MR thermometry in the human prostate gland at 3.0T for transurethral ultrasound therapy

Elizabeth Ramsay, Charles Mougenot, Max Köhler, Michael Bronskill, Laurence Klotz, Masoom A. Haider, Rajiv Chopra

Research output: Contribution to journalArticle

23 Citations (Scopus)

Abstract

Purpose: To investigate the spatial, temporal, and temperature resolution of a segmented gradient echo echo-planar imaging (EPI) technique as applied to proton resonance frequency (PRF) shift thermometry at 3 T in the human prostate gland, and to determine appropriate sequence parameters for magnetic resonance imaging (MRI)-controlled transurethral ultrasound thermal therapy. Materials and Methods: Eleven healthy volunteers (age range 23-58) were scanned at 3 T with a 16-channel torso coil to study the behavior of a gradient echo EPI thermometry sequence. The temperature stability and geometric distortion were assessed for 11 different parameter sets. In a further five volunteers, the prostate T2* was measured. Results: For all scan parameters investigated, the temperature standard deviation within the prostate was less than 1°C, while the distortion was less than 1 mm. Temperature stability was best with higher TE values (up to 25 msec), larger voxel sizes and lower EPI factors, but this had to be balanced against requirements for good spatial and temporal resolution. Prostate T2* values ranged from 30-50 msec. Conclusion: A good balance between temperature stability and temporal/spatial resolution is obtained with TE = 15 msec, voxel size = 1.14 mm, and EPI factor = 9, resulting in a dynamic scan time of 7.2 seconds for the nine slices.

Original languageEnglish (US)
Pages (from-to)1564-1571
Number of pages8
JournalJournal of Magnetic Resonance Imaging
Volume38
Issue number6
DOIs
StatePublished - Dec 2013

Fingerprint

Thermometry
Echo-Planar Imaging
Prostate
Temperature
Therapeutics
Torso
Protons
Volunteers
Healthy Volunteers
Hot Temperature
Magnetic Resonance Imaging

Keywords

  • interventional MRI
  • MR thermometry
  • proton resonance frequency shift
  • segmented EPI
  • thermal therapy

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

MR thermometry in the human prostate gland at 3.0T for transurethral ultrasound therapy. / Ramsay, Elizabeth; Mougenot, Charles; Köhler, Max; Bronskill, Michael; Klotz, Laurence; Haider, Masoom A.; Chopra, Rajiv.

In: Journal of Magnetic Resonance Imaging, Vol. 38, No. 6, 12.2013, p. 1564-1571.

Research output: Contribution to journalArticle

Ramsay, E, Mougenot, C, Köhler, M, Bronskill, M, Klotz, L, Haider, MA & Chopra, R 2013, 'MR thermometry in the human prostate gland at 3.0T for transurethral ultrasound therapy', Journal of Magnetic Resonance Imaging, vol. 38, no. 6, pp. 1564-1571. https://doi.org/10.1002/jmri.24063
Ramsay, Elizabeth ; Mougenot, Charles ; Köhler, Max ; Bronskill, Michael ; Klotz, Laurence ; Haider, Masoom A. ; Chopra, Rajiv. / MR thermometry in the human prostate gland at 3.0T for transurethral ultrasound therapy. In: Journal of Magnetic Resonance Imaging. 2013 ; Vol. 38, No. 6. pp. 1564-1571.
@article{67f8765e0e2b41afb682be09c02bd73c,
title = "MR thermometry in the human prostate gland at 3.0T for transurethral ultrasound therapy",
abstract = "Purpose: To investigate the spatial, temporal, and temperature resolution of a segmented gradient echo echo-planar imaging (EPI) technique as applied to proton resonance frequency (PRF) shift thermometry at 3 T in the human prostate gland, and to determine appropriate sequence parameters for magnetic resonance imaging (MRI)-controlled transurethral ultrasound thermal therapy. Materials and Methods: Eleven healthy volunteers (age range 23-58) were scanned at 3 T with a 16-channel torso coil to study the behavior of a gradient echo EPI thermometry sequence. The temperature stability and geometric distortion were assessed for 11 different parameter sets. In a further five volunteers, the prostate T2* was measured. Results: For all scan parameters investigated, the temperature standard deviation within the prostate was less than 1°C, while the distortion was less than 1 mm. Temperature stability was best with higher TE values (up to 25 msec), larger voxel sizes and lower EPI factors, but this had to be balanced against requirements for good spatial and temporal resolution. Prostate T2* values ranged from 30-50 msec. Conclusion: A good balance between temperature stability and temporal/spatial resolution is obtained with TE = 15 msec, voxel size = 1.14 mm, and EPI factor = 9, resulting in a dynamic scan time of 7.2 seconds for the nine slices.",
keywords = "interventional MRI, MR thermometry, proton resonance frequency shift, segmented EPI, thermal therapy",
author = "Elizabeth Ramsay and Charles Mougenot and Max K{\"o}hler and Michael Bronskill and Laurence Klotz and Haider, {Masoom A.} and Rajiv Chopra",
year = "2013",
month = "12",
doi = "10.1002/jmri.24063",
language = "English (US)",
volume = "38",
pages = "1564--1571",
journal = "Journal of Magnetic Resonance Imaging",
issn = "1053-1807",
publisher = "John Wiley and Sons Inc.",
number = "6",

