Static jaw collimation settings to minimize radiation dose to normal brain tissue during stereotactic radiosurgery

Eun Young Han, Xin Zhang, Yulong Yan, Sunil Sharma, Jose Penagaricano, Eduardo Moros, Peter Corry

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

At the University of Arkansas for Medical Sciences (UAMS) intracranial stereotactic radiosurgery (SRS) is performed by using a linear accelerator with an add-on micromultileaf collimator (mMLC). In our clinical setting, static jaws are automatically adapted to the furthest edge of the mMLC-defined segments with 2-mm (X jaw) and 5-mm (Y jaw) margin and the same jaw values are applied for all beam angles in the treatment planning system. This additional field gap between the static jaws and the mMLC allows additional radiation dose to normal brain tissue. Because a radiosurgery procedure consists of a single high dose to the planning target volume (PTV), reduction of unnecessary dose to normal brain tissue near the PTV is important, particularly for pediatric patients whose brains are still developing or when a critical organ, such as the optic chiasm, is near the PTV. The purpose of this study was to minimize dose to normal brain tissue by allowing minimal static jaw margin around the mMLC-defined fields and different static jaw values for each beam angle or arc. Dose output factors were measured with various static jaw margins and the results were compared with calculated doses in the treatment planning system. Ten patient plans were randomly selected and recalculated with zero static jaw margins without changing other parameters. Changes of PTV coverage, mean dose to predefined normal brain tissue volume adjacent to PTV, and monitor units were compared. It was found that the dose output percentage difference varied from 4.9-1.3% for the maximum static jaw opening vs. static jaw with zero margins. The mean dose to normal brain tissue at risk adjacent to the PTV was reduced by an average of 1.9%, with negligible PTV coverage loss. This dose reduction strategy may be meaningful in terms of late effects of radiation, particularly in pediatric patients. This study generated clinical knowledge and tools to consistently minimize dose to normal brain tissue.

Original languageEnglish (US)
Pages (from-to)391-395
Number of pages5
JournalMedical Dosimetry
Volume37
Issue number4
DOIs
StatePublished - 2012

Fingerprint

Radiosurgery
Jaw
Radiation
Brain
Pediatrics
Optic Chiasm
Particle Accelerators
Radiation Effects

Keywords

  • Output factor
  • Static jaw
  • Stereotactic radiosurgery

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Static jaw collimation settings to minimize radiation dose to normal brain tissue during stereotactic radiosurgery. / Han, Eun Young; Zhang, Xin; Yan, Yulong; Sharma, Sunil; Penagaricano, Jose; Moros, Eduardo; Corry, Peter.

In: Medical Dosimetry, Vol. 37, No. 4, 2012, p. 391-395.

Research output: Contribution to journalArticle

Han, Eun Young ; Zhang, Xin ; Yan, Yulong ; Sharma, Sunil ; Penagaricano, Jose ; Moros, Eduardo ; Corry, Peter. / Static jaw collimation settings to minimize radiation dose to normal brain tissue during stereotactic radiosurgery. In: Medical Dosimetry. 2012 ; Vol. 37, No. 4. pp. 391-395.
@article{4108c1efccd740948ffae751b6a39a24,
title = "Static jaw collimation settings to minimize radiation dose to normal brain tissue during stereotactic radiosurgery",
abstract = "At the University of Arkansas for Medical Sciences (UAMS) intracranial stereotactic radiosurgery (SRS) is performed by using a linear accelerator with an add-on micromultileaf collimator (mMLC). In our clinical setting, static jaws are automatically adapted to the furthest edge of the mMLC-defined segments with 2-mm (X jaw) and 5-mm (Y jaw) margin and the same jaw values are applied for all beam angles in the treatment planning system. This additional field gap between the static jaws and the mMLC allows additional radiation dose to normal brain tissue. Because a radiosurgery procedure consists of a single high dose to the planning target volume (PTV), reduction of unnecessary dose to normal brain tissue near the PTV is important, particularly for pediatric patients whose brains are still developing or when a critical organ, such as the optic chiasm, is near the PTV. The purpose of this study was to minimize dose to normal brain tissue by allowing minimal static jaw margin around the mMLC-defined fields and different static jaw values for each beam angle or arc. Dose output factors were measured with various static jaw margins and the results were compared with calculated doses in the treatment planning system. Ten patient plans were randomly selected and recalculated with zero static jaw margins without changing other parameters. Changes of PTV coverage, mean dose to predefined normal brain tissue volume adjacent to PTV, and monitor units were compared. It was found that the dose output percentage difference varied from 4.9-1.3{\%} for the maximum static jaw opening vs. static jaw with zero margins. The mean dose to normal brain tissue at risk adjacent to the PTV was reduced by an average of 1.9{\%}, with negligible PTV coverage loss. This dose reduction strategy may be meaningful in terms of late effects of radiation, particularly in pediatric patients. This study generated clinical knowledge and tools to consistently minimize dose to normal brain tissue.",
keywords = "Output factor, Static jaw, Stereotactic radiosurgery",
author = "Han, {Eun Young} and Xin Zhang and Yulong Yan and Sunil Sharma and Jose Penagaricano and Eduardo Moros and Peter Corry",
year = "2012",
doi = "10.1016/j.meddos.2012.02.005",
language = "English (US)",
volume = "37",
pages = "391--395",
journal = "Medical Dosimetry",
issn = "0958-3947",
publisher = "Elsevier Inc.",
number = "4",

