SU‐E‐T‐301: A Novel Daily QA Device for Proton Therapy

X. Ding, Y. Zheng, A. Mascia, W. Hsi, Y. Kang, E. Ramirez, O. Zeidan, R. Foster, M. Gao, S. Laub, M. Pankuch, N. Schreuder, B. Harris

Research output: Contribution to journalArticle

Abstract

Purpose: We describe the design and use of a daily QA device for proton therapy. The device is designed for therapists to check the readiness of the IBA Proton Therapy System (IBA, Louvain‐la‐Neuve, Belgium) during morning QA. The checks include connectivity, positioning, mechanical, imaging and dosimetric parameters of the proton therapy system. Methods: The device consists of a commercial QA device, (rf‐DailyQA3 ‐Sun Nuclear Corporation, Melbourne, FL), in conjunction with a home‐made acrylic phantom and mechanical indexing jig. The indexing jig indexes the rf‐DailyQA3 to treatment couch. Fiducial markers embedded in the phantom are used for checking the x‐ray image and alignment accuracy of the imaging system (VeriSuite, MedCom, Darmstadt‐ Germany). The rf‐ DailyQA3 is used to check the proton beam output, range and symmetry, which are acquired during one single beam delivery of 100 monitor units. We developed in‐house software to calculate the variation of beam range and symmetry, based on readings from the various ion chambers inside the rf‐DailyQA3. Results: The device has been employed to perform daily QA since June 2010 at two operational proton treatment centers and will soon be implemented at ProCure's New Jersey center. All QA tests are performed by radiation therapists and reviewed by the medical physicist on duty. Due to the simplicity of the device and the associated processing software, the QA time is less than 20 minutes per room. The measurement data collected by the device during daily QA are recorded in the OIS. The integrity of the data is validated by comparing against other independent measurements. Conclusions: The daily QA device has been proven to be robust, reliable and user‐friendly. The performance of this system has been proven to be stable and accurate using trend analyses. Key words:proton therapy, daily QA, output, range, symmetry.

Original languageEnglish (US)
Pages (from-to)3772-3773
Number of pages2
JournalMedical Physics
Volume39
Issue number6
DOIs
StatePublished - 2012

Fingerprint

Proton Therapy
Equipment and Supplies
Protons
Software
Fiducial Markers
Belgium
Solar System
Germany
Reading
X-Rays
Ions
Radiation

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

Ding, X., Zheng, Y., Mascia, A., Hsi, W., Kang, Y., Ramirez, E., ... Harris, B. (2012). SU‐E‐T‐301: A Novel Daily QA Device for Proton Therapy. Medical Physics, 39(6), 3772-3773. https://doi.org/10.1118/1.4735387

SU‐E‐T‐301 : A Novel Daily QA Device for Proton Therapy. / Ding, X.; Zheng, Y.; Mascia, A.; Hsi, W.; Kang, Y.; Ramirez, E.; Zeidan, O.; Foster, R.; Gao, M.; Laub, S.; Pankuch, M.; Schreuder, N.; Harris, B.

In: Medical Physics, Vol. 39, No. 6, 2012, p. 3772-3773.

Research output: Contribution to journalArticle

Ding, X, Zheng, Y, Mascia, A, Hsi, W, Kang, Y, Ramirez, E, Zeidan, O, Foster, R, Gao, M, Laub, S, Pankuch, M, Schreuder, N & Harris, B 2012, 'SU‐E‐T‐301: A Novel Daily QA Device for Proton Therapy', Medical Physics, vol. 39, no. 6, pp. 3772-3773. https://doi.org/10.1118/1.4735387
Ding X, Zheng Y, Mascia A, Hsi W, Kang Y, Ramirez E et al. SU‐E‐T‐301: A Novel Daily QA Device for Proton Therapy. Medical Physics. 2012;39(6):3772-3773. https://doi.org/10.1118/1.4735387
Ding, X. ; Zheng, Y. ; Mascia, A. ; Hsi, W. ; Kang, Y. ; Ramirez, E. ; Zeidan, O. ; Foster, R. ; Gao, M. ; Laub, S. ; Pankuch, M. ; Schreuder, N. ; Harris, B. / SU‐E‐T‐301 : A Novel Daily QA Device for Proton Therapy. In: Medical Physics. 2012 ; Vol. 39, No. 6. pp. 3772-3773.
@article{04e91a1b025b4edca55a07c92beba12e,
title = "SU‐E‐T‐301: A Novel Daily QA Device for Proton Therapy",
abstract = "Purpose: We describe the design and use of a daily QA device for proton therapy. The device is designed for therapists to check the readiness of the IBA Proton Therapy System (IBA, Louvain‐la‐Neuve, Belgium) during morning QA. The checks include connectivity, positioning, mechanical, imaging and dosimetric parameters of the proton therapy system. Methods: The device consists of a commercial QA device, (rf‐DailyQA3 ‐Sun Nuclear Corporation, Melbourne, FL), in conjunction with a home‐made acrylic phantom and mechanical indexing jig. The indexing jig indexes the rf‐DailyQA3 to treatment couch. Fiducial markers embedded in the phantom are used for checking the x‐ray image and alignment accuracy of the imaging system (VeriSuite, MedCom, Darmstadt‐ Germany). The rf‐ DailyQA3 is used to check the proton beam output, range and symmetry, which are acquired during one single beam delivery of 100 monitor units. We developed in‐house software to calculate the variation of beam range and symmetry, based on readings from the various ion chambers inside the rf‐DailyQA3. Results: The device has been employed to perform daily QA since June 2010 at two operational proton treatment centers and will soon be implemented at ProCure's New Jersey center. All QA tests are performed by radiation therapists and reviewed by the medical physicist on duty. Due to the simplicity of the device and the associated processing software, the QA time is less than 20 minutes per room. The measurement data collected by the device during daily QA are recorded in the OIS. The integrity of the data is validated by comparing against other independent measurements. Conclusions: The daily QA device has been proven to be robust, reliable and user‐friendly. The performance of this system has been proven to be stable and accurate using trend analyses. Key words:proton therapy, daily QA, output, range, symmetry.",
author = "X. Ding and Y. Zheng and A. Mascia and W. Hsi and Y. Kang and E. Ramirez and O. Zeidan and R. Foster and M. Gao and S. Laub and M. Pankuch and N. Schreuder and B. Harris",
year = "2012",
doi = "10.1118/1.4735387",
language = "English (US)",
volume = "39",
pages = "3772--3773",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - SU‐E‐T‐301

T2 - A Novel Daily QA Device for Proton Therapy

AU - Ding, X.

AU - Zheng, Y.

AU - Mascia, A.

AU - Hsi, W.

AU - Kang, Y.

AU - Ramirez, E.

AU - Zeidan, O.

AU - Foster, R.

AU - Gao, M.

AU - Laub, S.

AU - Pankuch, M.

AU - Schreuder, N.

AU - Harris, B.

PY - 2012

Y1 - 2012

N2 - Purpose: We describe the design and use of a daily QA device for proton therapy. The device is designed for therapists to check the readiness of the IBA Proton Therapy System (IBA, Louvain‐la‐Neuve, Belgium) during morning QA. The checks include connectivity, positioning, mechanical, imaging and dosimetric parameters of the proton therapy system. Methods: The device consists of a commercial QA device, (rf‐DailyQA3 ‐Sun Nuclear Corporation, Melbourne, FL), in conjunction with a home‐made acrylic phantom and mechanical indexing jig. The indexing jig indexes the rf‐DailyQA3 to treatment couch. Fiducial markers embedded in the phantom are used for checking the x‐ray image and alignment accuracy of the imaging system (VeriSuite, MedCom, Darmstadt‐ Germany). The rf‐ DailyQA3 is used to check the proton beam output, range and symmetry, which are acquired during one single beam delivery of 100 monitor units. We developed in‐house software to calculate the variation of beam range and symmetry, based on readings from the various ion chambers inside the rf‐DailyQA3. Results: The device has been employed to perform daily QA since June 2010 at two operational proton treatment centers and will soon be implemented at ProCure's New Jersey center. All QA tests are performed by radiation therapists and reviewed by the medical physicist on duty. Due to the simplicity of the device and the associated processing software, the QA time is less than 20 minutes per room. The measurement data collected by the device during daily QA are recorded in the OIS. The integrity of the data is validated by comparing against other independent measurements. Conclusions: The daily QA device has been proven to be robust, reliable and user‐friendly. The performance of this system has been proven to be stable and accurate using trend analyses. Key words:proton therapy, daily QA, output, range, symmetry.

AB - Purpose: We describe the design and use of a daily QA device for proton therapy. The device is designed for therapists to check the readiness of the IBA Proton Therapy System (IBA, Louvain‐la‐Neuve, Belgium) during morning QA. The checks include connectivity, positioning, mechanical, imaging and dosimetric parameters of the proton therapy system. Methods: The device consists of a commercial QA device, (rf‐DailyQA3 ‐Sun Nuclear Corporation, Melbourne, FL), in conjunction with a home‐made acrylic phantom and mechanical indexing jig. The indexing jig indexes the rf‐DailyQA3 to treatment couch. Fiducial markers embedded in the phantom are used for checking the x‐ray image and alignment accuracy of the imaging system (VeriSuite, MedCom, Darmstadt‐ Germany). The rf‐ DailyQA3 is used to check the proton beam output, range and symmetry, which are acquired during one single beam delivery of 100 monitor units. We developed in‐house software to calculate the variation of beam range and symmetry, based on readings from the various ion chambers inside the rf‐DailyQA3. Results: The device has been employed to perform daily QA since June 2010 at two operational proton treatment centers and will soon be implemented at ProCure's New Jersey center. All QA tests are performed by radiation therapists and reviewed by the medical physicist on duty. Due to the simplicity of the device and the associated processing software, the QA time is less than 20 minutes per room. The measurement data collected by the device during daily QA are recorded in the OIS. The integrity of the data is validated by comparing against other independent measurements. Conclusions: The daily QA device has been proven to be robust, reliable and user‐friendly. The performance of this system has been proven to be stable and accurate using trend analyses. Key words:proton therapy, daily QA, output, range, symmetry.

UR - http://www.scopus.com/inward/record.url?scp=85024821105&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024821105&partnerID=8YFLogxK

U2 - 10.1118/1.4735387

DO - 10.1118/1.4735387

M3 - Article

C2 - 28517302

AN - SCOPUS:85024821105

VL - 39

SP - 3772

EP - 3773

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -