Using a chronically implanted telemetric pressure sensor, we have determined the quantitative relationship between changes in body position and ventricular fluid pressure in normal subjects and subjects with shunts. The method allows accurate, reliable measurement of negative as well as positive pressures. We describe the derangement of postural intraventricular pressure regulation caused by placement of a shunt, as well as the influence of various shunt systems and the antisiphon device on this problem. Ventriculoatrial, venticuloperitoneal, and ventriculopleural shunt systems and the antisiphon device on this problem. Ventriculoatrial, ventriculoperitoneal, and ventriculopleural shunts all caused similar severely abnormal postural intracranial pressure relationships. The antisiphon device was generally effective in restoring normal pressures in patients in the upright position. We discuss the implications of our findings for understanding the mechanisms of postural intracranial pressure regulation in patients without hydrocephalus.

Original languageEnglish (US)
Pages (from-to)181-189
Number of pages9
JournalNeurosurgery
Volume26
Issue number2
StatePublished - 1990

Fingerprint

Body Fluids
Ventricular Pressure
Intracranial Pressure
Pressure
Equipment and Supplies
Ventriculoperitoneal Shunt
Hydrocephalus

Keywords

  • antisiphon device
  • cerebrospinal fluid shunts
  • hydrocephalus
  • intracranial pressure
  • telemetry

ASJC Scopus subject areas

  • Clinical Neurology
  • Surgery

Cite this

The relationship between ventricular fluid pressure and body position in normal subjects and subjects with shunts : A telemetric study. / Chapman, P. H.; Cosman, E. R.; Arnold, M. A.; Portnoy, H. D.; Sklar, F.

In: Neurosurgery, Vol. 26, No. 2, 1990, p. 181-189.

Research output: Contribution to journalArticle

Chapman, P. H. ; Cosman, E. R. ; Arnold, M. A. ; Portnoy, H. D. ; Sklar, F. / The relationship between ventricular fluid pressure and body position in normal subjects and subjects with shunts : A telemetric study. In: Neurosurgery. 1990 ; Vol. 26, No. 2. pp. 181-189.
@article{25432d761333458897c3e13e96c74691,
title = "The relationship between ventricular fluid pressure and body position in normal subjects and subjects with shunts: A telemetric study",
abstract = "Using a chronically implanted telemetric pressure sensor, we have determined the quantitative relationship between changes in body position and ventricular fluid pressure in normal subjects and subjects with shunts. The method allows accurate, reliable measurement of negative as well as positive pressures. We describe the derangement of postural intraventricular pressure regulation caused by placement of a shunt, as well as the influence of various shunt systems and the antisiphon device on this problem. Ventriculoatrial, venticuloperitoneal, and ventriculopleural shunt systems and the antisiphon device on this problem. Ventriculoatrial, ventriculoperitoneal, and ventriculopleural shunts all caused similar severely abnormal postural intracranial pressure relationships. The antisiphon device was generally effective in restoring normal pressures in patients in the upright position. We discuss the implications of our findings for understanding the mechanisms of postural intracranial pressure regulation in patients without hydrocephalus.",
keywords = "antisiphon device, cerebrospinal fluid shunts, hydrocephalus, intracranial pressure, telemetry",
author = "Chapman, {P. H.} and Cosman, {E. R.} and Arnold, {M. A.} and Portnoy, {H. D.} and F. Sklar",
year = "1990",
language = "English (US)",
volume = "26",
pages = "181--189",
journal = "Neurosurgery",
issn = "0148-396X",
publisher = "Lippincott Williams and Wilkins",
number = "2",

}

TY - JOUR

T1 - The relationship between ventricular fluid pressure and body position in normal subjects and subjects with shunts

T2 - A telemetric study

AU - Chapman, P. H.

AU - Cosman, E. R.

AU - Arnold, M. A.

AU - Portnoy, H. D.

AU - Sklar, F.

PY - 1990

Y1 - 1990

N2 - Using a chronically implanted telemetric pressure sensor, we have determined the quantitative relationship between changes in body position and ventricular fluid pressure in normal subjects and subjects with shunts. The method allows accurate, reliable measurement of negative as well as positive pressures. We describe the derangement of postural intraventricular pressure regulation caused by placement of a shunt, as well as the influence of various shunt systems and the antisiphon device on this problem. Ventriculoatrial, venticuloperitoneal, and ventriculopleural shunt systems and the antisiphon device on this problem. Ventriculoatrial, ventriculoperitoneal, and ventriculopleural shunts all caused similar severely abnormal postural intracranial pressure relationships. The antisiphon device was generally effective in restoring normal pressures in patients in the upright position. We discuss the implications of our findings for understanding the mechanisms of postural intracranial pressure regulation in patients without hydrocephalus.

AB - Using a chronically implanted telemetric pressure sensor, we have determined the quantitative relationship between changes in body position and ventricular fluid pressure in normal subjects and subjects with shunts. The method allows accurate, reliable measurement of negative as well as positive pressures. We describe the derangement of postural intraventricular pressure regulation caused by placement of a shunt, as well as the influence of various shunt systems and the antisiphon device on this problem. Ventriculoatrial, venticuloperitoneal, and ventriculopleural shunt systems and the antisiphon device on this problem. Ventriculoatrial, ventriculoperitoneal, and ventriculopleural shunts all caused similar severely abnormal postural intracranial pressure relationships. The antisiphon device was generally effective in restoring normal pressures in patients in the upright position. We discuss the implications of our findings for understanding the mechanisms of postural intracranial pressure regulation in patients without hydrocephalus.

KW - antisiphon device

KW - cerebrospinal fluid shunts

KW - hydrocephalus

KW - intracranial pressure

KW - telemetry

UR - http://www.scopus.com/inward/record.url?scp=0025321746&partnerID=8YFLogxK