}

TY - JOUR

T1 - Static jaw collimation settings to minimize radiation dose to normal brain tissue during stereotactic radiosurgery

AU - Han, Eun Young

AU - Zhang, Xin

AU - Yan, Yulong

AU - Sharma, Sunil

AU - Penagaricano, Jose

AU - Moros, Eduardo

AU - Corry, Peter

PY - 2012

Y1 - 2012

N2 - At the University of Arkansas for Medical Sciences (UAMS) intracranial stereotactic radiosurgery (SRS) is performed by using a linear accelerator with an add-on micromultileaf collimator (mMLC). In our clinical setting, static jaws are automatically adapted to the furthest edge of the mMLC-defined segments with 2-mm (X jaw) and 5-mm (Y jaw) margin and the same jaw values are applied for all beam angles in the treatment planning system. This additional field gap between the static jaws and the mMLC allows additional radiation dose to normal brain tissue. Because a radiosurgery procedure consists of a single high dose to the planning target volume (PTV), reduction of unnecessary dose to normal brain tissue near the PTV is important, particularly for pediatric patients whose brains are still developing or when a critical organ, such as the optic chiasm, is near the PTV. The purpose of this study was to minimize dose to normal brain tissue by allowing minimal static jaw margin around the mMLC-defined fields and different static jaw values for each beam angle or arc. Dose output factors were measured with various static jaw margins and the results were compared with calculated doses in the treatment planning system. Ten patient plans were randomly selected and recalculated with zero static jaw margins without changing other parameters. Changes of PTV coverage, mean dose to predefined normal brain tissue volume adjacent to PTV, and monitor units were compared. It was found that the dose output percentage difference varied from 4.9-1.3% for the maximum static jaw opening vs. static jaw with zero margins. The mean dose to normal brain tissue at risk adjacent to the PTV was reduced by an average of 1.9%, with negligible PTV coverage loss. This dose reduction strategy may be meaningful in terms of late effects of radiation, particularly in pediatric patients. This study generated clinical knowledge and tools to consistently minimize dose to normal brain tissue.

AB - At the University of Arkansas for Medical Sciences (UAMS) intracranial stereotactic radiosurgery (SRS) is performed by using a linear accelerator with an add-on micromultileaf collimator (mMLC). In our clinical setting, static jaws are automatically adapted to the furthest edge of the mMLC-defined segments with 2-mm (X jaw) and 5-mm (Y jaw) margin and the same jaw values are applied for all beam angles in the treatment planning system. This additional field gap between the static jaws and the mMLC allows additional radiation dose to normal brain tissue. Because a radiosurgery procedure consists of a single high dose to the planning target volume (PTV), reduction of unnecessary dose to normal brain tissue near the PTV is important, particularly for pediatric patients whose brains are still developing or when a critical organ, such as the optic chiasm, is near the PTV. The purpose of this study was to minimize dose to normal brain tissue by allowing minimal static jaw margin around the mMLC-defined fields and different static jaw values for each beam angle or arc. Dose output factors were measured with various static jaw margins and the results were compared with calculated doses in the treatment planning system. Ten patient plans were randomly selected and recalculated with zero static jaw margins without changing other parameters. Changes of PTV coverage, mean dose to predefined normal brain tissue volume adjacent to PTV, and monitor units were compared. It was found that the dose output percentage difference varied from 4.9-1.3% for the maximum static jaw opening vs. static jaw with zero margins. The mean dose to normal brain tissue at risk adjacent to the PTV was reduced by an average of 1.9%, with negligible PTV coverage loss. This dose reduction strategy may be meaningful in terms of late effects of radiation, particularly in pediatric patients. This study generated clinical knowledge and tools to consistently minimize dose to normal brain tissue.

KW - Output factor

KW - Static jaw

KW - Stereotactic radiosurgery

UR - http://www.scopus.com/inward/record.url?scp=84868207399&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84868207399&partnerID=8YFLogxK

U2 - 10.1016/j.meddos.2012.02.005

DO - 10.1016/j.meddos.2012.02.005

M3 - Article

C2 - 22552119

AN - SCOPUS:84868207399

VL - 37

SP - 391

EP - 395

JO - Medical Dosimetry

JF - Medical Dosimetry

SN - 0958-3947

IS - 4

ER -