Primary valve malfunction of the Krupin eye valve with disk

J. C. Burchfield, M. A. Kass, M. B. Wax

Research output: Contribution to journalArticle

Abstract

Purpose. To examine the management and possible causes of primary valve malfunction of the Krupin eye valve with disk. Methods. The authors reviewed the results of 113 patients undergoing implantation of the Krupin eye valve with disk and identified 8 patients with primary valve malfunction requiring surgical revison. Results. Valve revision involved manipulation (1 case), explantation of the malfunctioning valve and implantation of a new valve (2 cases), and amputation of the valve (5 cases). Six of eight patients had final intraocular pressures of less than 21 mm Hg on one or no medications at a mean interval of 15.9 months (range 5-36 months) after surgical revision. Transient postoperative hypotony was noted in 3 patients and chronic hypotony with loss of light perception in I patient. One explanted valve was examined and found to have partially fused leaflets. Conclusions. Surgical revision in cases of primary valve malfunction of the Krupin eye valve with disk may be accomplished relatively safely with an acceptable level of postoperative complicatons. The etiology of primary valve malfunction may be related to the sterilization process and prolonged storage prior to implantation.

Original languageEnglish (US)
JournalInvestigative Ophthalmology and Visual Science
Volume37
Issue number3
StatePublished - Feb 15 1996

Fingerprint

Glaucoma Drainage Implants
Reoperation
Intraocular Pressure
Amputation
Light

ASJC Scopus subject areas

  • Ophthalmology

Cite this

Primary valve malfunction of the Krupin eye valve with disk. / Burchfield, J. C.; Kass, M. A.; Wax, M. B.

In: Investigative Ophthalmology and Visual Science, Vol. 37, No. 3, 15.02.1996.

Research output: Contribution to journalArticle

Burchfield, J. C. ; Kass, M. A. ; Wax, M. B. / Primary valve malfunction of the Krupin eye valve with disk. In: Investigative Ophthalmology and Visual Science. 1996 ; Vol. 37, No. 3.
@article{196319aee3a54ed2a14ccd69613a185b,
title = "Primary valve malfunction of the Krupin eye valve with disk",
abstract = "Purpose. To examine the management and possible causes of primary valve malfunction of the Krupin eye valve with disk. Methods. The authors reviewed the results of 113 patients undergoing implantation of the Krupin eye valve with disk and identified 8 patients with primary valve malfunction requiring surgical revison. Results. Valve revision involved manipulation (1 case), explantation of the malfunctioning valve and implantation of a new valve (2 cases), and amputation of the valve (5 cases). Six of eight patients had final intraocular pressures of less than 21 mm Hg on one or no medications at a mean interval of 15.9 months (range 5-36 months) after surgical revision. Transient postoperative hypotony was noted in 3 patients and chronic hypotony with loss of light perception in I patient. One explanted valve was examined and found to have partially fused leaflets. Conclusions. Surgical revision in cases of primary valve malfunction of the Krupin eye valve with disk may be accomplished relatively safely with an acceptable level of postoperative complicatons. The etiology of primary valve malfunction may be related to the sterilization process and prolonged storage prior to implantation.",
author = "Burchfield, {J. C.} and Kass, {M. A.} and Wax, {M. B.}",
year = "1996",
month = "2",
day = "15",
language = "English (US)",
volume = "37",
journal = "Investigative Ophthalmology and Visual Science",
issn = "0146-0404",
publisher = "Association for Research in Vision and Ophthalmology Inc.",
number = "3",

}

TY - JOUR

T1 - Primary valve malfunction of the Krupin eye valve with disk

AU - Burchfield, J. C.

AU - Kass, M. A.

AU - Wax, M. B.

PY - 1996/2/15

Y1 - 1996/2/15

N2 - Purpose. To examine the management and possible causes of primary valve malfunction of the Krupin eye valve with disk. Methods. The authors reviewed the results of 113 patients undergoing implantation of the Krupin eye valve with disk and identified 8 patients with primary valve malfunction requiring surgical revison. Results. Valve revision involved manipulation (1 case), explantation of the malfunctioning valve and implantation of a new valve (2 cases), and amputation of the valve (5 cases). Six of eight patients had final intraocular pressures of less than 21 mm Hg on one or no medications at a mean interval of 15.9 months (range 5-36 months) after surgical revision. Transient postoperative hypotony was noted in 3 patients and chronic hypotony with loss of light perception in I patient. One explanted valve was examined and found to have partially fused leaflets. Conclusions. Surgical revision in cases of primary valve malfunction of the Krupin eye valve with disk may be accomplished relatively safely with an acceptable level of postoperative complicatons. The etiology of primary valve malfunction may be related to the sterilization process and prolonged storage prior to implantation.

AB - Purpose. To examine the management and possible causes of primary valve malfunction of the Krupin eye valve with disk. Methods. The authors reviewed the results of 113 patients undergoing implantation of the Krupin eye valve with disk and identified 8 patients with primary valve malfunction requiring surgical revison. Results. Valve revision involved manipulation (1 case), explantation of the malfunctioning valve and implantation of a new valve (2 cases), and amputation of the valve (5 cases). Six of eight patients had final intraocular pressures of less than 21 mm Hg on one or no medications at a mean interval of 15.9 months (range 5-36 months) after surgical revision. Transient postoperative hypotony was noted in 3 patients and chronic hypotony with loss of light perception in I patient. One explanted valve was examined and found to have partially fused leaflets. Conclusions. Surgical revision in cases of primary valve malfunction of the Krupin eye valve with disk may be accomplished relatively safely with an acceptable level of postoperative complicatons. The etiology of primary valve malfunction may be related to the sterilization process and prolonged storage prior to implantation.

UR - http://www.scopus.com/inward/record.url?scp=33750168697&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=33750168697&partnerID=8YFLogxK

M3 - Article

AN - SCOPUS:33750168697

VL - 37

JO - Investigative Ophthalmology and Visual Science

JF - Investigative Ophthalmology and Visual Science

SN - 0146-0404

IS - 3

ER -