Improving the evaluation of alveolar bone grafts with cone beam computerized tomography

Pollyana Marques De Moura, Rami Hallac, Alex Kane, James Seaward

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

Objective: Cone Beam computed tomography (CBCT) is used increasingly as a replacement for periapical x-rays when evaluating alveolar bone grafting. The manufacturer's standard settings for dental imaging may not, however, represent the optimal settings for evaluating postoperative alveolar bone grafts. We examined the influence of exposure parameters on CBCT image quality to optimize the quality of CBCT images while reducing the radiation dose to the minimum level necessary to obtain adequate images. Methods: A defect was created in a cadaver head to simulate an alveolar cleft, and the area was filled with a synthetic material to simulate an alveolar bone graft. Serial CBCT scans were acquired, systematically varying tube voltage and tube current settings from 72 to 96kV and 3 to 12mA. Region of interest analysis was undertaken, and image quality was evaluated by comparing the ratios of native alveolar bone to soft tissue and the ratios of synthetic bone graft to soft tissue and by assessing image noise. Results: Twenty-one CBCT data sets were obtained. Reducing tube voltage (kV) resulted in increased contrast ratio between bone and soft tissue and between synthetic bone graft and soft tissue, with maximal contrast at values of 76 kV/11 mA, 72 kV/12 mA, and 72 kV/11 mA. Of these, the setting with lowest image noise was 76 kV/11 mA. This setting also resulted in a radiation dose of less than half of the manufacturer's recommended settings for the same scan volume. Conclusions: There is potential to improve CBCT image quality significantly while dramatically reducing the radiation dose during postoperative examinations for alveolar bone grafting in patients with cleft lip and palate.

Original languageEnglish (US)
Pages (from-to)57-63
Number of pages7
JournalCleft Palate-Craniofacial Journal
Volume53
Issue number1
DOIs
StatePublished - Jan 1 2016

Fingerprint

Cone-Beam Computed Tomography
Transplants
Bone and Bones
Alveolar Bone Grafting
Radiation
Cleft Lip
Cleft Palate
Cadaver
Tooth
Head
X-Rays

Keywords

  • Alveolar bone graft
  • Cleft lip and palate
  • Cone-beam computed tomography
  • Radiation dose

ASJC Scopus subject areas

  • Otorhinolaryngology
  • Oral Surgery

Cite this

Improving the evaluation of alveolar bone grafts with cone beam computerized tomography. / De Moura, Pollyana Marques; Hallac, Rami; Kane, Alex; Seaward, James.

In: Cleft Palate-Craniofacial Journal, Vol. 53, No. 1, 01.01.2016, p. 57-63.

Research output: Contribution to journalArticle

@article{9ec52971a0bf4970927e15978d8f88d0,
title = "Improving the evaluation of alveolar bone grafts with cone beam computerized tomography",
abstract = "Objective: Cone Beam computed tomography (CBCT) is used increasingly as a replacement for periapical x-rays when evaluating alveolar bone grafting. The manufacturer's standard settings for dental imaging may not, however, represent the optimal settings for evaluating postoperative alveolar bone grafts. We examined the influence of exposure parameters on CBCT image quality to optimize the quality of CBCT images while reducing the radiation dose to the minimum level necessary to obtain adequate images. Methods: A defect was created in a cadaver head to simulate an alveolar cleft, and the area was filled with a synthetic material to simulate an alveolar bone graft. Serial CBCT scans were acquired, systematically varying tube voltage and tube current settings from 72 to 96kV and 3 to 12mA. Region of interest analysis was undertaken, and image quality was evaluated by comparing the ratios of native alveolar bone to soft tissue and the ratios of synthetic bone graft to soft tissue and by assessing image noise. Results: Twenty-one CBCT data sets were obtained. Reducing tube voltage (kV) resulted in increased contrast ratio between bone and soft tissue and between synthetic bone graft and soft tissue, with maximal contrast at values of 76 kV/11 mA, 72 kV/12 mA, and 72 kV/11 mA. Of these, the setting with lowest image noise was 76 kV/11 mA. This setting also resulted in a radiation dose of less than half of the manufacturer's recommended settings for the same scan volume. Conclusions: There is potential to improve CBCT image quality significantly while dramatically reducing the radiation dose during postoperative examinations for alveolar bone grafting in patients with cleft lip and palate.",
keywords = "Alveolar bone graft, Cleft lip and palate, Cone-beam computed tomography, Radiation dose",
author = "{De Moura}, {Pollyana Marques} and Rami Hallac and Alex Kane and James Seaward",
year = "2016",
month = "1",
day = "1",
doi = "10.1597/14-304",
language = "English (US)",
volume = "53",
pages = "57--63",
journal = "Cleft Palate-Craniofacial Journal",
issn = "1055-6656",
publisher = "American Cleft Palate Craniofacial Association",
number = "1",

