WE‐D‐224C‐07: A Comprehensive Patient‐Specific IMRT Quality Assurance Procedure On Hi‐Art Tomotherapy® Unit

Y. Yan, H. Jiang, X. Weng, J. Penagaricano, E. Moros, N. Papanikolaou, C. Shi, P. Novak, V. Ratanatharathorn

Research output: Contribution to journalArticle

Abstract

Purpose: To present a comprehensive patient‐specific IMRT QA procedure on Tomotherapy® unit using ion chamber and radiographic film for those facilities with RIT® film dosimetry system. Method and Materials: A cylindrical Virtual Water® phantom was scanned in CT with an ion chamber in the hole 1.5 cm below the middle coronal plane. Three‐dimensional dose calculation in fine grid using treatment beam was performed to the phantom which is appropriately positioned so that the measuring point and plane are within the volume of interest. A sheet of Kokak® EDR‐2 Readypack was sandwiched between two halves of the phantom and pricked on green lasers. Film exposure and ion chamber measurement were made simultaneously. To verify planar absolute dose distribution, the film was calibrated in a separate sheet of the same batch on the same day IMRT QA was performed. To facilitate the film calibration and film dosimetry analysis on Tomotherapy planning station, an in‐house software was developed to perform step‐valley film calibration and film image data conversion from RIT® format to DICOM format which is acceptable by the current version of Tomotherapy planning system. Results: 213 patient‐specific IMRT QA procedures have been performed on Tomotherapy unit since it was installed about two years ago. The average discrepancy between ion chamber measurement and dose calculation in phantom is 1.74%. Film dosimetry results agree with ion chamber measurements in an average discrepancy 1%. Three cases have failed the QA procedure based on 5% discrepancy criteria. Conclusion: Three‐dimensional dose calculation algorithm for helical IMRT works very well in homogeneous phantom. This QA procedure is clinically practical and necessary when independent dose calculation tools are not available. It is feasible to verify two‐dimensional absolute dose distribution using radiographic film if it is calibrated against ion chamber reading which is traceable to ADCL or NIST.

Original languageEnglish (US)
Number of pages1
JournalMedical Physics
Volume33
Issue number6
DOIs
StatePublished - 2006

Fingerprint

Film Dosimetry
Ions
X-Ray Film
Calibration
Motion Pictures
Reading
Lasers
Software
Water

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

WE‐D‐224C‐07 : A Comprehensive Patient‐Specific IMRT Quality Assurance Procedure On Hi‐Art Tomotherapy® Unit. / Yan, Y.; Jiang, H.; Weng, X.; Penagaricano, J.; Moros, E.; Papanikolaou, N.; Shi, C.; Novak, P.; Ratanatharathorn, V.

In: Medical Physics, Vol. 33, No. 6, 2006.

Research output: Contribution to journalArticle

Yan, Y, Jiang, H, Weng, X, Penagaricano, J, Moros, E, Papanikolaou, N, Shi, C, Novak, P & Ratanatharathorn, V 2006, 'WE‐D‐224C‐07: A Comprehensive Patient‐Specific IMRT Quality Assurance Procedure On Hi‐Art Tomotherapy® Unit', Medical Physics, vol. 33, no. 6. https://doi.org/10.1118/1.2241787
Yan, Y. ; Jiang, H. ; Weng, X. ; Penagaricano, J. ; Moros, E. ; Papanikolaou, N. ; Shi, C. ; Novak, P. ; Ratanatharathorn, V. / WE‐D‐224C‐07 : A Comprehensive Patient‐Specific IMRT Quality Assurance Procedure On Hi‐Art Tomotherapy® Unit. In: Medical Physics. 2006 ; Vol. 33, No. 6.
@article{07121bc9023447e9976baf2914f74f29,
title = "WE‐D‐224C‐07: A Comprehensive Patient‐Specific IMRT Quality Assurance Procedure On Hi‐Art Tomotherapy{\circledR} Unit",
abstract = "Purpose: To present a comprehensive patient‐specific IMRT QA procedure on Tomotherapy{\circledR} unit using ion chamber and radiographic film for those facilities with RIT{\circledR} film dosimetry system. Method and Materials: A cylindrical Virtual Water{\circledR} phantom was scanned in CT with an ion chamber in the hole 1.5 cm below the middle coronal plane. Three‐dimensional dose calculation in fine grid using treatment beam was performed to the phantom which is appropriately positioned so that the measuring point and plane are within the volume of interest. A sheet of Kokak{\circledR} EDR‐2 Readypack was sandwiched between two halves of the phantom and pricked on green lasers. Film exposure and ion chamber measurement were made simultaneously. To verify planar absolute dose distribution, the film was calibrated in a separate sheet of the same batch on the same day IMRT QA was performed. To facilitate the film calibration and film dosimetry analysis on Tomotherapy planning station, an in‐house software was developed to perform step‐valley film calibration and film image data conversion from RIT{\circledR} format to DICOM format which is acceptable by the current version of Tomotherapy planning system. Results: 213 patient‐specific IMRT QA procedures have been performed on Tomotherapy unit since it was installed about two years ago. The average discrepancy between ion chamber measurement and dose calculation in phantom is 1.74{\%}. Film dosimetry results agree with ion chamber measurements in an average discrepancy 1{\%}. Three cases have failed the QA procedure based on 5{\%} discrepancy criteria. Conclusion: Three‐dimensional dose calculation algorithm for helical IMRT works very well in homogeneous phantom. This QA procedure is clinically practical and necessary when independent dose calculation tools are not available. It is feasible to verify two‐dimensional absolute dose distribution using radiographic film if it is calibrated against ion chamber reading which is traceable to ADCL or NIST.",
author = "Y. Yan and H. Jiang and X. Weng and J. Penagaricano and E. Moros and N. Papanikolaou and C. Shi and P. Novak and V. Ratanatharathorn",
year = "2006",
doi = "10.1118/1.2241787",
language = "English (US)",
volume = "33",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - WE‐D‐224C‐07

T2 - A Comprehensive Patient‐Specific IMRT Quality Assurance Procedure On Hi‐Art Tomotherapy® Unit

AU - Yan, Y.

AU - Jiang, H.

AU - Weng, X.

AU - Penagaricano, J.

AU - Moros, E.

AU - Papanikolaou, N.

AU - Shi, C.

AU - Novak, P.

AU - Ratanatharathorn, V.

PY - 2006

Y1 - 2006

N2 - Purpose: To present a comprehensive patient‐specific IMRT QA procedure on Tomotherapy® unit using ion chamber and radiographic film for those facilities with RIT® film dosimetry system. Method and Materials: A cylindrical Virtual Water® phantom was scanned in CT with an ion chamber in the hole 1.5 cm below the middle coronal plane. Three‐dimensional dose calculation in fine grid using treatment beam was performed to the phantom which is appropriately positioned so that the measuring point and plane are within the volume of interest. A sheet of Kokak® EDR‐2 Readypack was sandwiched between two halves of the phantom and pricked on green lasers. Film exposure and ion chamber measurement were made simultaneously. To verify planar absolute dose distribution, the film was calibrated in a separate sheet of the same batch on the same day IMRT QA was performed. To facilitate the film calibration and film dosimetry analysis on Tomotherapy planning station, an in‐house software was developed to perform step‐valley film calibration and film image data conversion from RIT® format to DICOM format which is acceptable by the current version of Tomotherapy planning system. Results: 213 patient‐specific IMRT QA procedures have been performed on Tomotherapy unit since it was installed about two years ago. The average discrepancy between ion chamber measurement and dose calculation in phantom is 1.74%. Film dosimetry results agree with ion chamber measurements in an average discrepancy 1%. Three cases have failed the QA procedure based on 5% discrepancy criteria. Conclusion: Three‐dimensional dose calculation algorithm for helical IMRT works very well in homogeneous phantom. This QA procedure is clinically practical and necessary when independent dose calculation tools are not available. It is feasible to verify two‐dimensional absolute dose distribution using radiographic film if it is calibrated against ion chamber reading which is traceable to ADCL or NIST.

AB - Purpose: To present a comprehensive patient‐specific IMRT QA procedure on Tomotherapy® unit using ion chamber and radiographic film for those facilities with RIT® film dosimetry system. Method and Materials: A cylindrical Virtual Water® phantom was scanned in CT with an ion chamber in the hole 1.5 cm below the middle coronal plane. Three‐dimensional dose calculation in fine grid using treatment beam was performed to the phantom which is appropriately positioned so that the measuring point and plane are within the volume of interest. A sheet of Kokak® EDR‐2 Readypack was sandwiched between two halves of the phantom and pricked on green lasers. Film exposure and ion chamber measurement were made simultaneously. To verify planar absolute dose distribution, the film was calibrated in a separate sheet of the same batch on the same day IMRT QA was performed. To facilitate the film calibration and film dosimetry analysis on Tomotherapy planning station, an in‐house software was developed to perform step‐valley film calibration and film image data conversion from RIT® format to DICOM format which is acceptable by the current version of Tomotherapy planning system. Results: 213 patient‐specific IMRT QA procedures have been performed on Tomotherapy unit since it was installed about two years ago. The average discrepancy between ion chamber measurement and dose calculation in phantom is 1.74%. Film dosimetry results agree with ion chamber measurements in an average discrepancy 1%. Three cases have failed the QA procedure based on 5% discrepancy criteria. Conclusion: Three‐dimensional dose calculation algorithm for helical IMRT works very well in homogeneous phantom. This QA procedure is clinically practical and necessary when independent dose calculation tools are not available. It is feasible to verify two‐dimensional absolute dose distribution using radiographic film if it is calibrated against ion chamber reading which is traceable to ADCL or NIST.

UR - http://www.scopus.com/inward/record.url?scp=85024821791&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024821791&partnerID=8YFLogxK

U2 - 10.1118/1.2241787

DO - 10.1118/1.2241787

M3 - Article

AN - SCOPUS:85024821791

VL - 33

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -