Localised hyperthermia in rodent models using an MRI-compatible high-intensity focused ultrasound system

Chenchen Bing, Joris Nofiele, Robert Staruch, Michelle Ladouceur-Wodzak, Yonatan Chatzinoff, Ashish Ranjan, Rajiv Chopra

Research output: Contribution to journalArticle

11 Citations (Scopus)

Abstract

Purpose: Localised hyperthermia in rodent studies is challenging due to the small target size. This study describes the development and characterisation of an MRI-compatible high-intensity focused ultrasound (HIFU) system to perform localised mild hyperthermia treatments in rodent models. Material and methods: The hyperthermia platform consisted of an MRI-compatible small animal HIFU system, focused transducers with sector-vortex lenses, a custom-made receive coil, and means to maintain systemic temperatures of rodents. The system was integrated into a 3T MR imager. Control software was developed to acquire images, process temperature maps, and adjust output power using a proportional-integral-derivative feedback control algorithm. Hyperthermia exposures were performed in tissue-mimicking phantoms and in a rodent model (n = 9). During heating, an ROI was assigned in the heated region for temperature control and the target temperature was 42 °C; 30 min mild hyperthermia treatment followed by a 10-min cooling procedure was performed on each animal. Results: 3D-printed sector-vortex lenses were successful at creating annular focal regions which enables customisation of the heating volume. Localised mild hyperthermia performed in rats produced a mean ROI temperature of 42.1 ± 0.3 °C. The T10 and T90 percentiles were 43.2 ± 0.4 °C and 41.0 ± 0.3 °C, respectively. For a 30-min treatment, the mean time duration between 41-45 °C was 31.1 min within the ROI. Conclusions: The MRI-compatible HIFU system was successfully adapted to perform localised mild hyperthermia treatment in rodent models. A target temperature of 42 °C was well-maintained in a rat thigh model for 30 min.

Original languageEnglish (US)
Pages (from-to)813-822
Number of pages10
JournalInternational Journal of Hyperthermia
Volume31
Issue number8
DOIs
StatePublished - Nov 17 2015

Fingerprint

Rodentia
Fever
Temperature
Heating
Lenses
Thigh
Transducers
Software

Keywords

  • Heat-targeted drug delivery
  • mild hyperthermia
  • MR-HIFU
  • nanoparticles
  • non-invasive thermometry

ASJC Scopus subject areas

  • Cancer Research
  • Physiology
  • Radiological and Ultrasound Technology
  • Physiology (medical)

Cite this

Localised hyperthermia in rodent models using an MRI-compatible high-intensity focused ultrasound system. / Bing, Chenchen; Nofiele, Joris; Staruch, Robert; Ladouceur-Wodzak, Michelle; Chatzinoff, Yonatan; Ranjan, Ashish; Chopra, Rajiv.

In: International Journal of Hyperthermia, Vol. 31, No. 8, 17.11.2015, p. 813-822.

Research output: Contribution to journalArticle

Bing, Chenchen ; Nofiele, Joris ; Staruch, Robert ; Ladouceur-Wodzak, Michelle ; Chatzinoff, Yonatan ; Ranjan, Ashish ; Chopra, Rajiv. / Localised hyperthermia in rodent models using an MRI-compatible high-intensity focused ultrasound system. In: International Journal of Hyperthermia. 2015 ; Vol. 31, No. 8. pp. 813-822.
@article{0df976daf1f84ba584cb2ab6f8321b16,
title = "Localised hyperthermia in rodent models using an MRI-compatible high-intensity focused ultrasound system",
abstract = "Purpose: Localised hyperthermia in rodent studies is challenging due to the small target size. This study describes the development and characterisation of an MRI-compatible high-intensity focused ultrasound (HIFU) system to perform localised mild hyperthermia treatments in rodent models. Material and methods: The hyperthermia platform consisted of an MRI-compatible small animal HIFU system, focused transducers with sector-vortex lenses, a custom-made receive coil, and means to maintain systemic temperatures of rodents. The system was integrated into a 3T MR imager. Control software was developed to acquire images, process temperature maps, and adjust output power using a proportional-integral-derivative feedback control algorithm. Hyperthermia exposures were performed in tissue-mimicking phantoms and in a rodent model (n = 9). During heating, an ROI was assigned in the heated region for temperature control and the target temperature was 42 °C; 30 min mild hyperthermia treatment followed by a 10-min cooling procedure was performed on each animal. Results: 3D-printed sector-vortex lenses were successful at creating annular focal regions which enables customisation of the heating volume. Localised mild hyperthermia performed in rats produced a mean ROI temperature of 42.1 ± 0.3 °C. The T10 and T90 percentiles were 43.2 ± 0.4 °C and 41.0 ± 0.3 °C, respectively. For a 30-min treatment, the mean time duration between 41-45 °C was 31.1 min within the ROI. Conclusions: The MRI-compatible HIFU system was successfully adapted to perform localised mild hyperthermia treatment in rodent models. A target temperature of 42 °C was well-maintained in a rat thigh model for 30 min.",
keywords = "Heat-targeted drug delivery, mild hyperthermia, MR-HIFU, nanoparticles, non-invasive thermometry",
author = "Chenchen Bing and Joris Nofiele and Robert Staruch and Michelle Ladouceur-Wodzak and Yonatan Chatzinoff and Ashish Ranjan and Rajiv Chopra",
year = "2015",
month = "11",
day = "17",
doi = "10.3109/02656736.2015.1094833",
language = "English (US)",
volume = "31",
pages = "813--822",
journal = "International Journal of Hyperthermia",
issn = "0265-6736",
publisher = "Informa Healthcare",
number = "8",

