Temporal performance of amorphous selenium mammography detectors

Bo Zhao, Wei Zhao

Research output: Contribution to journalArticle

22 Citations (Scopus)

Abstract

We investigated temporal performance of amorphous selenium (a-Se) detectors specifically designed for mammographic imaging. Our goal is to quantify the inherent lag and ghosting of a-Se photoconductor as a function of imaging conditions. Two small area electroded a-Se samples, one positively and the other negatively biased on the entrance side of x rays, were used in the experiments. The study of lag and ghosting was performed by delivering a number of raw exposures as experienced in screening mammography to the samples at different electric field strength ESe while measuring the current through the a-Se sample. Ghosting at different operational conditions was quantified as the percentage x-ray sensitivity (x-ray generated photocurrent measured from the sample) reduction compared to before irradiation. Lag was determined by measuring the residual current of a-Se at a given time after the end of each x-ray exposure. Both lag and ghosting were measured as a function of E Se and cumulative exposure. The values of ESe used in our experiments ranged from 1 to 20 V/μm. It was found that ghosting increases with exposure and decreases with ESe for both samples because of the dominant effect of recombination between trapped electrons and x-ray generated holes. Lag on the other hand has different dependence on ESe and cumulative exposure. At ESe ≤ 10 V/μm, the first frame lag for both samples changed slowly with cumulative exposure, with a range of 0.2%-1.7% for the positively biased sample and 0.5%-8% for the negatively biased sample. Overall the positively biased sample has better temporal performance than the negatively biased sample due to the lower density of trapped electrons. The impact of time interval between exposures on the temporal performance was also investigated. Recovery of ghosting with longer time interval was observed, which was attributed to the neutralization of trapped electrons by injected holes through dark current.

Original languageEnglish (US)
Pages (from-to)128-136
Number of pages9
JournalMedical Physics
Volume32
Issue number1
DOIs
StatePublished - Jan 2005

Fingerprint

Mammography
Selenium
X-Rays
Electrons
Genetic Recombination

Keywords

  • Amorphous selenium
  • Bulk trapping
  • Ghosting
  • Lag
  • Temporal performance
  • Trapped electrons

ASJC Scopus subject areas

  • Biophysics

Cite this

Temporal performance of amorphous selenium mammography detectors. / Zhao, Bo; Zhao, Wei.

In: Medical Physics, Vol. 32, No. 1, 01.2005, p. 128-136.

Research output: Contribution to journalArticle

Zhao, Bo ; Zhao, Wei. / Temporal performance of amorphous selenium mammography detectors. In: Medical Physics. 2005 ; Vol. 32, No. 1. pp. 128-136.
@article{cc5407c8d13e44bda8b0301578af32ca,
title = "Temporal performance of amorphous selenium mammography detectors",
abstract = "We investigated temporal performance of amorphous selenium (a-Se) detectors specifically designed for mammographic imaging. Our goal is to quantify the inherent lag and ghosting of a-Se photoconductor as a function of imaging conditions. Two small area electroded a-Se samples, one positively and the other negatively biased on the entrance side of x rays, were used in the experiments. The study of lag and ghosting was performed by delivering a number of raw exposures as experienced in screening mammography to the samples at different electric field strength ESe while measuring the current through the a-Se sample. Ghosting at different operational conditions was quantified as the percentage x-ray sensitivity (x-ray generated photocurrent measured from the sample) reduction compared to before irradiation. Lag was determined by measuring the residual current of a-Se at a given time after the end of each x-ray exposure. Both lag and ghosting were measured as a function of E Se and cumulative exposure. The values of ESe used in our experiments ranged from 1 to 20 V/μm. It was found that ghosting increases with exposure and decreases with ESe for both samples because of the dominant effect of recombination between trapped electrons and x-ray generated holes. Lag on the other hand has different dependence on ESe and cumulative exposure. At ESe ≤ 10 V/μm, the first frame lag for both samples changed slowly with cumulative exposure, with a range of 0.2{\%}-1.7{\%} for the positively biased sample and 0.5{\%}-8{\%} for the negatively biased sample. Overall the positively biased sample has better temporal performance than the negatively biased sample due to the lower density of trapped electrons. The impact of time interval between exposures on the temporal performance was also investigated. Recovery of ghosting with longer time interval was observed, which was attributed to the neutralization of trapped electrons by injected holes through dark current.",
keywords = "Amorphous selenium, Bulk trapping, Ghosting, Lag, Temporal performance, Trapped electrons",
author = "Bo Zhao and Wei Zhao",
year = "2005",
month = "1",
doi = "10.1118/1.1827791",
language = "English (US)",
volume = "32",
pages = "128--136",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "1",

