Megavoltage image-based dynamic multileaf collimator tracking of a NiTi stent in porcine lungs on a linear accelerator

Per R. Poulsen, Jesper Carl, Jane Nielsen, Martin S. Nielsen, Jakob B. Thomsen, Henrik K. Jensen, Benedict Kjærgaard, Peter R. Zepernick, Esben Worm, Walther Fledelius, Byungchul Cho, Amit Sawant, Dan Ruan, Paul J. Keall

Research output: Contribution to journalArticle

18 Citations (Scopus)

Abstract

Purpose: To investigate the accuracy and potential limitations of MV image-based dynamic multileaf collimator (DMLC) tracking in a porcine model on a linear accelerator. Methods and Materials: A thermo-expandable NiTi stent designed for kilovoltage (kV) X-ray visualization of lung lesions was inserted into the bronchia of three anaesthetized Göttingen minipigs. A four-dimensional computed tomography scan was used for planning a five-field conformal treatment with circular multileaf collimator (MLC) apertures. A 22.5 Gy single fraction treatment was delivered to the pigs. The peak-to-peak stent motion was 3 to 8 mm, with breathing periods of 1.2 to 4 s. Before treatment, X-ray images were used for image-guided setup based on the stent. During treatment delivery, continuous megavoltage (MV) portal images were acquired at 7.5 Hz. The stent was segmented in the images and used for continuous adaptation of the MLC aperture. Offline, the tracking error in beam's eye view of the treatment beam was calculated for each MV image as the difference between the MLC aperture center and the segmented stent position. The standard deviations of the systematic error Σ and the random error σ were determined and compared with the would-be errors for a nontracking treatment with pretreatment image-guided setup. Results: Reliable stent segmentation was obtained for 11 of 15 fields. Segmentation failures occurred when image contrast was dominated by overlapping anatomical structures (ribs, diaphragm) rather than by the stent, which was designed for kV rather than MV X-ray visibility. For the 11 fields with reliable segmentation, Σ was 0.5 mm/0.4 mm in the two imager directions, whereas σ was 0.5 mm/1.1 mm. Without tracking, Σ and σ would have been 1.7 mm/1.4 mm and 0.8 mm/1.4 mm, respectively. Conclusion: For the first time, in vivo DMLC tracking has been demonstrated on a linear accelerator showing the potential for improved targeting accuracy. The study mimicked the envisioned patient workflow of future patient treatments. Clinical implementation of MV image-based tracking would require markers designed for MV visibility.

Original languageEnglish (US)
JournalInternational Journal of Radiation Oncology Biology Physics
Volume82
Issue number2
DOIs
StatePublished - Feb 1 2012

Fingerprint

Particle Accelerators
linear accelerators
collimators
lungs
Stents
Swine
Lung
X-Rays
apertures
visibility
Four-Dimensional Computed Tomography
Therapeutics
Miniature Swine
Workflow
Ribs
x rays
swine
Diaphragm
random errors
image contrast

Keywords

  • Dynamic MLC tracking
  • Image-guided radiotherapy
  • Intrafraction motion
  • Tumor tracking

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging
  • Radiation
  • Cancer Research

Cite this

Megavoltage image-based dynamic multileaf collimator tracking of a NiTi stent in porcine lungs on a linear accelerator. / Poulsen, Per R.; Carl, Jesper; Nielsen, Jane; Nielsen, Martin S.; Thomsen, Jakob B.; Jensen, Henrik K.; Kjærgaard, Benedict; Zepernick, Peter R.; Worm, Esben; Fledelius, Walther; Cho, Byungchul; Sawant, Amit; Ruan, Dan; Keall, Paul J.

In: International Journal of Radiation Oncology Biology Physics, Vol. 82, No. 2, 01.02.2012.

