Generation and verification of QFix kVue Calypso-compatible couch top model for a dedicated stereotactic linear accelerator with FFF beams

Stephen J. Gardner, Misbah Gulam, Kwang Song, Haisen Li, Yimei Huang, Bo Zhao, Yujiao Qin, Karen Snyder, Jinkoo Kim, James Gordon, Indrin J. Chetty, Ning Wen

Research output: Contribution to journalArticle

4 Citations (Scopus)

Abstract

This study details the generation, verification, and implementation of a treatment planning system (TPS) couch top model for patient support system used in conjunction with a dedicated stereotactic linear accelerator. Couch top model was created within the TPS using CT simulation images of the kVue Calpyso-compatible couchtop (with rails). Verification measurements were compared to TPS dose prediction for different energies (6 MV FFF and 10 MV FFF) and rail configurations (rails in and rails out) using: 1) central axis point-dose measurements with pinpoint chamber in water-equivalent phantom at 42 gantry angles for various field sizes (2 × 2 cm2, 4 × 4 cm2, 10 × 10 cm2); and 2) Gafchromic EBT3 film parallel to beam in acrylic slab to assess changes in surface and percent depth doses in PA geometry. To assess sensitivity of delivered dose to variations in patient lateral position, measurements at central axis using the pinpoint chamber geometry were taken at lateral couch displacements of 2, 5, and 10 mm for 6 MV FFF. The maximum percent difference for point-dose measurements was 3.24% (6 MV FFF) and 2.30% (10 MV FFF). The average percent difference for point-dose measurements was less than 1.10% for all beam energies and rail geometries. The maximum percent difference between calculated and measured dose can be as large as 13.0% if no couch model is used for dose calculation. The presence of the couch structures also impacts surface dose and PDD, which was evaluated with Gafchromic film measurements. The upstream shift in the depth of dose maximum (dmax) was found to be 10.5 mm for 6 MV FFF and 5.5 mm for 10 MV FFF for 'Rails In' configuration. Transmission of the treatment beam through the couch results in an increase in surface dose (absolute percentage) of approximately 50% for both photon energies (6 MV FFF and 10MV FFF). The largest sensitivity to lateral shifts occurred at the lateral boundary of the rail structures. The mean magnitude (standard deviation) of the deviation between shifted and centered measurements over all field sizes tested was 0.61% (0.61%) for 2 mm shifts, 0.46% (0.67%) for 5 mm shifts, and 0.86% (1.46%) for 10 mm shifts.

Original languageEnglish (US)
Pages (from-to)163-180
Number of pages18
JournalJournal of Applied Clinical Medical Physics
Volume16
Issue number4
StatePublished - 2015

Fingerprint

Calypso
couches
Particle Accelerators
Linear accelerators
linear accelerators
Rails
rails
dosage
Planning
Therapeutics
shift
Geometry
Photons
planning
Position measurement
thiacloprid
Water
geometry
chambers
Acrylics

Keywords

  • Couch modeling
  • Dose delivery uncertainties
  • SRS/SBRT delivery
  • VMAT delivery

ASJC Scopus subject areas

  • Radiation
  • Instrumentation
  • Radiology Nuclear Medicine and imaging

Cite this

Generation and verification of QFix kVue Calypso-compatible couch top model for a dedicated stereotactic linear accelerator with FFF beams. / Gardner, Stephen J.; Gulam, Misbah; Song, Kwang; Li, Haisen; Huang, Yimei; Zhao, Bo; Qin, Yujiao; Snyder, Karen; Kim, Jinkoo; Gordon, James; Chetty, Indrin J.; Wen, Ning.

In: Journal of Applied Clinical Medical Physics, Vol. 16, No. 4, 2015, p. 163-180.

