Evaluation of the Elekta Agility MLC performance using high-resolution log files

Christopher N. Kabat, Dewayne L. Defoor, Pamela Myers, Neil Kirby, Karl Rasmussen, Daniel L. Saenz, Panayiotis Mavroidis, Niko Papanikolaou, Sotirios Stathakis

Research output: Contribution to journalArticlepeer-review

18 Scopus citations

Abstract

Purpose: With the advent of volumetric modulated arc therapy (VMAT) and intensity-modulated radiation therapy (IMRT) treatment techniques, the requirement for more elaborate approaches in reviewing linac components’ integrity has become even more stringent. A possible solution to this challenge is to employ the usage of log files generated during treatment. The log files generated by the new generation of Elekta linacs record events at a higher frequency (25 Hz) than their predecessors, which allows for retrospective analysis and identification of subtle changes and provides another means of quality assurance. The ability to track machine components based on log files for each treatment can allow for constant monitoring of fraction consistency in addition to machine reliability. Using Elekta Agility log files, a set of tests were developed to evaluate the reliability and robustness of the multileaf collimators (MLCs). Methods: To evaluate Elekta log file utilization for linac MLC QA effectiveness, five MLC test patterns were constructed to review the effects of leaf velocity and acceleration on positional accuracy, including gravitational effects for the Elekta MLC system. Each test was run five times in a particular setting to obtain reproducibility data and statistical averages. This study was performed on two identical Versa HD machines, each delivering a full set of test plans with all possible variations. Plans were delivered using Elekta's iCOMcat software and recorded log files were extracted. Log files were reformatted for readability and automatically analyzed in Matlab®. Results: The Elekta Agility MLC system was shown to be capable of obtaining speeds within the range of 5–35 mm/s. MLC step and shoot tests have demonstrated the MLC system's capability of having positional repeatability, averaging 0.03- and 0.08-mm offsets with and without gravitational effects, respectively. The IMRT-specific tests have shown that gravitational effects are negligible with all positional tests averaging 0.5-mm offsets. The largest speed root-mean-square error (RMSE) for the MLC system was found at the maximum speed of 35 mm/s with an average error of 0.8 mm. For slower speeds, the value was found to be much lower. Conclusion: Utilizing log files has demonstrated the feasibility for higher precision of MLC motions to be reviewed, based on the performance tests that were instituted. Log files provide insight on the effects of friction, acceleration, and gravity, with MU's delivered that previously could not be reviewed in such detail. Based on our results, log file-based QA has enhanced our ability to review performance, functionality, and perform QA on Elekta's MLC system.

Original languageEnglish (US)
Pages (from-to)1397-1407
Number of pages11
JournalMedical physics
Volume46
Issue number3
DOIs
StatePublished - Mar 2019
Externally publishedYes

Keywords

  • Elekta
  • LogFiles
  • MLC
  • QA

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Evaluation of the Elekta Agility MLC performance using high-resolution log files'. Together they form a unique fingerprint.

Cite this