50, R20), OAR maximum/mean doses, and the monitor units. Results: Phantom measurement showed that both the AAA and the AXB algorithms provided adequate dose calculation accuracy (94.7{\%} gamma pass rate on 2{\%}/2 mm criteria for AAA vs. 97.3{\%} for AXB), though AXB provided better accuracy in the air cavity. The LINAC-based VMAT plans achieved similar dosimetric endpoints as the Cyberknife planning, and all plans met the larynx SBRT dosimetric constraints. Cyberknife plans achieved an average conformity index of 1.13, compared to 1.20 of VMAT-AXB and 1.19 of VMAT-AAA. The VMAT plans spared the thyroid gland better with average Dmean of 2.4 Gy (VMAT-AXB) and 2.7 Gy (VMAT-AAA), as compared to 4.3 Gy for Cyberknife plans. The VMAT-AAA plans had a slightly lower contralateral arytenoid Dmax (average: 15.2 Gy) than Cyberknife plans (average: 17.9 Gy) with statistical significance, while the contralateral arytenoid Dmax was similar between VMAT-AXB and Cyberknife plans with no statistically significant difference. Cyberknife plans offered slightly better R50 (average: 5.0) than VMAT-AXB (5.9) and VMAT-AAA (5.7) plans. The VMAT plans substantially reduced the plan MUs to less than 1/3 of the Cyberknife plans, and the differences were statistically significant. The other metrics were similar between VMAT and Cyberknife plans with no statistically significant differences. Conclusions: Gantry-based LINACs can achieve similar plan quality to Cyberknife systems. Treatment outcome with both methods remains to be investigated.",
author = "You Zhang and Tsuicheng Chiu and Jeffrey Dubas and Zhen Tian and Pam Lee and Xuejun Gu and Yulong Yan and David Sher and Robert Timmerman and Bo Zhao",
year = "2019",
month = "11",
day = "4",
doi = "10.1186/s13014-019-1404-z",
language = "English (US)",
volume = "14",
journal = "Radiation Oncology",
issn = "1748-717X",
publisher = "BioMed Central",
number = "1",

}

TY - JOUR

T1 - Benchmarking techniques for stereotactic body radiotherapy for early-stage glottic laryngeal cancer

T2 - LINAC-based non-coplanar VMAT vs. Cyberknife planning

AU - Zhang, You

AU - Chiu, Tsuicheng

AU - Dubas, Jeffrey

AU - Tian, Zhen

AU - Lee, Pam

AU - Gu, Xuejun

AU - Yan, Yulong

AU - Sher, David

AU - Timmerman, Robert

AU - Zhao, Bo

PY - 2019/11/4

Y1 - 2019/11/4

N2 - Introduction: Stereotactic body radiation therapy (SBRT) was found effective in treating laryngeal cancer with only five treatment fractions by a recent clinical trial (NCT01984502, ClinicalTrials.gov). Nevertheless, this trial used the Cyberknife system, which is not widely accessible enough to benefit all patients affected by laryngeal cancer. Our study investigates the feasibility of larynx SBRT treatment planning on a conventional gantry-based LINAC and compares its plan quality with that from the Cyberknife. Materials & methods: Ten larynx SBRT cases were originally treated by Cyberknife using fixed cones in our institution, with plans created and optimized using the Monte-Carlo algorithm in the MultiPlan treatment planning system. These cases were retrospectively planned in the Eclipse planning system for a LINAC with the same prescription dose. We used volumetric modulated arc therapy (VMAT) for larynx SBRT planning in Eclipse and incorporated non-coplanar arcs to approach the Cyberknife's large solid angle delivery space. We used both anisotropic analytical algorithm (AAA) and Acuros XB (AXB) algorithm for dose calculation and compared their accuracy by measurements on an in-house larynx phantom. We compared the LINAC VMAT plans (VMAT-AAA and VMAT-AXB) with the original Cyberknife plans using dosimetric endpoints such as the conformity index, gradient indices (R50, R20), OAR maximum/mean doses, and the monitor units. Results: Phantom measurement showed that both the AAA and the AXB algorithms provided adequate dose calculation accuracy (94.7% gamma pass rate on 2%/2 mm criteria for AAA vs. 97.3% for AXB), though AXB provided better accuracy in the air cavity. The LINAC-based VMAT plans achieved similar dosimetric endpoints as the Cyberknife planning, and all plans met the larynx SBRT dosimetric constraints. Cyberknife plans achieved an average conformity index of 1.13, compared to 1.20 of VMAT-AXB and 1.19 of VMAT-AAA. The VMAT plans spared the thyroid gland better with average Dmean of 2.4 Gy (VMAT-AXB) and 2.7 Gy (VMAT-AAA), as compared to 4.3 Gy for Cyberknife plans. The VMAT-AAA plans had a slightly lower contralateral arytenoid Dmax (average: 15.2 Gy) than Cyberknife plans (average: 17.9 Gy) with statistical significance, while the contralateral arytenoid Dmax was similar between VMAT-AXB and Cyberknife plans with no statistically significant difference. Cyberknife plans offered slightly better R50 (average: 5.0) than VMAT-AXB (5.9) and VMAT-AAA (5.7) plans. The VMAT plans substantially reduced the plan MUs to less than 1/3 of the Cyberknife plans, and the differences were statistically significant. The other metrics were similar between VMAT and Cyberknife plans with no statistically significant differences. Conclusions: Gantry-based LINACs can achieve similar plan quality to Cyberknife systems. Treatment outcome with both methods remains to be investigated.