}

TY - JOUR

T1 - MR thermometry in the human prostate gland at 3.0T for transurethral ultrasound therapy

AU - Ramsay, Elizabeth

AU - Mougenot, Charles

AU - Köhler, Max

AU - Bronskill, Michael

AU - Klotz, Laurence

AU - Haider, Masoom A.

AU - Chopra, Rajiv

PY - 2013/12

Y1 - 2013/12

N2 - Purpose: To investigate the spatial, temporal, and temperature resolution of a segmented gradient echo echo-planar imaging (EPI) technique as applied to proton resonance frequency (PRF) shift thermometry at 3 T in the human prostate gland, and to determine appropriate sequence parameters for magnetic resonance imaging (MRI)-controlled transurethral ultrasound thermal therapy. Materials and Methods: Eleven healthy volunteers (age range 23-58) were scanned at 3 T with a 16-channel torso coil to study the behavior of a gradient echo EPI thermometry sequence. The temperature stability and geometric distortion were assessed for 11 different parameter sets. In a further five volunteers, the prostate T2* was measured. Results: For all scan parameters investigated, the temperature standard deviation within the prostate was less than 1°C, while the distortion was less than 1 mm. Temperature stability was best with higher TE values (up to 25 msec), larger voxel sizes and lower EPI factors, but this had to be balanced against requirements for good spatial and temporal resolution. Prostate T2* values ranged from 30-50 msec. Conclusion: A good balance between temperature stability and temporal/spatial resolution is obtained with TE = 15 msec, voxel size = 1.14 mm, and EPI factor = 9, resulting in a dynamic scan time of 7.2 seconds for the nine slices.

AB - Purpose: To investigate the spatial, temporal, and temperature resolution of a segmented gradient echo echo-planar imaging (EPI) technique as applied to proton resonance frequency (PRF) shift thermometry at 3 T in the human prostate gland, and to determine appropriate sequence parameters for magnetic resonance imaging (MRI)-controlled transurethral ultrasound thermal therapy. Materials and Methods: Eleven healthy volunteers (age range 23-58) were scanned at 3 T with a 16-channel torso coil to study the behavior of a gradient echo EPI thermometry sequence. The temperature stability and geometric distortion were assessed for 11 different parameter sets. In a further five volunteers, the prostate T2* was measured. Results: For all scan parameters investigated, the temperature standard deviation within the prostate was less than 1°C, while the distortion was less than 1 mm. Temperature stability was best with higher TE values (up to 25 msec), larger voxel sizes and lower EPI factors, but this had to be balanced against requirements for good spatial and temporal resolution. Prostate T2* values ranged from 30-50 msec. Conclusion: A good balance between temperature stability and temporal/spatial resolution is obtained with TE = 15 msec, voxel size = 1.14 mm, and EPI factor = 9, resulting in a dynamic scan time of 7.2 seconds for the nine slices.

KW - interventional MRI

KW - MR thermometry

KW - proton resonance frequency shift

KW - segmented EPI

KW - thermal therapy

UR - http://www.scopus.com/inward/record.url?scp=84890014952&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84890014952&partnerID=8YFLogxK

U2 - 10.1002/jmri.24063

DO - 10.1002/jmri.24063

M3 - Article

C2 - 23440850

AN - SCOPUS:84890014952

VL - 38

SP - 1564

EP - 1571

JO - Journal of Magnetic Resonance Imaging

JF - Journal of Magnetic Resonance Imaging

SN - 1053-1807

IS - 6

ER -