}

TY - JOUR

T1 - Localised hyperthermia in rodent models using an MRI-compatible high-intensity focused ultrasound system

AU - Bing, Chenchen

AU - Nofiele, Joris

AU - Staruch, Robert

AU - Ladouceur-Wodzak, Michelle

AU - Chatzinoff, Yonatan

AU - Ranjan, Ashish

AU - Chopra, Rajiv

PY - 2015/11/17

Y1 - 2015/11/17

N2 - Purpose: Localised hyperthermia in rodent studies is challenging due to the small target size. This study describes the development and characterisation of an MRI-compatible high-intensity focused ultrasound (HIFU) system to perform localised mild hyperthermia treatments in rodent models. Material and methods: The hyperthermia platform consisted of an MRI-compatible small animal HIFU system, focused transducers with sector-vortex lenses, a custom-made receive coil, and means to maintain systemic temperatures of rodents. The system was integrated into a 3T MR imager. Control software was developed to acquire images, process temperature maps, and adjust output power using a proportional-integral-derivative feedback control algorithm. Hyperthermia exposures were performed in tissue-mimicking phantoms and in a rodent model (n = 9). During heating, an ROI was assigned in the heated region for temperature control and the target temperature was 42 °C; 30 min mild hyperthermia treatment followed by a 10-min cooling procedure was performed on each animal. Results: 3D-printed sector-vortex lenses were successful at creating annular focal regions which enables customisation of the heating volume. Localised mild hyperthermia performed in rats produced a mean ROI temperature of 42.1 ± 0.3 °C. The T10 and T90 percentiles were 43.2 ± 0.4 °C and 41.0 ± 0.3 °C, respectively. For a 30-min treatment, the mean time duration between 41-45 °C was 31.1 min within the ROI. Conclusions: The MRI-compatible HIFU system was successfully adapted to perform localised mild hyperthermia treatment in rodent models. A target temperature of 42 °C was well-maintained in a rat thigh model for 30 min.

AB - Purpose: Localised hyperthermia in rodent studies is challenging due to the small target size. This study describes the development and characterisation of an MRI-compatible high-intensity focused ultrasound (HIFU) system to perform localised mild hyperthermia treatments in rodent models. Material and methods: The hyperthermia platform consisted of an MRI-compatible small animal HIFU system, focused transducers with sector-vortex lenses, a custom-made receive coil, and means to maintain systemic temperatures of rodents. The system was integrated into a 3T MR imager. Control software was developed to acquire images, process temperature maps, and adjust output power using a proportional-integral-derivative feedback control algorithm. Hyperthermia exposures were performed in tissue-mimicking phantoms and in a rodent model (n = 9). During heating, an ROI was assigned in the heated region for temperature control and the target temperature was 42 °C; 30 min mild hyperthermia treatment followed by a 10-min cooling procedure was performed on each animal. Results: 3D-printed sector-vortex lenses were successful at creating annular focal regions which enables customisation of the heating volume. Localised mild hyperthermia performed in rats produced a mean ROI temperature of 42.1 ± 0.3 °C. The T10 and T90 percentiles were 43.2 ± 0.4 °C and 41.0 ± 0.3 °C, respectively. For a 30-min treatment, the mean time duration between 41-45 °C was 31.1 min within the ROI. Conclusions: The MRI-compatible HIFU system was successfully adapted to perform localised mild hyperthermia treatment in rodent models. A target temperature of 42 °C was well-maintained in a rat thigh model for 30 min.

KW - Heat-targeted drug delivery

KW - mild hyperthermia

KW - MR-HIFU

KW - nanoparticles

KW - non-invasive thermometry

UR - http://www.scopus.com/inward/record.url?scp=84953355035&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84953355035&partnerID=8YFLogxK

U2 - 10.3109/02656736.2015.1094833

DO - 10.3109/02656736.2015.1094833

M3 - Article

C2 - 26540488

AN - SCOPUS:84953355035

VL - 31

SP - 813

EP - 822

JO - International Journal of Hyperthermia

JF - International Journal of Hyperthermia

SN - 0265-6736

IS - 8

ER -