}

TY - JOUR

T1 - Improving the evaluation of alveolar bone grafts with cone beam computerized tomography

AU - De Moura, Pollyana Marques

AU - Hallac, Rami

AU - Kane, Alex

AU - Seaward, James

PY - 2016/1/1

Y1 - 2016/1/1

N2 - Objective: Cone Beam computed tomography (CBCT) is used increasingly as a replacement for periapical x-rays when evaluating alveolar bone grafting. The manufacturer's standard settings for dental imaging may not, however, represent the optimal settings for evaluating postoperative alveolar bone grafts. We examined the influence of exposure parameters on CBCT image quality to optimize the quality of CBCT images while reducing the radiation dose to the minimum level necessary to obtain adequate images. Methods: A defect was created in a cadaver head to simulate an alveolar cleft, and the area was filled with a synthetic material to simulate an alveolar bone graft. Serial CBCT scans were acquired, systematically varying tube voltage and tube current settings from 72 to 96kV and 3 to 12mA. Region of interest analysis was undertaken, and image quality was evaluated by comparing the ratios of native alveolar bone to soft tissue and the ratios of synthetic bone graft to soft tissue and by assessing image noise. Results: Twenty-one CBCT data sets were obtained. Reducing tube voltage (kV) resulted in increased contrast ratio between bone and soft tissue and between synthetic bone graft and soft tissue, with maximal contrast at values of 76 kV/11 mA, 72 kV/12 mA, and 72 kV/11 mA. Of these, the setting with lowest image noise was 76 kV/11 mA. This setting also resulted in a radiation dose of less than half of the manufacturer's recommended settings for the same scan volume. Conclusions: There is potential to improve CBCT image quality significantly while dramatically reducing the radiation dose during postoperative examinations for alveolar bone grafting in patients with cleft lip and palate.

AB - Objective: Cone Beam computed tomography (CBCT) is used increasingly as a replacement for periapical x-rays when evaluating alveolar bone grafting. The manufacturer's standard settings for dental imaging may not, however, represent the optimal settings for evaluating postoperative alveolar bone grafts. We examined the influence of exposure parameters on CBCT image quality to optimize the quality of CBCT images while reducing the radiation dose to the minimum level necessary to obtain adequate images. Methods: A defect was created in a cadaver head to simulate an alveolar cleft, and the area was filled with a synthetic material to simulate an alveolar bone graft. Serial CBCT scans were acquired, systematically varying tube voltage and tube current settings from 72 to 96kV and 3 to 12mA. Region of interest analysis was undertaken, and image quality was evaluated by comparing the ratios of native alveolar bone to soft tissue and the ratios of synthetic bone graft to soft tissue and by assessing image noise. Results: Twenty-one CBCT data sets were obtained. Reducing tube voltage (kV) resulted in increased contrast ratio between bone and soft tissue and between synthetic bone graft and soft tissue, with maximal contrast at values of 76 kV/11 mA, 72 kV/12 mA, and 72 kV/11 mA. Of these, the setting with lowest image noise was 76 kV/11 mA. This setting also resulted in a radiation dose of less than half of the manufacturer's recommended settings for the same scan volume. Conclusions: There is potential to improve CBCT image quality significantly while dramatically reducing the radiation dose during postoperative examinations for alveolar bone grafting in patients with cleft lip and palate.

KW - Alveolar bone graft

KW - Cleft lip and palate

KW - Cone-beam computed tomography

KW - Radiation dose

UR - http://www.scopus.com/inward/record.url?scp=84957655188&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84957655188&partnerID=8YFLogxK

U2 - 10.1597/14-304

DO - 10.1597/14-304

M3 - Article

C2 - 26101811

AN - SCOPUS:84957655188

VL - 53

SP - 57

EP - 63

JO - Cleft Palate-Craniofacial Journal

JF - Cleft Palate-Craniofacial Journal

SN - 1055-6656

IS - 1

ER -