Validation for tremor quantification of an electromagnetic tracking device

Research output: Contribution to journalArticle

41 Citations (Scopus)

Abstract

An electromagnetic tracking system was used to record arm motion in subjects with Parkinson's disease (n = 23), essential tremor (n = 28) or without neurological disease (n = 4). Tremor magnitude was calculated by averaging the three-dimensional displacement of individual tremor bursts. Tremor magnitude calculated in this manner was quite closely correlated with a clinician's estimate (r = 0.88 and 0.86 for Parkinsonian and essential tremors, respectively) and was reproducible (r = 0.93 for repeated recordings). The accuracy of the device and algorithm was confirmed by mechanically generating oscillations of known magnitudes and frequencies. This device is adaptable for quantifying different types of tremors and its accuracy is easy to verify. Because position rather than acceleration is tracked, tremor amplitude can be stated in readily comprehensible units.

Original languageEnglish (US)
Pages (from-to)265-271
Number of pages7
JournalMovement Disorders
Volume16
Issue number2
DOIs
StatePublished - Mar 2001

Fingerprint

Electromagnetic Phenomena
Tremor
Equipment and Supplies
Essential Tremor
Parkinson Disease

Keywords

  • Electromagnetic tracking device
  • Essential tremor
  • Parkinson's disease
  • Tremor quantification

ASJC Scopus subject areas

  • Clinical Neurology
  • Neuroscience(all)

Cite this

Validation for tremor quantification of an electromagnetic tracking device. / O'Suilleabhain, Padraig E.; Dewey, Richard B.

In: Movement Disorders, Vol. 16, No. 2, 03.2001, p. 265-271.

Research output: Contribution to journalArticle

@article{83ba20c600f0466cb21defedcf5ab402,
title = "Validation for tremor quantification of an electromagnetic tracking device",
abstract = "An electromagnetic tracking system was used to record arm motion in subjects with Parkinson's disease (n = 23), essential tremor (n = 28) or without neurological disease (n = 4). Tremor magnitude was calculated by averaging the three-dimensional displacement of individual tremor bursts. Tremor magnitude calculated in this manner was quite closely correlated with a clinician's estimate (r = 0.88 and 0.86 for Parkinsonian and essential tremors, respectively) and was reproducible (r = 0.93 for repeated recordings). The accuracy of the device and algorithm was confirmed by mechanically generating oscillations of known magnitudes and frequencies. This device is adaptable for quantifying different types of tremors and its accuracy is easy to verify. Because position rather than acceleration is tracked, tremor amplitude can be stated in readily comprehensible units.",
keywords = "Electromagnetic tracking device, Essential tremor, Parkinson's disease, Tremor quantification",
author = "O'Suilleabhain, {Padraig E.} and Dewey, {Richard B.}",
year = "2001",
month = "3",
doi = "10.1002/mds.1064",
language = "English (US)",
volume = "16",
pages = "265--271",
journal = "Movement Disorders",
issn = "0885-3185",
publisher = "John Wiley and Sons Inc.",
number = "2",

}

TY - JOUR

T1 - Validation for tremor quantification of an electromagnetic tracking device

AU - O'Suilleabhain, Padraig E.

AU - Dewey, Richard B.

PY - 2001/3

Y1 - 2001/3

N2 - An electromagnetic tracking system was used to record arm motion in subjects with Parkinson's disease (n = 23), essential tremor (n = 28) or without neurological disease (n = 4). Tremor magnitude was calculated by averaging the three-dimensional displacement of individual tremor bursts. Tremor magnitude calculated in this manner was quite closely correlated with a clinician's estimate (r = 0.88 and 0.86 for Parkinsonian and essential tremors, respectively) and was reproducible (r = 0.93 for repeated recordings). The accuracy of the device and algorithm was confirmed by mechanically generating oscillations of known magnitudes and frequencies. This device is adaptable for quantifying different types of tremors and its accuracy is easy to verify. Because position rather than acceleration is tracked, tremor amplitude can be stated in readily comprehensible units.

AB - An electromagnetic tracking system was used to record arm motion in subjects with Parkinson's disease (n = 23), essential tremor (n = 28) or without neurological disease (n = 4). Tremor magnitude was calculated by averaging the three-dimensional displacement of individual tremor bursts. Tremor magnitude calculated in this manner was quite closely correlated with a clinician's estimate (r = 0.88 and 0.86 for Parkinsonian and essential tremors, respectively) and was reproducible (r = 0.93 for repeated recordings). The accuracy of the device and algorithm was confirmed by mechanically generating oscillations of known magnitudes and frequencies. This device is adaptable for quantifying different types of tremors and its accuracy is easy to verify. Because position rather than acceleration is tracked, tremor amplitude can be stated in readily comprehensible units.

KW - Electromagnetic tracking device

KW - Essential tremor

KW - Parkinson's disease

KW - Tremor quantification

UR - http://www.scopus.com/inward/record.url?scp=0035297097&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0035297097&partnerID=8YFLogxK

U2 - 10.1002/mds.1064

DO - 10.1002/mds.1064

M3 - Article

C2 - 11295779

AN - SCOPUS:0035297097

VL - 16

SP - 265

EP - 271

JO - Movement Disorders

JF - Movement Disorders

SN - 0885-3185

IS - 2

ER -