@article{4ae4d8cb2b774313a3f03826518778c3,
title = "Effect of biphasic endocardial countershock on pacing thresholds in humans",
abstract = "The use of non-thoracotomy endocardial implantable defibrillation with pacing capabilities has increased substantially over the past 2 years. This report demonstrates that the pacing threshold increases in some patients after endocardial defibrillation, and substantiates the practice of using maximal pacing output after endocardial defibrillation.",
author = "Kessler, {D. J.} and Canby, {R. C.} and Horton, {R. P.} and Joglar, {Jose A} and Jessen, {Michael E} and Page, {R. L.}",
note = "Funding Information: From the University ofTexas Southwestern Medical Center, Division of Cardiology, CS 7.102, 5323 Harry Hines Boulevard, Dallas, Tex6s 75235-9047. This study was supported in rt by Training in Cardiovascular Research Grant 5T32HL07360 r rom the National Institutes of Health, Bethesda, Mayland. Manuscript received August 24, 1995; revised manuscript received October 18, 1995, and accepted October 20.",
year = "1996",
doi = "10.1016/S0002-9149(97)89350-2",
language = "English (US)",
volume = "77",
pages = "527--528",
journal = "American Journal of Cardiology",
issn = "0002-9149",
publisher = "Elsevier Inc.",
number = "7",
}