}

TY - JOUR

T1 - Temporal performance of amorphous selenium mammography detectors

AU - Zhao, Bo

AU - Zhao, Wei

PY - 2005/1

Y1 - 2005/1

N2 - We investigated temporal performance of amorphous selenium (a-Se) detectors specifically designed for mammographic imaging. Our goal is to quantify the inherent lag and ghosting of a-Se photoconductor as a function of imaging conditions. Two small area electroded a-Se samples, one positively and the other negatively biased on the entrance side of x rays, were used in the experiments. The study of lag and ghosting was performed by delivering a number of raw exposures as experienced in screening mammography to the samples at different electric field strength ESe while measuring the current through the a-Se sample. Ghosting at different operational conditions was quantified as the percentage x-ray sensitivity (x-ray generated photocurrent measured from the sample) reduction compared to before irradiation. Lag was determined by measuring the residual current of a-Se at a given time after the end of each x-ray exposure. Both lag and ghosting were measured as a function of E Se and cumulative exposure. The values of ESe used in our experiments ranged from 1 to 20 V/μm. It was found that ghosting increases with exposure and decreases with ESe for both samples because of the dominant effect of recombination between trapped electrons and x-ray generated holes. Lag on the other hand has different dependence on ESe and cumulative exposure. At ESe ≤ 10 V/μm, the first frame lag for both samples changed slowly with cumulative exposure, with a range of 0.2%-1.7% for the positively biased sample and 0.5%-8% for the negatively biased sample. Overall the positively biased sample has better temporal performance than the negatively biased sample due to the lower density of trapped electrons. The impact of time interval between exposures on the temporal performance was also investigated. Recovery of ghosting with longer time interval was observed, which was attributed to the neutralization of trapped electrons by injected holes through dark current.

AB - We investigated temporal performance of amorphous selenium (a-Se) detectors specifically designed for mammographic imaging. Our goal is to quantify the inherent lag and ghosting of a-Se photoconductor as a function of imaging conditions. Two small area electroded a-Se samples, one positively and the other negatively biased on the entrance side of x rays, were used in the experiments. The study of lag and ghosting was performed by delivering a number of raw exposures as experienced in screening mammography to the samples at different electric field strength ESe while measuring the current through the a-Se sample. Ghosting at different operational conditions was quantified as the percentage x-ray sensitivity (x-ray generated photocurrent measured from the sample) reduction compared to before irradiation. Lag was determined by measuring the residual current of a-Se at a given time after the end of each x-ray exposure. Both lag and ghosting were measured as a function of E Se and cumulative exposure. The values of ESe used in our experiments ranged from 1 to 20 V/μm. It was found that ghosting increases with exposure and decreases with ESe for both samples because of the dominant effect of recombination between trapped electrons and x-ray generated holes. Lag on the other hand has different dependence on ESe and cumulative exposure. At ESe ≤ 10 V/μm, the first frame lag for both samples changed slowly with cumulative exposure, with a range of 0.2%-1.7% for the positively biased sample and 0.5%-8% for the negatively biased sample. Overall the positively biased sample has better temporal performance than the negatively biased sample due to the lower density of trapped electrons. The impact of time interval between exposures on the temporal performance was also investigated. Recovery of ghosting with longer time interval was observed, which was attributed to the neutralization of trapped electrons by injected holes through dark current.

KW - Amorphous selenium

KW - Bulk trapping

KW - Ghosting

KW - Lag

KW - Temporal performance

KW - Trapped electrons

UR - http://www.scopus.com/inward/record.url?scp=12844262148&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=12844262148&partnerID=8YFLogxK

U2 - 10.1118/1.1827791

DO - 10.1118/1.1827791

M3 - Article

C2 - 15719963

AN - SCOPUS:12844262148

VL - 32

SP - 128

EP - 136

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 1

ER -