Use of 3D adaptive raw-data filter in CT of the lung: Effect on radiation dose reduction

Takeshi Kubo, Yoshiharu Ohno, Shiva Gautam, Pei Jan P Lin, Hans Ulrich Kauczor, Hiroto Hatabu, Masaya Takahashi, Mizuki Nishino, Arkadiusz Sitek, Daisuke Takenaka, Munenobu Nogami, Hisanobu Koyama, Julia Ley-Zaporozhan, Wolfram Stiller, Sebastian Ley, Hiroyasu Inokawa, Miwa Okumura, Yasuko Fujisawa, Hiroyuki Kura, Toshihiro RifuVassilios Raptopoulos, Kazuro Sugimura

Research output: Contribution to journalArticle

24 Citations (Scopus)

Abstract

OBJECTIVE. The purpose of this study was to determine the effectiveness of a 3D adaptive raw-data filter in improving image quality and the role of the filter in radiation dose reduction in lung CT. MATERIALS AND METHODS. Fifty-eight chest CT examinations were performed with a 16-MDCT scanner. Two acquisitions were performed with different tube current-exposure time settings (50 and 150 mAs, 120 kVp). Four series of lung images were prepared from two sets of raw data with and without application of a 3D adaptive filter (50 mAs, 50 mAs with filter, 150 mAs, 150 mAs with filter). Three blinded readers using a 5-point scale from 1 (nondiagnostic) to 5 (excellent) independently evaluated image quality in five lobes and the lingula. A set of images was considered acceptable when scores in all six regions were 3 (acceptable) or higher. The SD of attenuation was calculated in 24 regions of interest. RESULTS. The overall mean image quality scores were 3.09, 3.53, 4.02, and 4.38 for the 50 mAs, 50 mAs with filter, 150 mAs, and 150 mAs with filter sets, respectively. Scores were significantly better with filter application (p < 0.001). A significant decrease in SD of attenuation was observed with filter application (p < 0.001). Among the respective series of images, 18, 52, 50, and 58 sets were judged acceptable with no significant difference in acceptability between images obtained at 50 mAs with a filter and at 150 mAs (p = 0.72). With filter application, the acceptability of 50-mAs images became comparable with that of 150-mAs images, making dose reduction to 50 mAs practical. CONCLUSION. Use of a 3D adaptive raw-data filter improved the quality of lung images, making dose reduction to 50 mAs attainable with use of the filter.

Original languageEnglish (US)
JournalAmerican Journal of Roentgenology
Volume191
Issue number4
DOIs
StatePublished - Oct 2008

Fingerprint

Radiation Effects
Lung
Thorax
Radiation

Keywords

  • Artifacts
  • Chest CT
  • Imaging filter
  • Radiation dose

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

Kubo, T., Ohno, Y., Gautam, S., Lin, P. J. P., Kauczor, H. U., Hatabu, H., ... Sugimura, K. (2008). Use of 3D adaptive raw-data filter in CT of the lung: Effect on radiation dose reduction. American Journal of Roentgenology, 191(4). https://doi.org/10.2214/AJR/07.2630

Use of 3D adaptive raw-data filter in CT of the lung : Effect on radiation dose reduction. / Kubo, Takeshi; Ohno, Yoshiharu; Gautam, Shiva; Lin, Pei Jan P; Kauczor, Hans Ulrich; Hatabu, Hiroto; Takahashi, Masaya; Nishino, Mizuki; Sitek, Arkadiusz; Takenaka, Daisuke; Nogami, Munenobu; Koyama, Hisanobu; Ley-Zaporozhan, Julia; Stiller, Wolfram; Ley, Sebastian; Inokawa, Hiroyasu; Okumura, Miwa; Fujisawa, Yasuko; Kura, Hiroyuki; Rifu, Toshihiro; Raptopoulos, Vassilios; Sugimura, Kazuro.

In: American Journal of Roentgenology, Vol. 191, No. 4, 10.2008.

Research output: Contribution to journalArticle

Kubo, T, Ohno, Y, Gautam, S, Lin, PJP, Kauczor, HU, Hatabu, H, Takahashi, M, Nishino, M, Sitek, A, Takenaka, D, Nogami, M, Koyama, H, Ley-Zaporozhan, J, Stiller, W, Ley, S, Inokawa, H, Okumura, M, Fujisawa, Y, Kura, H, Rifu, T, Raptopoulos, V & Sugimura, K 2008, 'Use of 3D adaptive raw-data filter in CT of the lung: Effect on radiation dose reduction', American Journal of Roentgenology, vol. 191, no. 4. https://doi.org/10.2214/AJR/07.2630
Kubo, Takeshi ; Ohno, Yoshiharu ; Gautam, Shiva ; Lin, Pei Jan P ; Kauczor, Hans Ulrich ; Hatabu, Hiroto ; Takahashi, Masaya ; Nishino, Mizuki ; Sitek, Arkadiusz ; Takenaka, Daisuke ; Nogami, Munenobu ; Koyama, Hisanobu ; Ley-Zaporozhan, Julia ; Stiller, Wolfram ; Ley, Sebastian ; Inokawa, Hiroyasu ; Okumura, Miwa ; Fujisawa, Yasuko ; Kura, Hiroyuki ; Rifu, Toshihiro ; Raptopoulos, Vassilios ; Sugimura, Kazuro. / Use of 3D adaptive raw-data filter in CT of the lung : Effect on radiation dose reduction. In: American Journal of Roentgenology. 2008 ; Vol. 191, No. 4.
@article{95ede3a0e7864fbaabfcba3fe3d8609b,
title = "Use of 3D adaptive raw-data filter in CT of the lung: Effect on radiation dose reduction",
abstract = "OBJECTIVE. The purpose of this study was to determine the effectiveness of a 3D adaptive raw-data filter in improving image quality and the role of the filter in radiation dose reduction in lung CT. MATERIALS AND METHODS. Fifty-eight chest CT examinations were performed with a 16-MDCT scanner. Two acquisitions were performed with different tube current-exposure time settings (50 and 150 mAs, 120 kVp). Four series of lung images were prepared from two sets of raw data with and without application of a 3D adaptive filter (50 mAs, 50 mAs with filter, 150 mAs, 150 mAs with filter). Three blinded readers using a 5-point scale from 1 (nondiagnostic) to 5 (excellent) independently evaluated image quality in five lobes and the lingula. A set of images was considered acceptable when scores in all six regions were 3 (acceptable) or higher. The SD of attenuation was calculated in 24 regions of interest. RESULTS. The overall mean image quality scores were 3.09, 3.53, 4.02, and 4.38 for the 50 mAs, 50 mAs with filter, 150 mAs, and 150 mAs with filter sets, respectively. Scores were significantly better with filter application (p < 0.001). A significant decrease in SD of attenuation was observed with filter application (p < 0.001). Among the respective series of images, 18, 52, 50, and 58 sets were judged acceptable with no significant difference in acceptability between images obtained at 50 mAs with a filter and at 150 mAs (p = 0.72). With filter application, the acceptability of 50-mAs images became comparable with that of 150-mAs images, making dose reduction to 50 mAs practical. CONCLUSION. Use of a 3D adaptive raw-data filter improved the quality of lung images, making dose reduction to 50 mAs attainable with use of the filter.",
keywords = "Artifacts, Chest CT, Imaging filter, Radiation dose",
author = "Takeshi Kubo and Yoshiharu Ohno and Shiva Gautam and Lin, {Pei Jan P} and Kauczor, {Hans Ulrich} and Hiroto Hatabu and Masaya Takahashi and Mizuki Nishino and Arkadiusz Sitek and Daisuke Takenaka and Munenobu Nogami and Hisanobu Koyama and Julia Ley-Zaporozhan and Wolfram Stiller and Sebastian Ley and Hiroyasu Inokawa and Miwa Okumura and Yasuko Fujisawa and Hiroyuki Kura and Toshihiro Rifu and Vassilios Raptopoulos and Kazuro Sugimura",
year = "2008",
month = "10",
doi = "10.2214/AJR/07.2630",
language = "English (US)",
volume = "191",
journal = "American Journal of Roentgenology",
issn = "0361-803X",
publisher = "American Roentgen Ray Society",
number = "4",

}

TY - JOUR

T1 - Use of 3D adaptive raw-data filter in CT of the lung

T2 - Effect on radiation dose reduction

AU - Kubo, Takeshi

AU - Ohno, Yoshiharu

AU - Gautam, Shiva

AU - Lin, Pei Jan P

AU - Kauczor, Hans Ulrich

AU - Hatabu, Hiroto

AU - Takahashi, Masaya

AU - Nishino, Mizuki

AU - Sitek, Arkadiusz

AU - Takenaka, Daisuke

AU - Nogami, Munenobu

AU - Koyama, Hisanobu

AU - Ley-Zaporozhan, Julia

AU - Stiller, Wolfram

AU - Ley, Sebastian

AU - Inokawa, Hiroyasu

AU - Okumura, Miwa

AU - Fujisawa, Yasuko

AU - Kura, Hiroyuki

AU - Rifu, Toshihiro

AU - Raptopoulos, Vassilios

AU - Sugimura, Kazuro

PY - 2008/10

Y1 - 2008/10

N2 - OBJECTIVE. The purpose of this study was to determine the effectiveness of a 3D adaptive raw-data filter in improving image quality and the role of the filter in radiation dose reduction in lung CT. MATERIALS AND METHODS. Fifty-eight chest CT examinations were performed with a 16-MDCT scanner. Two acquisitions were performed with different tube current-exposure time settings (50 and 150 mAs, 120 kVp). Four series of lung images were prepared from two sets of raw data with and without application of a 3D adaptive filter (50 mAs, 50 mAs with filter, 150 mAs, 150 mAs with filter). Three blinded readers using a 5-point scale from 1 (nondiagnostic) to 5 (excellent) independently evaluated image quality in five lobes and the lingula. A set of images was considered acceptable when scores in all six regions were 3 (acceptable) or higher. The SD of attenuation was calculated in 24 regions of interest. RESULTS. The overall mean image quality scores were 3.09, 3.53, 4.02, and 4.38 for the 50 mAs, 50 mAs with filter, 150 mAs, and 150 mAs with filter sets, respectively. Scores were significantly better with filter application (p < 0.001). A significant decrease in SD of attenuation was observed with filter application (p < 0.001). Among the respective series of images, 18, 52, 50, and 58 sets were judged acceptable with no significant difference in acceptability between images obtained at 50 mAs with a filter and at 150 mAs (p = 0.72). With filter application, the acceptability of 50-mAs images became comparable with that of 150-mAs images, making dose reduction to 50 mAs practical. CONCLUSION. Use of a 3D adaptive raw-data filter improved the quality of lung images, making dose reduction to 50 mAs attainable with use of the filter.

AB - OBJECTIVE. The purpose of this study was to determine the effectiveness of a 3D adaptive raw-data filter in improving image quality and the role of the filter in radiation dose reduction in lung CT. MATERIALS AND METHODS. Fifty-eight chest CT examinations were performed with a 16-MDCT scanner. Two acquisitions were performed with different tube current-exposure time settings (50 and 150 mAs, 120 kVp). Four series of lung images were prepared from two sets of raw data with and without application of a 3D adaptive filter (50 mAs, 50 mAs with filter, 150 mAs, 150 mAs with filter). Three blinded readers using a 5-point scale from 1 (nondiagnostic) to 5 (excellent) independently evaluated image quality in five lobes and the lingula. A set of images was considered acceptable when scores in all six regions were 3 (acceptable) or higher. The SD of attenuation was calculated in 24 regions of interest. RESULTS. The overall mean image quality scores were 3.09, 3.53, 4.02, and 4.38 for the 50 mAs, 50 mAs with filter, 150 mAs, and 150 mAs with filter sets, respectively. Scores were significantly better with filter application (p < 0.001). A significant decrease in SD of attenuation was observed with filter application (p < 0.001). Among the respective series of images, 18, 52, 50, and 58 sets were judged acceptable with no significant difference in acceptability between images obtained at 50 mAs with a filter and at 150 mAs (p = 0.72). With filter application, the acceptability of 50-mAs images became comparable with that of 150-mAs images, making dose reduction to 50 mAs practical. CONCLUSION. Use of a 3D adaptive raw-data filter improved the quality of lung images, making dose reduction to 50 mAs attainable with use of the filter.

KW - Artifacts

KW - Chest CT

KW - Imaging filter

KW - Radiation dose

UR - http://www.scopus.com/inward/record.url?scp=53249140708&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=53249140708&partnerID=8YFLogxK

U2 - 10.2214/AJR/07.2630

DO - 10.2214/AJR/07.2630

M3 - Article

C2 - 18806144

AN - SCOPUS:53249140708

VL - 191

JO - American Journal of Roentgenology

JF - American Journal of Roentgenology

SN - 0361-803X

IS - 4

ER -