Research output: Contribution to journalArticle

Poulsen, PR, Carl, J, Nielsen, J, Nielsen, MS, Thomsen, JB, Jensen, HK, Kjærgaard, B, Zepernick, PR, Worm, E, Fledelius, W, Cho, B, Sawant, A, Ruan, D & Keall, PJ 2012, 'Megavoltage image-based dynamic multileaf collimator tracking of a NiTi stent in porcine lungs on a linear accelerator', International Journal of Radiation Oncology Biology Physics, vol. 82, no. 2. https://doi.org/10.1016/j.ijrobp.2011.03.023
Poulsen, Per R. ; Carl, Jesper ; Nielsen, Jane ; Nielsen, Martin S. ; Thomsen, Jakob B. ; Jensen, Henrik K. ; Kjærgaard, Benedict ; Zepernick, Peter R. ; Worm, Esben ; Fledelius, Walther ; Cho, Byungchul ; Sawant, Amit ; Ruan, Dan ; Keall, Paul J. / Megavoltage image-based dynamic multileaf collimator tracking of a NiTi stent in porcine lungs on a linear accelerator. In: International Journal of Radiation Oncology Biology Physics. 2012 ; Vol. 82, No. 2.
@article{dcda4babab844c01b5477e55bd348cd1,
title = "Megavoltage image-based dynamic multileaf collimator tracking of a NiTi stent in porcine lungs on a linear accelerator",
abstract = "Purpose: To investigate the accuracy and potential limitations of MV image-based dynamic multileaf collimator (DMLC) tracking in a porcine model on a linear accelerator. Methods and Materials: A thermo-expandable NiTi stent designed for kilovoltage (kV) X-ray visualization of lung lesions was inserted into the bronchia of three anaesthetized G{\"o}ttingen minipigs. A four-dimensional computed tomography scan was used for planning a five-field conformal treatment with circular multileaf collimator (MLC) apertures. A 22.5 Gy single fraction treatment was delivered to the pigs. The peak-to-peak stent motion was 3 to 8 mm, with breathing periods of 1.2 to 4 s. Before treatment, X-ray images were used for image-guided setup based on the stent. During treatment delivery, continuous megavoltage (MV) portal images were acquired at 7.5 Hz. The stent was segmented in the images and used for continuous adaptation of the MLC aperture. Offline, the tracking error in beam's eye view of the treatment beam was calculated for each MV image as the difference between the MLC aperture center and the segmented stent position. The standard deviations of the systematic error Σ and the random error σ were determined and compared with the would-be errors for a nontracking treatment with pretreatment image-guided setup. Results: Reliable stent segmentation was obtained for 11 of 15 fields. Segmentation failures occurred when image contrast was dominated by overlapping anatomical structures (ribs, diaphragm) rather than by the stent, which was designed for kV rather than MV X-ray visibility. For the 11 fields with reliable segmentation, Σ was 0.5 mm/0.4 mm in the two imager directions, whereas σ was 0.5 mm/1.1 mm. Without tracking, Σ and σ would have been 1.7 mm/1.4 mm and 0.8 mm/1.4 mm, respectively. Conclusion: For the first time, in vivo DMLC tracking has been demonstrated on a linear accelerator showing the potential for improved targeting accuracy. The study mimicked the envisioned patient workflow of future patient treatments. Clinical implementation of MV image-based tracking would require markers designed for MV visibility.",
keywords = "Dynamic MLC tracking, Image-guided radiotherapy, Intrafraction motion, Tumor tracking",
author = "Poulsen, {Per R.} and Jesper Carl and Jane Nielsen and Nielsen, {Martin S.} and Thomsen, {Jakob B.} and Jensen, {Henrik K.} and Benedict Kj{\ae}rgaard and Zepernick, {Peter R.} and Esben Worm and Walther Fledelius and Byungchul Cho and Amit Sawant and Dan Ruan and Keall, {Paul J.}",
year = "2012",
month = "2",
day = "1",
doi = "10.1016/j.ijrobp.2011.03.023",
language = "English (US)",
volume = "82",
journal = "International Journal of Radiation Oncology Biology Physics",
issn = "0360-3016",
publisher = "Elsevier Inc.",
number = "2",

}

TY - JOUR

T1 - Megavoltage image-based dynamic multileaf collimator tracking of a NiTi stent in porcine lungs on a linear accelerator

AU - Poulsen, Per R.

AU - Carl, Jesper

AU - Nielsen, Jane

AU - Nielsen, Martin S.

AU - Thomsen, Jakob B.

AU - Jensen, Henrik K.

AU - Kjærgaard, Benedict

AU - Zepernick, Peter R.

AU - Worm, Esben

AU - Fledelius, Walther

AU - Cho, Byungchul

AU - Sawant, Amit

AU - Ruan, Dan

AU - Keall, Paul J.

