SU‐E‐J‐16: Noise Reduction with Detail Preservation for Low‐Dose KV CBCT Using Non‐Local Means: Simulated Patient Study

W. lu, W. Yao, J. Wang, D. Yang

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

Purpose: With the frequent uses of CBCT in IGRT, the cumulative imaging dose to normal tissues may not be insignificant. A lower mAs protocol in CBCT acquisition reduces the dose, but dramatically degrades image quality due to excessive noise. The purpose of this study is to examine the effectiveness of the nonlocal means (NL‐means) denoising algorithm in reducing noise while preserving details in simulated low‐dose patient CBCT. Method and Materials:NL‐means algorithm estimates the true value of a pixel as a weighted average of all pixels in the image, where the weights depend on the similarity between the pixels. Compared with the local smoothing or filtering methods, NL‐means can reduce noise while preserve details. Low‐dose patient CBCT images were generated by adding noise to the normal‐dose images based on stochastic property of incident photons from the x‐ray source, simulating data acquisition with reduced mAs. The low‐dose images were normalized to be within [0 1] and then fed to the NL‐means for denoising. The optimal parameters for NL‐means were experimentally determined. Results: The simulated low‐dose CBCT had only 6.3% of the patient CBCT dose. NL‐means clearly reduced the noise without obvious blurring, and the images appear to have similar quality as normal‐dose images. The suppressed noise resembled the desired white noise except at sharp edges. The mean signal‐to‐noise ratio in homogeneous regions was increased from 7.6 to 28.1. NL‐means preserved anatomic details as fine as 2 pixels but blurred single‐pixel details. Conclusion: Results on the patient data demonstrated that NL‐means effectively reduced the noise while preserving most fine details in simulated low‐dose patient CBCT, consistent with results on phantom study. A single set of optimal parameters was suitable for various CBCTs. This post‐processing method can be straightforwardly implemented in OBI software.

Original languageEnglish (US)
Number of pages1
JournalMedical Physics
Volume38
Issue number6
DOIs
StatePublished - 2011

Fingerprint

Noise
Information Storage and Retrieval
Photons
Software
X-Rays
Weights and Measures

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

SU‐E‐J‐16 : Noise Reduction with Detail Preservation for Low‐Dose KV CBCT Using Non‐Local Means: Simulated Patient Study. / lu, W.; Yao, W.; Wang, J.; Yang, D.

In: Medical Physics, Vol. 38, No. 6, 2011.

Research output: Contribution to journalArticle

@article{9621e8e81c5342e389eefe2c87c66c82,
title = "SU‐E‐J‐16: Noise Reduction with Detail Preservation for Low‐Dose KV CBCT Using Non‐Local Means: Simulated Patient Study",
abstract = "Purpose: With the frequent uses of CBCT in IGRT, the cumulative imaging dose to normal tissues may not be insignificant. A lower mAs protocol in CBCT acquisition reduces the dose, but dramatically degrades image quality due to excessive noise. The purpose of this study is to examine the effectiveness of the nonlocal means (NL‐means) denoising algorithm in reducing noise while preserving details in simulated low‐dose patient CBCT. Method and Materials:NL‐means algorithm estimates the true value of a pixel as a weighted average of all pixels in the image, where the weights depend on the similarity between the pixels. Compared with the local smoothing or filtering methods, NL‐means can reduce noise while preserve details. Low‐dose patient CBCT images were generated by adding noise to the normal‐dose images based on stochastic property of incident photons from the x‐ray source, simulating data acquisition with reduced mAs. The low‐dose images were normalized to be within [0 1] and then fed to the NL‐means for denoising. The optimal parameters for NL‐means were experimentally determined. Results: The simulated low‐dose CBCT had only 6.3{\%} of the patient CBCT dose. NL‐means clearly reduced the noise without obvious blurring, and the images appear to have similar quality as normal‐dose images. The suppressed noise resembled the desired white noise except at sharp edges. The mean signal‐to‐noise ratio in homogeneous regions was increased from 7.6 to 28.1. NL‐means preserved anatomic details as fine as 2 pixels but blurred single‐pixel details. Conclusion: Results on the patient data demonstrated that NL‐means effectively reduced the noise while preserving most fine details in simulated low‐dose patient CBCT, consistent with results on phantom study. A single set of optimal parameters was suitable for various CBCTs. This post‐processing method can be straightforwardly implemented in OBI software.",
author = "W. lu and W. Yao and J. Wang and D. Yang",
year = "2011",
doi = "10.1118/1.3611784",
language = "English (US)",
volume = "38",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - SU‐E‐J‐16

T2 - Noise Reduction with Detail Preservation for Low‐Dose KV CBCT Using Non‐Local Means: Simulated Patient Study

AU - lu, W.

AU - Yao, W.

AU - Wang, J.

AU - Yang, D.

PY - 2011

Y1 - 2011

N2 - Purpose: With the frequent uses of CBCT in IGRT, the cumulative imaging dose to normal tissues may not be insignificant. A lower mAs protocol in CBCT acquisition reduces the dose, but dramatically degrades image quality due to excessive noise. The purpose of this study is to examine the effectiveness of the nonlocal means (NL‐means) denoising algorithm in reducing noise while preserving details in simulated low‐dose patient CBCT. Method and Materials:NL‐means algorithm estimates the true value of a pixel as a weighted average of all pixels in the image, where the weights depend on the similarity between the pixels. Compared with the local smoothing or filtering methods, NL‐means can reduce noise while preserve details. Low‐dose patient CBCT images were generated by adding noise to the normal‐dose images based on stochastic property of incident photons from the x‐ray source, simulating data acquisition with reduced mAs. The low‐dose images were normalized to be within [0 1] and then fed to the NL‐means for denoising. The optimal parameters for NL‐means were experimentally determined. Results: The simulated low‐dose CBCT had only 6.3% of the patient CBCT dose. NL‐means clearly reduced the noise without obvious blurring, and the images appear to have similar quality as normal‐dose images. The suppressed noise resembled the desired white noise except at sharp edges. The mean signal‐to‐noise ratio in homogeneous regions was increased from 7.6 to 28.1. NL‐means preserved anatomic details as fine as 2 pixels but blurred single‐pixel details. Conclusion: Results on the patient data demonstrated that NL‐means effectively reduced the noise while preserving most fine details in simulated low‐dose patient CBCT, consistent with results on phantom study. A single set of optimal parameters was suitable for various CBCTs. This post‐processing method can be straightforwardly implemented in OBI software.

AB - Purpose: With the frequent uses of CBCT in IGRT, the cumulative imaging dose to normal tissues may not be insignificant. A lower mAs protocol in CBCT acquisition reduces the dose, but dramatically degrades image quality due to excessive noise. The purpose of this study is to examine the effectiveness of the nonlocal means (NL‐means) denoising algorithm in reducing noise while preserving details in simulated low‐dose patient CBCT. Method and Materials:NL‐means algorithm estimates the true value of a pixel as a weighted average of all pixels in the image, where the weights depend on the similarity between the pixels. Compared with the local smoothing or filtering methods, NL‐means can reduce noise while preserve details. Low‐dose patient CBCT images were generated by adding noise to the normal‐dose images based on stochastic property of incident photons from the x‐ray source, simulating data acquisition with reduced mAs. The low‐dose images were normalized to be within [0 1] and then fed to the NL‐means for denoising. The optimal parameters for NL‐means were experimentally determined. Results: The simulated low‐dose CBCT had only 6.3% of the patient CBCT dose. NL‐means clearly reduced the noise without obvious blurring, and the images appear to have similar quality as normal‐dose images. The suppressed noise resembled the desired white noise except at sharp edges. The mean signal‐to‐noise ratio in homogeneous regions was increased from 7.6 to 28.1. NL‐means preserved anatomic details as fine as 2 pixels but blurred single‐pixel details. Conclusion: Results on the patient data demonstrated that NL‐means effectively reduced the noise while preserving most fine details in simulated low‐dose patient CBCT, consistent with results on phantom study. A single set of optimal parameters was suitable for various CBCTs. This post‐processing method can be straightforwardly implemented in OBI software.

UR - http://www.scopus.com/inward/record.url?scp=84866098323&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84866098323&partnerID=8YFLogxK

U2 - 10.1118/1.3611784

DO - 10.1118/1.3611784

M3 - Article

AN - SCOPUS:84866098323

VL - 38

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -