MO‐EE‐A1‐02: Dosimetry for an Ophthalmic Applicator of P‐32

C. Choi, J. Kim, J. Park, S. Park, Y. Choi, H. Kim, Y. Park, H. Kim, S. ye

Research output: Contribution to journalArticle

Abstract

Purpose: A 90Sr/Y applicator has been used as a ‐source for postoperative irradiation after pterygium excision. As an alternative to 90Sr/Y irradiation, we proposed treatments with 32P. This study aims to provide the dosimetry for this new applicator. Method and Materials: In order to optimize the design and materials of 32P ophthalmic applicators, Monte Carlo simulations were performed. The absorbed dose at the surface of a sealed beta source is often measured by using an extrapolation ionization chamber. Radiochromic film (RCF) was used to measure depth dose distributions and dose profiles at various depths. A micro‐MOSFET detector was used for depth dose measurements. Results: The absorbed dose rates to the reference point were 0.238 ± 0.012 cGy/s for an extrapolation ionization chamber, 0.280 ± 0.001 cGy/s for radiochromic films, and 0.257 ± 0.020 cGy/s for MOSFET. The axial depth dose rate was reduced into approximately 1/10 as 32P betas penetrate every 2 mm depth. Measured data sets in depths of 1 mm to 3.5 mm agreed with Monte Carlo data. Due to non‐uniform absorption of 32P into an absorbent disk, the dose at the center of transaxial plane were 2%–4% less than the peak dose around the periphery. We confirmed no leakage of 32P activities and negligible exposure rate around the hand grip of the applicator. Conclusions: The 32P applicator can deliver uniform therapeutic doses to the surface of the conjunctiva, while sparing the lens better than 90Sr/Y applicators. The doses at any points from the 32P applicator can be calculated by using these measured data sets. The safety of 32P applicator was confirmed. However, prior to the clinical application of every new applicator, safety, dose uniformity, and absorbed dose rate at the reference point should be carefully evaluated by the method developed in this study.

Original languageEnglish (US)
Number of pages1
JournalMedical Physics
Volume37
Issue number6
DOIs
StatePublished - Jan 1 2010

Fingerprint

Pterygium
Safety
Conjunctiva
Hand Strength
Lenses
Hand
Datasets
Therapeutics

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

Choi, C., Kim, J., Park, J., Park, S., Choi, Y., Kim, H., ... ye, S. (2010). MO‐EE‐A1‐02: Dosimetry for an Ophthalmic Applicator of P‐32. Medical Physics, 37(6). https://doi.org/10.1118/1.3469080

MO‐EE‐A1‐02 : Dosimetry for an Ophthalmic Applicator of P‐32. / Choi, C.; Kim, J.; Park, J.; Park, S.; Choi, Y.; Kim, H.; Park, Y.; Kim, H.; ye, S.

In: Medical Physics, Vol. 37, No. 6, 01.01.2010.

Research output: Contribution to journalArticle

Choi, C, Kim, J, Park, J, Park, S, Choi, Y, Kim, H, Park, Y, Kim, H & ye, S 2010, 'MO‐EE‐A1‐02: Dosimetry for an Ophthalmic Applicator of P‐32', Medical Physics, vol. 37, no. 6. https://doi.org/10.1118/1.3469080
Choi, C. ; Kim, J. ; Park, J. ; Park, S. ; Choi, Y. ; Kim, H. ; Park, Y. ; Kim, H. ; ye, S. / MO‐EE‐A1‐02 : Dosimetry for an Ophthalmic Applicator of P‐32. In: Medical Physics. 2010 ; Vol. 37, No. 6.
@article{472410a1d7a14e749694dd944e9a50fb,
title = "MO‐EE‐A1‐02: Dosimetry for an Ophthalmic Applicator of P‐32",
abstract = "Purpose: A 90Sr/Y applicator has been used as a ‐source for postoperative irradiation after pterygium excision. As an alternative to 90Sr/Y irradiation, we proposed treatments with 32P. This study aims to provide the dosimetry for this new applicator. Method and Materials: In order to optimize the design and materials of 32P ophthalmic applicators, Monte Carlo simulations were performed. The absorbed dose at the surface of a sealed beta source is often measured by using an extrapolation ionization chamber. Radiochromic film (RCF) was used to measure depth dose distributions and dose profiles at various depths. A micro‐MOSFET detector was used for depth dose measurements. Results: The absorbed dose rates to the reference point were 0.238 ± 0.012 cGy/s for an extrapolation ionization chamber, 0.280 ± 0.001 cGy/s for radiochromic films, and 0.257 ± 0.020 cGy/s for MOSFET. The axial depth dose rate was reduced into approximately 1/10 as 32P betas penetrate every 2 mm depth. Measured data sets in depths of 1 mm to 3.5 mm agreed with Monte Carlo data. Due to non‐uniform absorption of 32P into an absorbent disk, the dose at the center of transaxial plane were 2{\%}–4{\%} less than the peak dose around the periphery. We confirmed no leakage of 32P activities and negligible exposure rate around the hand grip of the applicator. Conclusions: The 32P applicator can deliver uniform therapeutic doses to the surface of the conjunctiva, while sparing the lens better than 90Sr/Y applicators. The doses at any points from the 32P applicator can be calculated by using these measured data sets. The safety of 32P applicator was confirmed. However, prior to the clinical application of every new applicator, safety, dose uniformity, and absorbed dose rate at the reference point should be carefully evaluated by the method developed in this study.",
author = "C. Choi and J. Kim and J. Park and S. Park and Y. Choi and H. Kim and Y. Park and H. Kim and S. ye",
year = "2010",
month = "1",
day = "1",
doi = "10.1118/1.3469080",
language = "English (US)",
volume = "37",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - MO‐EE‐A1‐02