PY - 2012/2/1

Y1 - 2012/2/1

N2 - Purpose: To investigate the accuracy and potential limitations of MV image-based dynamic multileaf collimator (DMLC) tracking in a porcine model on a linear accelerator. Methods and Materials: A thermo-expandable NiTi stent designed for kilovoltage (kV) X-ray visualization of lung lesions was inserted into the bronchia of three anaesthetized Göttingen minipigs. A four-dimensional computed tomography scan was used for planning a five-field conformal treatment with circular multileaf collimator (MLC) apertures. A 22.5 Gy single fraction treatment was delivered to the pigs. The peak-to-peak stent motion was 3 to 8 mm, with breathing periods of 1.2 to 4 s. Before treatment, X-ray images were used for image-guided setup based on the stent. During treatment delivery, continuous megavoltage (MV) portal images were acquired at 7.5 Hz. The stent was segmented in the images and used for continuous adaptation of the MLC aperture. Offline, the tracking error in beam's eye view of the treatment beam was calculated for each MV image as the difference between the MLC aperture center and the segmented stent position. The standard deviations of the systematic error Σ and the random error σ were determined and compared with the would-be errors for a nontracking treatment with pretreatment image-guided setup. Results: Reliable stent segmentation was obtained for 11 of 15 fields. Segmentation failures occurred when image contrast was dominated by overlapping anatomical structures (ribs, diaphragm) rather than by the stent, which was designed for kV rather than MV X-ray visibility. For the 11 fields with reliable segmentation, Σ was 0.5 mm/0.4 mm in the two imager directions, whereas σ was 0.5 mm/1.1 mm. Without tracking, Σ and σ would have been 1.7 mm/1.4 mm and 0.8 mm/1.4 mm, respectively. Conclusion: For the first time, in vivo DMLC tracking has been demonstrated on a linear accelerator showing the potential for improved targeting accuracy. The study mimicked the envisioned patient workflow of future patient treatments. Clinical implementation of MV image-based tracking would require markers designed for MV visibility.

AB - Purpose: To investigate the accuracy and potential limitations of MV image-based dynamic multileaf collimator (DMLC) tracking in a porcine model on a linear accelerator. Methods and Materials: A thermo-expandable NiTi stent designed for kilovoltage (kV) X-ray visualization of lung lesions was inserted into the bronchia of three anaesthetized Göttingen minipigs. A four-dimensional computed tomography scan was used for planning a five-field conformal treatment with circular multileaf collimator (MLC) apertures. A 22.5 Gy single fraction treatment was delivered to the pigs. The peak-to-peak stent motion was 3 to 8 mm, with breathing periods of 1.2 to 4 s. Before treatment, X-ray images were used for image-guided setup based on the stent. During treatment delivery, continuous megavoltage (MV) portal images were acquired at 7.5 Hz. The stent was segmented in the images and used for continuous adaptation of the MLC aperture. Offline, the tracking error in beam's eye view of the treatment beam was calculated for each MV image as the difference between the MLC aperture center and the segmented stent position. The standard deviations of the systematic error Σ and the random error σ were determined and compared with the would-be errors for a nontracking treatment with pretreatment image-guided setup. Results: Reliable stent segmentation was obtained for 11 of 15 fields. Segmentation failures occurred when image contrast was dominated by overlapping anatomical structures (ribs, diaphragm) rather than by the stent, which was designed for kV rather than MV X-ray visibility. For the 11 fields with reliable segmentation, Σ was 0.5 mm/0.4 mm in the two imager directions, whereas σ was 0.5 mm/1.1 mm. Without tracking, Σ and σ would have been 1.7 mm/1.4 mm and 0.8 mm/1.4 mm, respectively. Conclusion: For the first time, in vivo DMLC tracking has been demonstrated on a linear accelerator showing the potential for improved targeting accuracy. The study mimicked the envisioned patient workflow of future patient treatments. Clinical implementation of MV image-based tracking would require markers designed for MV visibility.

KW - Dynamic MLC tracking

KW - Image-guided radiotherapy

KW - Intrafraction motion

KW - Tumor tracking

UR - http://www.scopus.com/inward/record.url?scp=84855841957&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84855841957&partnerID=8YFLogxK

U2 - 10.1016/j.ijrobp.2011.03.023

DO - 10.1016/j.ijrobp.2011.03.023

M3 - Article

VL - 82

JO - International Journal of Radiation Oncology Biology Physics

JF - International Journal of Radiation Oncology Biology Physics

SN - 0360-3016

IS - 2

ER -