Characterization of electron beams for modulated electron beam radiotherapy

M. C. Lee, A. Kapur, S. B. Jiang, C. M. Ma

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Abstract

A set of clinical electron beams have been characterized in anticipation of their use in modulated electron radiotherapy (MERT). Using the EGS4/BEAM and DOSXYZ simulation codes, a Monte Carlo model of a Varian Clinac 2100C was generated for electron beams of nominal energies of 6, 9, 12, 16, and 20 MeV. Dose calculations using the Monte Carlo generated phase space data demonstrate profile and depth-dose agreement with measured data to within 2% of the maximum dose values. Simulations were then performed using a proposed electron multileaf collimator to examine beam characteristics relevant to MERT. In particular, abutting fields have been constructed and the presence of hot and cold spots due to penumbral mismatches is shown to be significant. The effect of field setup errors is also considered and it is shown that small setup uncertainties of 3mm may lead to significant hot and cold spots of up to 25%. Furthermore, electron beamlets were shown to differ in penumbras and symmetry based on where they were located in the field. Intensity modulated electron fields were then constructed, demonstrating the ability to create fields with tilted dose profiles using a sequential abutting field method and a dynamic delivery method.

Original languageEnglish (US)
Title of host publicationAnnual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings
EditorsJ.D Enderle
Pages1497-1500
Number of pages4
Volume2
StatePublished - 2000
Event22nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society - Chicago, IL, United States
Duration: Jul 23 2000Jul 28 2000

Other

Other22nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society
CountryUnited States
CityChicago, IL
Period7/23/007/28/00

Fingerprint

Radiotherapy
Electron beams
Electrons

Keywords

  • EGS4
  • IMRT
  • Modulated electron beams (MERT)
  • Monte Carlo

ASJC Scopus subject areas

  • Bioengineering

Cite this

Lee, M. C., Kapur, A., Jiang, S. B., & Ma, C. M. (2000). Characterization of electron beams for modulated electron beam radiotherapy. In J. D. Enderle (Ed.), Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings (Vol. 2, pp. 1497-1500)

Characterization of electron beams for modulated electron beam radiotherapy. / Lee, M. C.; Kapur, A.; Jiang, S. B.; Ma, C. M.

Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings. ed. / J.D Enderle. Vol. 2 2000. p. 1497-1500.

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Lee, MC, Kapur, A, Jiang, SB & Ma, CM 2000, Characterization of electron beams for modulated electron beam radiotherapy. in JD Enderle (ed.), Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings. vol. 2, pp. 1497-1500, 22nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Chicago, IL, United States, 7/23/00.
Lee MC, Kapur A, Jiang SB, Ma CM. Characterization of electron beams for modulated electron beam radiotherapy. In Enderle JD, editor, Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings. Vol. 2. 2000. p. 1497-1500
Lee, M. C. ; Kapur, A. ; Jiang, S. B. ; Ma, C. M. / Characterization of electron beams for modulated electron beam radiotherapy. Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings. editor / J.D Enderle. Vol. 2 2000. pp. 1497-1500
@inproceedings{97f32ff3f84143639c043e294c3a0838,
title = "Characterization of electron beams for modulated electron beam radiotherapy",
abstract = "A set of clinical electron beams have been characterized in anticipation of their use in modulated electron radiotherapy (MERT). Using the EGS4/BEAM and DOSXYZ simulation codes, a Monte Carlo model of a Varian Clinac 2100C was generated for electron beams of nominal energies of 6, 9, 12, 16, and 20 MeV. Dose calculations using the Monte Carlo generated phase space data demonstrate profile and depth-dose agreement with measured data to within 2{\%} of the maximum dose values. Simulations were then performed using a proposed electron multileaf collimator to examine beam characteristics relevant to MERT. In particular, abutting fields have been constructed and the presence of hot and cold spots due to penumbral mismatches is shown to be significant. The effect of field setup errors is also considered and it is shown that small setup uncertainties of 3mm may lead to significant hot and cold spots of up to 25{\%}. Furthermore, electron beamlets were shown to differ in penumbras and symmetry based on where they were located in the field. Intensity modulated electron fields were then constructed, demonstrating the ability to create fields with tilted dose profiles using a sequential abutting field method and a dynamic delivery method.",
keywords = "EGS4, IMRT, Modulated electron beams (MERT), Monte Carlo",
author = "Lee, {M. C.} and A. Kapur and Jiang, {S. B.} and Ma, {C. M.}",
year = "2000",
language = "English (US)",
volume = "2",
pages = "1497--1500",
editor = "J.D Enderle",
booktitle = "Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings",

}

TY - GEN

T1 - Characterization of electron beams for modulated electron beam radiotherapy

AU - Lee, M. C.

AU - Kapur, A.

AU - Jiang, S. B.

AU - Ma, C. M.

PY - 2000

Y1 - 2000

N2 - A set of clinical electron beams have been characterized in anticipation of their use in modulated electron radiotherapy (MERT). Using the EGS4/BEAM and DOSXYZ simulation codes, a Monte Carlo model of a Varian Clinac 2100C was generated for electron beams of nominal energies of 6, 9, 12, 16, and 20 MeV. Dose calculations using the Monte Carlo generated phase space data demonstrate profile and depth-dose agreement with measured data to within 2% of the maximum dose values. Simulations were then performed using a proposed electron multileaf collimator to examine beam characteristics relevant to MERT. In particular, abutting fields have been constructed and the presence of hot and cold spots due to penumbral mismatches is shown to be significant. The effect of field setup errors is also considered and it is shown that small setup uncertainties of 3mm may lead to significant hot and cold spots of up to 25%. Furthermore, electron beamlets were shown to differ in penumbras and symmetry based on where they were located in the field. Intensity modulated electron fields were then constructed, demonstrating the ability to create fields with tilted dose profiles using a sequential abutting field method and a dynamic delivery method.

AB - A set of clinical electron beams have been characterized in anticipation of their use in modulated electron radiotherapy (MERT). Using the EGS4/BEAM and DOSXYZ simulation codes, a Monte Carlo model of a Varian Clinac 2100C was generated for electron beams of nominal energies of 6, 9, 12, 16, and 20 MeV. Dose calculations using the Monte Carlo generated phase space data demonstrate profile and depth-dose agreement with measured data to within 2% of the maximum dose values. Simulations were then performed using a proposed electron multileaf collimator to examine beam characteristics relevant to MERT. In particular, abutting fields have been constructed and the presence of hot and cold spots due to penumbral mismatches is shown to be significant. The effect of field setup errors is also considered and it is shown that small setup uncertainties of 3mm may lead to significant hot and cold spots of up to 25%. Furthermore, electron beamlets were shown to differ in penumbras and symmetry based on where they were located in the field. Intensity modulated electron fields were then constructed, demonstrating the ability to create fields with tilted dose profiles using a sequential abutting field method and a dynamic delivery method.

KW - EGS4

KW - IMRT

KW - Modulated electron beams (MERT)

KW - Monte Carlo

UR - http://www.scopus.com/inward/record.url?scp=0034445680&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0034445680&partnerID=8YFLogxK

M3 - Conference contribution

VL - 2

SP - 1497

EP - 1500

BT - Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings

A2 - Enderle, J.D

ER -