Transurethral prostate magnetic resonance elastography: Prospective imaging requirements

Arvin Arani, Donald Plewes, Rajiv Chopra

Research output: Contribution to journalArticle

20 Citations (Scopus)

Abstract

Tissue stiffness is known to undergo alterations when affected by prostate cancer and may serve as an indicator of the disease. Stiffness measurements can be made with magnetic resonance elastography performed using a transurethral actuator to generate shear waves in the prostate gland. The goal of this study was to help determine the imaging requirements of transurethral magnetic resonance elastography and to evaluate whether the spatial and stiffness resolution of this technique overlapped with the requirements for prostate cancer detection. Through the use of prostate-mimicking gelatin phantoms, frequencies of at least 400 Hz were necessary to obtain accurate stiffness measurements of 10 mm diameter inclusions, but the detection of inclusions with diameters as small as 4.75 mm was possible at 200 Hz. The shear wave attenuation coefficient was measured in vivo in the canine prostate gland, and was used to predict the detectable penetration depth of shear waves in prostate tissue. These results suggested that frequencies below 200 Hz could propagate to the prostate boundary with a signal to noise ratio (SNR) of 60 and an actuator capable of producing 60 μm displacements. These requirements are achievable with current imaging and actuator technologies, and motivate further investigation of magnetic resonance elastography for the targeting of prostate cancer.

Original languageEnglish (US)
Pages (from-to)340-349
Number of pages10
JournalMagnetic Resonance in Medicine
Volume65
Issue number2
DOIs
StatePublished - Feb 2011

Fingerprint

Elasticity Imaging Techniques
Prostate
Prostatic Neoplasms
Signal-To-Noise Ratio
Gelatin
Canidae
Technology

Keywords

  • cancer
  • MR elastography
  • prostate imaging
  • tissue stiffness
  • transurethral

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

Transurethral prostate magnetic resonance elastography : Prospective imaging requirements. / Arani, Arvin; Plewes, Donald; Chopra, Rajiv.

In: Magnetic Resonance in Medicine, Vol. 65, No. 2, 02.2011, p. 340-349.

Research output: Contribution to journalArticle

@article{62b2143839684992824fbb90b98148e9,
title = "Transurethral prostate magnetic resonance elastography: Prospective imaging requirements",
abstract = "Tissue stiffness is known to undergo alterations when affected by prostate cancer and may serve as an indicator of the disease. Stiffness measurements can be made with magnetic resonance elastography performed using a transurethral actuator to generate shear waves in the prostate gland. The goal of this study was to help determine the imaging requirements of transurethral magnetic resonance elastography and to evaluate whether the spatial and stiffness resolution of this technique overlapped with the requirements for prostate cancer detection. Through the use of prostate-mimicking gelatin phantoms, frequencies of at least 400 Hz were necessary to obtain accurate stiffness measurements of 10 mm diameter inclusions, but the detection of inclusions with diameters as small as 4.75 mm was possible at 200 Hz. The shear wave attenuation coefficient was measured in vivo in the canine prostate gland, and was used to predict the detectable penetration depth of shear waves in prostate tissue. These results suggested that frequencies below 200 Hz could propagate to the prostate boundary with a signal to noise ratio (SNR) of 60 and an actuator capable of producing 60 μm displacements. These requirements are achievable with current imaging and actuator technologies, and motivate further investigation of magnetic resonance elastography for the targeting of prostate cancer.",
keywords = "cancer, MR elastography, prostate imaging, tissue stiffness, transurethral",
author = "Arvin Arani and Donald Plewes and Rajiv Chopra",
year = "2011",
month = "2",
doi = "10.1002/mrm.22633",
language = "English (US)",
volume = "65",
pages = "340--349",
journal = "Magnetic Resonance in Medicine",
issn = "0740-3194",
publisher = "John Wiley and Sons Inc.",
number = "2",

}

TY - JOUR

T1 - Transurethral prostate magnetic resonance elastography

T2 - Prospective imaging requirements

AU - Arani, Arvin

AU - Plewes, Donald

AU - Chopra, Rajiv

PY - 2011/2

Y1 - 2011/2

N2 - Tissue stiffness is known to undergo alterations when affected by prostate cancer and may serve as an indicator of the disease. Stiffness measurements can be made with magnetic resonance elastography performed using a transurethral actuator to generate shear waves in the prostate gland. The goal of this study was to help determine the imaging requirements of transurethral magnetic resonance elastography and to evaluate whether the spatial and stiffness resolution of this technique overlapped with the requirements for prostate cancer detection. Through the use of prostate-mimicking gelatin phantoms, frequencies of at least 400 Hz were necessary to obtain accurate stiffness measurements of 10 mm diameter inclusions, but the detection of inclusions with diameters as small as 4.75 mm was possible at 200 Hz. The shear wave attenuation coefficient was measured in vivo in the canine prostate gland, and was used to predict the detectable penetration depth of shear waves in prostate tissue. These results suggested that frequencies below 200 Hz could propagate to the prostate boundary with a signal to noise ratio (SNR) of 60 and an actuator capable of producing 60 μm displacements. These requirements are achievable with current imaging and actuator technologies, and motivate further investigation of magnetic resonance elastography for the targeting of prostate cancer.

AB - Tissue stiffness is known to undergo alterations when affected by prostate cancer and may serve as an indicator of the disease. Stiffness measurements can be made with magnetic resonance elastography performed using a transurethral actuator to generate shear waves in the prostate gland. The goal of this study was to help determine the imaging requirements of transurethral magnetic resonance elastography and to evaluate whether the spatial and stiffness resolution of this technique overlapped with the requirements for prostate cancer detection. Through the use of prostate-mimicking gelatin phantoms, frequencies of at least 400 Hz were necessary to obtain accurate stiffness measurements of 10 mm diameter inclusions, but the detection of inclusions with diameters as small as 4.75 mm was possible at 200 Hz. The shear wave attenuation coefficient was measured in vivo in the canine prostate gland, and was used to predict the detectable penetration depth of shear waves in prostate tissue. These results suggested that frequencies below 200 Hz could propagate to the prostate boundary with a signal to noise ratio (SNR) of 60 and an actuator capable of producing 60 μm displacements. These requirements are achievable with current imaging and actuator technologies, and motivate further investigation of magnetic resonance elastography for the targeting of prostate cancer.

KW - cancer

KW - MR elastography

KW - prostate imaging

KW - tissue stiffness

KW - transurethral

UR - http://www.scopus.com/inward/record.url?scp=79251558800&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=79251558800&partnerID=8YFLogxK

U2 - 10.1002/mrm.22633

DO - 10.1002/mrm.22633

M3 - Article

C2 - 20882673

AN - SCOPUS:79251558800

VL - 65

SP - 340

EP - 349

JO - Magnetic Resonance in Medicine

JF - Magnetic Resonance in Medicine

SN - 0740-3194

IS - 2

ER -