AB - Introduction: Stereotactic body radiation therapy (SBRT) was found effective in treating laryngeal cancer with only five treatment fractions by a recent clinical trial (NCT01984502, ClinicalTrials.gov). Nevertheless, this trial used the Cyberknife system, which is not widely accessible enough to benefit all patients affected by laryngeal cancer. Our study investigates the feasibility of larynx SBRT treatment planning on a conventional gantry-based LINAC and compares its plan quality with that from the Cyberknife. Materials & methods: Ten larynx SBRT cases were originally treated by Cyberknife using fixed cones in our institution, with plans created and optimized using the Monte-Carlo algorithm in the MultiPlan treatment planning system. These cases were retrospectively planned in the Eclipse planning system for a LINAC with the same prescription dose. We used volumetric modulated arc therapy (VMAT) for larynx SBRT planning in Eclipse and incorporated non-coplanar arcs to approach the Cyberknife's large solid angle delivery space. We used both anisotropic analytical algorithm (AAA) and Acuros XB (AXB) algorithm for dose calculation and compared their accuracy by measurements on an in-house larynx phantom. We compared the LINAC VMAT plans (VMAT-AAA and VMAT-AXB) with the original Cyberknife plans using dosimetric endpoints such as the conformity index, gradient indices (R50, R20), OAR maximum/mean doses, and the monitor units. Results: Phantom measurement showed that both the AAA and the AXB algorithms provided adequate dose calculation accuracy (94.7% gamma pass rate on 2%/2 mm criteria for AAA vs. 97.3% for AXB), though AXB provided better accuracy in the air cavity. The LINAC-based VMAT plans achieved similar dosimetric endpoints as the Cyberknife planning, and all plans met the larynx SBRT dosimetric constraints. Cyberknife plans achieved an average conformity index of 1.13, compared to 1.20 of VMAT-AXB and 1.19 of VMAT-AAA. The VMAT plans spared the thyroid gland better with average Dmean of 2.4 Gy (VMAT-AXB) and 2.7 Gy (VMAT-AAA), as compared to 4.3 Gy for Cyberknife plans. The VMAT-AAA plans had a slightly lower contralateral arytenoid Dmax (average: 15.2 Gy) than Cyberknife plans (average: 17.9 Gy) with statistical significance, while the contralateral arytenoid Dmax was similar between VMAT-AXB and Cyberknife plans with no statistically significant difference. Cyberknife plans offered slightly better R50 (average: 5.0) than VMAT-AXB (5.9) and VMAT-AAA (5.7) plans. The VMAT plans substantially reduced the plan MUs to less than 1/3 of the Cyberknife plans, and the differences were statistically significant. The other metrics were similar between VMAT and Cyberknife plans with no statistically significant differences. Conclusions: Gantry-based LINACs can achieve similar plan quality to Cyberknife systems. Treatment outcome with both methods remains to be investigated.

UR - http://www.scopus.com/inward/record.url?scp=85074550970&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85074550970&partnerID=8YFLogxK

U2 - 10.1186/s13014-019-1404-z

DO - 10.1186/s13014-019-1404-z

M3 - Article

C2 - 31684993

AN - SCOPUS:85074550970

VL - 14

JO - Radiation Oncology

JF - Radiation Oncology

SN - 1748-717X

IS - 1

M1 - 193

ER -