Research output: Contribution to journalArticle

Gardner, SJ, Gulam, M, Song, K, Li, H, Huang, Y, Zhao, B, Qin, Y, Snyder, K, Kim, J, Gordon, J, Chetty, IJ & Wen, N 2015, 'Generation and verification of QFix kVue Calypso-compatible couch top model for a dedicated stereotactic linear accelerator with FFF beams', Journal of Applied Clinical Medical Physics, vol. 16, no. 4, pp. 163-180.
Gardner, Stephen J. ; Gulam, Misbah ; Song, Kwang ; Li, Haisen ; Huang, Yimei ; Zhao, Bo ; Qin, Yujiao ; Snyder, Karen ; Kim, Jinkoo ; Gordon, James ; Chetty, Indrin J. ; Wen, Ning. / Generation and verification of QFix kVue Calypso-compatible couch top model for a dedicated stereotactic linear accelerator with FFF beams. In: Journal of Applied Clinical Medical Physics. 2015 ; Vol. 16, No. 4. pp. 163-180.
@article{5cc748eb9b6c40fc8d5fc129921356c6,
title = "Generation and verification of QFix kVue Calypso-compatible couch top model for a dedicated stereotactic linear accelerator with FFF beams",
abstract = "This study details the generation, verification, and implementation of a treatment planning system (TPS) couch top model for patient support system used in conjunction with a dedicated stereotactic linear accelerator. Couch top model was created within the TPS using CT simulation images of the kVue Calpyso-compatible couchtop (with rails). Verification measurements were compared to TPS dose prediction for different energies (6 MV FFF and 10 MV FFF) and rail configurations (rails in and rails out) using: 1) central axis point-dose measurements with pinpoint chamber in water-equivalent phantom at 42 gantry angles for various field sizes (2 × 2 cm2, 4 × 4 cm2, 10 × 10 cm2); and 2) Gafchromic EBT3 film parallel to beam in acrylic slab to assess changes in surface and percent depth doses in PA geometry. To assess sensitivity of delivered dose to variations in patient lateral position, measurements at central axis using the pinpoint chamber geometry were taken at lateral couch displacements of 2, 5, and 10 mm for 6 MV FFF. The maximum percent difference for point-dose measurements was 3.24{\%} (6 MV FFF) and 2.30{\%} (10 MV FFF). The average percent difference for point-dose measurements was less than 1.10{\%} for all beam energies and rail geometries. The maximum percent difference between calculated and measured dose can be as large as 13.0{\%} if no couch model is used for dose calculation. The presence of the couch structures also impacts surface dose and PDD, which was evaluated with Gafchromic film measurements. The upstream shift in the depth of dose maximum (dmax) was found to be 10.5 mm for 6 MV FFF and 5.5 mm for 10 MV FFF for 'Rails In' configuration. Transmission of the treatment beam through the couch results in an increase in surface dose (absolute percentage) of approximately 50{\%} for both photon energies (6 MV FFF and 10MV FFF). The largest sensitivity to lateral shifts occurred at the lateral boundary of the rail structures. The mean magnitude (standard deviation) of the deviation between shifted and centered measurements over all field sizes tested was 0.61{\%} (0.61{\%}) for 2 mm shifts, 0.46{\%} (0.67{\%}) for 5 mm shifts, and 0.86{\%} (1.46{\%}) for 10 mm shifts.",
keywords = "Couch modeling, Dose delivery uncertainties, SRS/SBRT delivery, VMAT delivery",
author = "Gardner, {Stephen J.} and Misbah Gulam and Kwang Song and Haisen Li and Yimei Huang and Bo Zhao and Yujiao Qin and Karen Snyder and Jinkoo Kim and James Gordon and Chetty, {Indrin J.} and Ning Wen",
year = "2015",
language = "English (US)",
volume = "16",
pages = "163--180",
journal = "Journal of Applied Clinical Medical Physics",
issn = "1526-9914",
publisher = "American Institute of Physics Publising LLC",
number = "4",

}

TY - JOUR

T1 - Generation and verification of QFix kVue Calypso-compatible couch top model for a dedicated stereotactic linear accelerator with FFF beams

AU - Gardner, Stephen J.

AU - Gulam, Misbah

AU - Song, Kwang

AU - Li, Haisen

AU - Huang, Yimei

AU - Zhao, Bo

AU - Qin, Yujiao

AU - Snyder, Karen

AU - Kim, Jinkoo

AU - Gordon, James

AU - Chetty, Indrin J.

AU - Wen, Ning

PY - 2015

Y1 - 2015

N2 - This study details the generation, verification, and implementation of a treatment planning system (TPS) couch top model for patient support system used in conjunction with a dedicated stereotactic linear accelerator. Couch top model was created within the TPS using CT simulation images of the kVue Calpyso-compatible couchtop (with rails). Verification measurements were compared to TPS dose prediction for different energies (6 MV FFF and 10 MV FFF) and rail configurations (rails in and rails out) using: 1) central axis point-dose measurements with pinpoint chamber in water-equivalent phantom at 42 gantry angles for various field sizes (2 × 2 cm2, 4 × 4 cm2, 10 × 10 cm2); and 2) Gafchromic EBT3 film parallel to beam in acrylic slab to assess changes in surface and percent depth doses in PA geometry. To assess sensitivity of delivered dose to variations in patient lateral position, measurements at central axis using the pinpoint chamber geometry were taken at lateral couch displacements of 2, 5, and 10 mm for 6 MV FFF. The maximum percent difference for point-dose measurements was 3.24% (6 MV FFF) and 2.30% (10 MV FFF). The average percent difference for point-dose measurements was less than 1.10% for all beam energies and rail geometries. The maximum percent difference between calculated and measured dose can be as large as 13.0% if no couch model is used for dose calculation. The presence of the couch structures also impacts surface dose and PDD, which was evaluated with Gafchromic film measurements. The upstream shift in the depth of dose maximum (dmax) was found to be 10.5 mm for 6 MV FFF and 5.5 mm for 10 MV FFF for 'Rails In' configuration. Transmission of the treatment beam through the couch results in an increase in surface dose (absolute percentage) of approximately 50% for both photon energies (6 MV FFF and 10MV FFF). The largest sensitivity to lateral shifts occurred at the lateral boundary of the rail structures. The mean magnitude (standard deviation) of the deviation between shifted and centered measurements over all field sizes tested was 0.61% (0.61%) for 2 mm shifts, 0.46% (0.67%) for 5 mm shifts, and 0.86% (1.46%) for 10 mm shifts.

AB - This study details the generation, verification, and implementation of a treatment planning system (TPS) couch top model for patient support system used in conjunction with a dedicated stereotactic linear accelerator. Couch top model was created within the TPS using CT simulation images of the kVue Calpyso-compatible couchtop (with rails). Verification measurements were compared to TPS dose prediction for different energies (6 MV FFF and 10 MV FFF) and rail configurations (rails in and rails out) using: 1) central axis point-dose measurements with pinpoint chamber in water-equivalent phantom at 42 gantry angles for various field sizes (2 × 2 cm2, 4 × 4 cm2, 10 × 10 cm2); and 2) Gafchromic EBT3 film parallel to beam in acrylic slab to assess changes in surface and percent depth doses in PA geometry. To assess sensitivity of delivered dose to variations in patient lateral position, measurements at central axis using the pinpoint chamber geometry were taken at lateral couch displacements of 2, 5, and 10 mm for 6 MV FFF. The maximum percent difference for point-dose measurements was 3.24% (6 MV FFF) and 2.30% (10 MV FFF). The average percent difference for point-dose measurements was less than 1.10% for all beam energies and rail geometries. The maximum percent difference between calculated and measured dose can be as large as 13.0% if no couch model is used for dose calculation. The presence of the couch structures also impacts surface dose and PDD, which was evaluated with Gafchromic film measurements. The upstream shift in the depth of dose maximum (dmax) was found to be 10.5 mm for 6 MV FFF and 5.5 mm for 10 MV FFF for 'Rails In' configuration. Transmission of the treatment beam through the couch results in an increase in surface dose (absolute percentage) of approximately 50% for both photon energies (6 MV FFF and 10MV FFF). The largest sensitivity to lateral shifts occurred at the lateral boundary of the rail structures. The mean magnitude (standard deviation) of the deviation between shifted and centered measurements over all field sizes tested was 0.61% (0.61%) for 2 mm shifts, 0.46% (0.67%) for 5 mm shifts, and 0.86% (1.46%) for 10 mm shifts.

KW - Couch modeling

KW - Dose delivery uncertainties

KW - SRS/SBRT delivery

KW - VMAT delivery

UR - http://www.scopus.com/inward/record.url?scp=84938942390&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84938942390&partnerID=8YFLogxK

M3 - Article

VL - 16

SP - 163

EP - 180

JO - Journal of Applied Clinical Medical Physics

JF - Journal of Applied Clinical Medical Physics

SN - 1526-9914

IS - 4

ER -