Validation of a near-infrared probe for detection of thin intracranial white matter structures

Cole A. Giller, Hanli Liu, Prem Gurnani, Sundar Victor, Umar Yazdani, Dwight C. German

Research output: Contribution to journalArticle

36 Citations (Scopus)

Abstract

Object. The authors have developed an intracranial near-infrared (NIR) probe that analyzes the scattering of light emitted from its tip to measure the optical properties of cerebral tissue. Despite its success in distinguishing gray matter from white matter in humans during stereotactic surgery, the limits of this instrument's resolution remain unclear. In this study, the authors determined the spatial resolution of this new probe by using a rodent model supplemented with phantom measurements and computer simulation. Methods. A phantom consisting of Intralipid and gelatin was constructed to resemble a layer of white matter overlying a layer of gray matter. Near-infrared measurements were obtained as the probe was inserted through the gray-white matter transition. A computer simulation of NIR measurements through a gray-white matter transition was also performed using Monte Carlo techniques. The NIR probe was then used to study 19 tracks from the cortical surface through the corpus callosum in an in vivo rodent preparation. The animals were killed and histological sections through the tracks were obtained. Data from the phantom models and computer simulations showed that the NIR probe samples a volume of tissue extending 1 to 1.5 mm in front of the probe tip (this distance is termed the "lookthrough" distance). Measurements obtained from an NIR probe passing through a thin layer of white matter consisted of an initial segment of increasing values, a maximum (peak) value, and a trailing segment of decreasing values. The length of the initial segment is the look-through distance, the position of the peak indicates the location of the superficial white matter boundary, and the length of the trailing segment is the thickness of the layer. These considerations were confirmed in experiments with rodents. All tracks passed through the corpus callosum, which was demonstrated as a broad peak on each NIR graph. The position of the dorsal boundary of the corpus callosum and its width (based on histological measurements) correlated well with the peak of the NIR curve and its trailing segment, respectively. The initial segments correlated well with estimates of the lookthrough distance. Five of the tracks transected the smaller anterior commissure (diameter 0.2 mm), producing a narrow NIR peak at the correct depth. Conclusions. Data in this study confirm that the NIR probe can reliably detect and measure the thickness of layers of white matter as thin as 0.2 mm. Such resolution should be adequate to detect larger structures of interest encountered during stereotactic surgery in humans.

Original languageEnglish (US)
Pages (from-to)1299-1306
Number of pages8
JournalJournal of Neurosurgery
Volume98
Issue number6
StatePublished - Jun 1 2003

Fingerprint

Corpus Callosum
Computer Simulation
Rodentia
Supine Position
Gelatin
White Matter
Light
Gray Matter

Keywords

  • Deep brain stimulation
  • Near-infrared spectroscopy
  • Rat
  • Stereotactic surgery

ASJC Scopus subject areas

  • Clinical Neurology
  • Neuroscience(all)

Cite this

Giller, C. A., Liu, H., Gurnani, P., Victor, S., Yazdani, U., & German, D. C. (2003). Validation of a near-infrared probe for detection of thin intracranial white matter structures. Journal of Neurosurgery, 98(6), 1299-1306.

Validation of a near-infrared probe for detection of thin intracranial white matter structures. / Giller, Cole A.; Liu, Hanli; Gurnani, Prem; Victor, Sundar; Yazdani, Umar; German, Dwight C.

In: Journal of Neurosurgery, Vol. 98, No. 6, 01.06.2003, p. 1299-1306.

Research output: Contribution to journalArticle

Giller, CA, Liu, H, Gurnani, P, Victor, S, Yazdani, U & German, DC 2003, 'Validation of a near-infrared probe for detection of thin intracranial white matter structures', Journal of Neurosurgery, vol. 98, no. 6, pp. 1299-1306.
Giller, Cole A. ; Liu, Hanli ; Gurnani, Prem ; Victor, Sundar ; Yazdani, Umar ; German, Dwight C. / Validation of a near-infrared probe for detection of thin intracranial white matter structures. In: Journal of Neurosurgery. 2003 ; Vol. 98, No. 6. pp. 1299-1306.
@article{8685192bf5aa4a23aa0ed5024d7e8f79,
title = "Validation of a near-infrared probe for detection of thin intracranial white matter structures",
abstract = "Object. The authors have developed an intracranial near-infrared (NIR) probe that analyzes the scattering of light emitted from its tip to measure the optical properties of cerebral tissue. Despite its success in distinguishing gray matter from white matter in humans during stereotactic surgery, the limits of this instrument's resolution remain unclear. In this study, the authors determined the spatial resolution of this new probe by using a rodent model supplemented with phantom measurements and computer simulation. Methods. A phantom consisting of Intralipid and gelatin was constructed to resemble a layer of white matter overlying a layer of gray matter. Near-infrared measurements were obtained as the probe was inserted through the gray-white matter transition. A computer simulation of NIR measurements through a gray-white matter transition was also performed using Monte Carlo techniques. The NIR probe was then used to study 19 tracks from the cortical surface through the corpus callosum in an in vivo rodent preparation. The animals were killed and histological sections through the tracks were obtained. Data from the phantom models and computer simulations showed that the NIR probe samples a volume of tissue extending 1 to 1.5 mm in front of the probe tip (this distance is termed the {"}lookthrough{"} distance). Measurements obtained from an NIR probe passing through a thin layer of white matter consisted of an initial segment of increasing values, a maximum (peak) value, and a trailing segment of decreasing values. The length of the initial segment is the look-through distance, the position of the peak indicates the location of the superficial white matter boundary, and the length of the trailing segment is the thickness of the layer. These considerations were confirmed in experiments with rodents. All tracks passed through the corpus callosum, which was demonstrated as a broad peak on each NIR graph. The position of the dorsal boundary of the corpus callosum and its width (based on histological measurements) correlated well with the peak of the NIR curve and its trailing segment, respectively. The initial segments correlated well with estimates of the lookthrough distance. Five of the tracks transected the smaller anterior commissure (diameter 0.2 mm), producing a narrow NIR peak at the correct depth. Conclusions. Data in this study confirm that the NIR probe can reliably detect and measure the thickness of layers of white matter as thin as 0.2 mm. Such resolution should be adequate to detect larger structures of interest encountered during stereotactic surgery in humans.",
keywords = "Deep brain stimulation, Near-infrared spectroscopy, Rat, Stereotactic surgery",
author = "Giller, {Cole A.} and Hanli Liu and Prem Gurnani and Sundar Victor and Umar Yazdani and German, {Dwight C.}",
year = "2003",
month = "6",
day = "1",
language = "English (US)",
volume = "98",
pages = "1299--1306",
journal = "Journal of Neurosurgery",
issn = "0022-3085",
publisher = "American Association of Neurological Surgeons",
number = "6",

}

TY - JOUR

T1 - Validation of a near-infrared probe for detection of thin intracranial white matter structures

AU - Giller, Cole A.

AU - Liu, Hanli

AU - Gurnani, Prem

AU - Victor, Sundar

AU - Yazdani, Umar

AU - German, Dwight C.