T2 - Dosimetry for an Ophthalmic Applicator of P‐32

AU - Choi, C.

AU - Kim, J.

AU - Park, J.

AU - Park, S.

AU - Choi, Y.

AU - Kim, H.

AU - Park, Y.

AU - Kim, H.

AU - ye, S.

PY - 2010/1/1

Y1 - 2010/1/1

N2 - Purpose: A 90Sr/Y applicator has been used as a ‐source for postoperative irradiation after pterygium excision. As an alternative to 90Sr/Y irradiation, we proposed treatments with 32P. This study aims to provide the dosimetry for this new applicator. Method and Materials: In order to optimize the design and materials of 32P ophthalmic applicators, Monte Carlo simulations were performed. The absorbed dose at the surface of a sealed beta source is often measured by using an extrapolation ionization chamber. Radiochromic film (RCF) was used to measure depth dose distributions and dose profiles at various depths. A micro‐MOSFET detector was used for depth dose measurements. Results: The absorbed dose rates to the reference point were 0.238 ± 0.012 cGy/s for an extrapolation ionization chamber, 0.280 ± 0.001 cGy/s for radiochromic films, and 0.257 ± 0.020 cGy/s for MOSFET. The axial depth dose rate was reduced into approximately 1/10 as 32P betas penetrate every 2 mm depth. Measured data sets in depths of 1 mm to 3.5 mm agreed with Monte Carlo data. Due to non‐uniform absorption of 32P into an absorbent disk, the dose at the center of transaxial plane were 2%–4% less than the peak dose around the periphery. We confirmed no leakage of 32P activities and negligible exposure rate around the hand grip of the applicator. Conclusions: The 32P applicator can deliver uniform therapeutic doses to the surface of the conjunctiva, while sparing the lens better than 90Sr/Y applicators. The doses at any points from the 32P applicator can be calculated by using these measured data sets. The safety of 32P applicator was confirmed. However, prior to the clinical application of every new applicator, safety, dose uniformity, and absorbed dose rate at the reference point should be carefully evaluated by the method developed in this study.

AB - Purpose: A 90Sr/Y applicator has been used as a ‐source for postoperative irradiation after pterygium excision. As an alternative to 90Sr/Y irradiation, we proposed treatments with 32P. This study aims to provide the dosimetry for this new applicator. Method and Materials: In order to optimize the design and materials of 32P ophthalmic applicators, Monte Carlo simulations were performed. The absorbed dose at the surface of a sealed beta source is often measured by using an extrapolation ionization chamber. Radiochromic film (RCF) was used to measure depth dose distributions and dose profiles at various depths. A micro‐MOSFET detector was used for depth dose measurements. Results: The absorbed dose rates to the reference point were 0.238 ± 0.012 cGy/s for an extrapolation ionization chamber, 0.280 ± 0.001 cGy/s for radiochromic films, and 0.257 ± 0.020 cGy/s for MOSFET. The axial depth dose rate was reduced into approximately 1/10 as 32P betas penetrate every 2 mm depth. Measured data sets in depths of 1 mm to 3.5 mm agreed with Monte Carlo data. Due to non‐uniform absorption of 32P into an absorbent disk, the dose at the center of transaxial plane were 2%–4% less than the peak dose around the periphery. We confirmed no leakage of 32P activities and negligible exposure rate around the hand grip of the applicator. Conclusions: The 32P applicator can deliver uniform therapeutic doses to the surface of the conjunctiva, while sparing the lens better than 90Sr/Y applicators. The doses at any points from the 32P applicator can be calculated by using these measured data sets. The safety of 32P applicator was confirmed. However, prior to the clinical application of every new applicator, safety, dose uniformity, and absorbed dose rate at the reference point should be carefully evaluated by the method developed in this study.

UR - http://www.scopus.com/inward/record.url?scp=85024805307&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024805307&partnerID=8YFLogxK

U2 - 10.1118/1.3469080

DO - 10.1118/1.3469080

M3 - Article

AN - SCOPUS:85024805307

VL - 37

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -