TY - JOUR
T1 - Design and development of 1 mm resolution PET detectors with position-sensitive PMTs
AU - Shao, Yiping
AU - Cherry, Simon R.
AU - Chatziioannou, Arion F.
N1 - Funding Information:
We would like to thank K. Meadors and R.W. Silverman for technical support. This work was supported by grants from the National Cancer Institute (CA69370 and CA74036) and by Department of Energy contract DE-FC03-87-ER60615.
PY - 2002/1/21
Y1 - 2002/1/21
N2 - We report our investigation of a positron emission tomography (PET) detector with 1m spatial resolution. The prototype detector consists of a 9 × 9 array of 1 × 1 × 10mm3 lutetium oxyorthosilicate (LSO) scintillator crystals coupled to Hamamatsu R5900-M64 or R5900-C12 position sensitive PMT by either optical fibers or an optical fiber bundle. With a 511eV gamma source, the intrinsic spatial resolution of this detector was measured to be 0.92mm. All crystals were well resolved in the flood source histogram. The measured energy and coincidence timing resolutions were around 26% and 4ns, respectively, demonstrating that sufficient light can be extracted from these small crystals for PET applications.
AB - We report our investigation of a positron emission tomography (PET) detector with 1m spatial resolution. The prototype detector consists of a 9 × 9 array of 1 × 1 × 10mm3 lutetium oxyorthosilicate (LSO) scintillator crystals coupled to Hamamatsu R5900-M64 or R5900-C12 position sensitive PMT by either optical fibers or an optical fiber bundle. With a 511eV gamma source, the intrinsic spatial resolution of this detector was measured to be 0.92mm. All crystals were well resolved in the flood source histogram. The measured energy and coincidence timing resolutions were around 26% and 4ns, respectively, demonstrating that sufficient light can be extracted from these small crystals for PET applications.
KW - Gamma detector
KW - Multi-channel photomultiplier (MC-PMT)
KW - Optical fiber bundle
KW - Position-sensitive photomultiplier (PS-PMT)
KW - Positron emission tomography (PET)
KW - Scintillator
UR - http://www.scopus.com/inward/record.url?scp=0037147906&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0037147906&partnerID=8YFLogxK
U2 - 10.1016/S0168-9002(01)01785-5
DO - 10.1016/S0168-9002(01)01785-5
M3 - Conference article
AN - SCOPUS:0037147906
SN - 0168-9002
VL - 477
SP - 486
EP - 490
JO - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
JF - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
IS - 1-3
T2 - 5th International Conference on Position-Sensitive Detectors PSD-5 1999
Y2 - 13 September 1999 through 17 September 1999
ER -