PY - 2003/6/1

Y1 - 2003/6/1

N2 - Object. The authors have developed an intracranial near-infrared (NIR) probe that analyzes the scattering of light emitted from its tip to measure the optical properties of cerebral tissue. Despite its success in distinguishing gray matter from white matter in humans during stereotactic surgery, the limits of this instrument's resolution remain unclear. In this study, the authors determined the spatial resolution of this new probe by using a rodent model supplemented with phantom measurements and computer simulation. Methods. A phantom consisting of Intralipid and gelatin was constructed to resemble a layer of white matter overlying a layer of gray matter. Near-infrared measurements were obtained as the probe was inserted through the gray-white matter transition. A computer simulation of NIR measurements through a gray-white matter transition was also performed using Monte Carlo techniques. The NIR probe was then used to study 19 tracks from the cortical surface through the corpus callosum in an in vivo rodent preparation. The animals were killed and histological sections through the tracks were obtained. Data from the phantom models and computer simulations showed that the NIR probe samples a volume of tissue extending 1 to 1.5 mm in front of the probe tip (this distance is termed the "lookthrough" distance). Measurements obtained from an NIR probe passing through a thin layer of white matter consisted of an initial segment of increasing values, a maximum (peak) value, and a trailing segment of decreasing values. The length of the initial segment is the look-through distance, the position of the peak indicates the location of the superficial white matter boundary, and the length of the trailing segment is the thickness of the layer. These considerations were confirmed in experiments with rodents. All tracks passed through the corpus callosum, which was demonstrated as a broad peak on each NIR graph. The position of the dorsal boundary of the corpus callosum and its width (based on histological measurements) correlated well with the peak of the NIR curve and its trailing segment, respectively. The initial segments correlated well with estimates of the lookthrough distance. Five of the tracks transected the smaller anterior commissure (diameter 0.2 mm), producing a narrow NIR peak at the correct depth. Conclusions. Data in this study confirm that the NIR probe can reliably detect and measure the thickness of layers of white matter as thin as 0.2 mm. Such resolution should be adequate to detect larger structures of interest encountered during stereotactic surgery in humans.

AB - Object. The authors have developed an intracranial near-infrared (NIR) probe that analyzes the scattering of light emitted from its tip to measure the optical properties of cerebral tissue. Despite its success in distinguishing gray matter from white matter in humans during stereotactic surgery, the limits of this instrument's resolution remain unclear. In this study, the authors determined the spatial resolution of this new probe by using a rodent model supplemented with phantom measurements and computer simulation. Methods. A phantom consisting of Intralipid and gelatin was constructed to resemble a layer of white matter overlying a layer of gray matter. Near-infrared measurements were obtained as the probe was inserted through the gray-white matter transition. A computer simulation of NIR measurements through a gray-white matter transition was also performed using Monte Carlo techniques. The NIR probe was then used to study 19 tracks from the cortical surface through the corpus callosum in an in vivo rodent preparation. The animals were killed and histological sections through the tracks were obtained. Data from the phantom models and computer simulations showed that the NIR probe samples a volume of tissue extending 1 to 1.5 mm in front of the probe tip (this distance is termed the "lookthrough" distance). Measurements obtained from an NIR probe passing through a thin layer of white matter consisted of an initial segment of increasing values, a maximum (peak) value, and a trailing segment of decreasing values. The length of the initial segment is the look-through distance, the position of the peak indicates the location of the superficial white matter boundary, and the length of the trailing segment is the thickness of the layer. These considerations were confirmed in experiments with rodents. All tracks passed through the corpus callosum, which was demonstrated as a broad peak on each NIR graph. The position of the dorsal boundary of the corpus callosum and its width (based on histological measurements) correlated well with the peak of the NIR curve and its trailing segment, respectively. The initial segments correlated well with estimates of the lookthrough distance. Five of the tracks transected the smaller anterior commissure (diameter 0.2 mm), producing a narrow NIR peak at the correct depth. Conclusions. Data in this study confirm that the NIR probe can reliably detect and measure the thickness of layers of white matter as thin as 0.2 mm. Such resolution should be adequate to detect larger structures of interest encountered during stereotactic surgery in humans.

KW - Deep brain stimulation

KW - Near-infrared spectroscopy

KW - Rat

KW - Stereotactic surgery

UR - http://www.scopus.com/inward/record.url?scp=0037512469&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0037512469&partnerID=8YFLogxK

M3 - Article

C2 - 12816278

AN - SCOPUS:0037512469

VL - 98

SP - 1299

EP - 1306

JO - Journal of Neurosurgery

JF - Journal of Neurosurgery

SN - 0022-3085

IS - 6

ER -