Computer-aided analysis of star shot films for high-accuracy radiation therapy treatment units

Tom Depuydt, Rudi Penne, Dirk Verellen, Jan Hrbacek, Stephanie Lang, Katrien Leysen, Iwein Vandevondel, Kenneth Poels, Truus Reynders, Thierry Gevaert, Michael Duchateau, Koen Tournel, Marlies Boussaer, Dorian Cosentino, Cristina Garibaldi, Timothy Solberg, Mark De Ridder

Research output: Contribution to journalArticlepeer-review

37 Scopus citations

Abstract

As mechanical stability of radiation therapy treatment devices has gone beyond sub-millimeter levels, there is a rising demand for simple yet highly accurate measurement techniques to support the routine quality control of these devices. A combination of using high-resolution radiosensitive film and computer-aided analysis could provide an answer. One generally known technique is the acquisition of star shot films to determine the mechanical stability of rotations of gantries and the therapeutic beam. With computer-aided analysis, mechanical performance can be quantified as a radiation isocenter radius size. In this work, computer-aided analysis of star shot film is further refined by applying an analytical solution for the smallest intersecting circle problem, in contrast to the gradient optimization approaches used until today. An algorithm is presented and subjected to a performance test using two different types of radiosensitive film, the Kodak EDR2 radiographic film and the ISP EBT2 radiochromic film. Artificial star shots with a priori known radiation isocenter size are used to determine the systematic errors introduced by the digitization of the film and the computer analysis. The estimated uncertainty on the isocenter size measurement with the presented technique was 0.04 mm (2σ) and 0.06 mm (2σ) for radiographic and radiochromic films, respectively. As an application of the technique, a study was conducted to compare the mechanical stability of O-ring gantry systems with C-arm-based gantries. In total ten systems of five different institutions were included in this study and star shots were acquired for gantry, collimator, ring, couch rotations and gantry wobble. It was not possible to draw general conclusions about differences in mechanical performance between O-ring and C-arm gantry systems, mainly due to differences in the beamMLC alignment procedure accuracy. Nevertheless, the best performing O-ring system in this study, a BrainLab/MHI Vero system, and the best performing C-arm system, a Varian Truebeam system, showed comparable mechanical performance: gantry isocenter radius of 0.12 and 0.09mm, respectively, ring/couch rotation of below 0.10 mm for both systems and a wobble of 0.06 and 0.18 mm, respectively. The methodology described in this work can be used to monitor mechanical performance constancy of high-accuracy treatment devices, with means available in a clinical radiation therapy environment.

Original languageEnglish (US)
Pages (from-to)2997-3011
Number of pages15
JournalPhysics in medicine and biology
Volume57
Issue number10
DOIs
StatePublished - May 21 2012

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Computer-aided analysis of star shot films for high-accuracy radiation therapy treatment units'. Together they form a unique fingerprint.

Cite this