Active MR-temperature feedback control of dynamic interstitial ultrasound therapy in brain: in vivo experiments and modeling in native and coagulated tissues

W. A. N'Djin, M. Burtnyk, N. Lipsman, M. Bronskill, W. Kucharczyk, M. L. Schwartz, R. Chopra

Research output: Contribution to journalArticle

10 Citations (Scopus)

Abstract

PURPOSE: The recent clinical emergence of minimally invasive image-guided therapy has demonstrated promise in the management of brain metastasis, although control over the spatial pattern of heating currently remains limited. Based on experience in other organs, the delivery of high-intensity contact ultrasound energy from minimally invasive applicators can enable accurate spatial control of energy deposition, large treatment volumes, and high treatment rate. In this acute study, the feasibility of active MR-Temperature feedback control of dynamic ultrasound heat deposition for interstitial thermal ablation in brain was evaluatedin vivo.

METHODS: A four-element linear ultrasound transducer (f=8.2 MHz) originally developed for transurethral ultrasound therapy was used in a porcine model for generating thermal ablations in brain interstitially. First, the feasibility of treating and retreating preciselyin vivo brain tissues using stationary (non-rotating device) ultrasound exposures was studied in two pigs. Experimental results were compared to numerical simulations for maximum surface acoustic intensities ranging from 5 to 20 W cm(-2). Second, active MRT feedback-controlled ultrasound treatments were performed in three pigs with a rotating device to coagulate target volumes of various shapes. The acoustic power and rotation rate of the device were adjusted in real-time based on MR-thermometry feedback control to optimize heat deposition at the target boundary. Modeling of in vivo treatments were performed and compared to observed experimental results.

RESULTS: Overall, the time-space evolution of the temperature profiles observedin vivo could be well estimated from numerical simulations for both stationary and dynamic interstitial ultrasound exposures. Dynamic exposures performed under closed-loop temperature control enabled accurate elevation of the brain tissues within the targeted region above the 55 °C threshold necessary for the creation of irreversible thermal damage. Treatment volumes ranging from 1 to 9 cm3 were completed within 8±3 min with a radial targeting error<2 mm on average (treatment rate: 0.7±0.5 cm3/min). Tissue changes were visible on T1-weighted contrast-enhanced (T1w-CE) images immediately after treatment. These changes were also evident on T2-weighted (T2w) images acquired 2 h after the 1st treatment and correlated well with the MR-thermometry measurements.

CONCLUSIONS: These results support the feasibility of active MRT feedback control of dynamic interstitial ultrasound therapy ofin vivo brain tissues and confirm the feasibility of using simulations to predict spatial heating patterns in the brain.

Original languageEnglish (US)
Pages (from-to)93301
Number of pages1
JournalMedical Physics
Volume41
Issue number9
DOIs
StatePublished - Sep 1 2014

Fingerprint

Temperature
Hot Temperature
Brain
Thermometry
Swine
Acoustics
Equipment and Supplies
Heating
Therapeutics
Transducers
Neoplasm Metastasis

ASJC Scopus subject areas

  • Medicine(all)

Cite this

Active MR-temperature feedback control of dynamic interstitial ultrasound therapy in brain : in vivo experiments and modeling in native and coagulated tissues. / N'Djin, W. A.; Burtnyk, M.; Lipsman, N.; Bronskill, M.; Kucharczyk, W.; Schwartz, M. L.; Chopra, R.

In: Medical Physics, Vol. 41, No. 9, 01.09.2014, p. 93301.

Research output: Contribution to journalArticle

N'Djin, W. A. ; Burtnyk, M. ; Lipsman, N. ; Bronskill, M. ; Kucharczyk, W. ; Schwartz, M. L. ; Chopra, R. / Active MR-temperature feedback control of dynamic interstitial ultrasound therapy in brain : in vivo experiments and modeling in native and coagulated tissues. In: Medical Physics. 2014 ; Vol. 41, No. 9. pp. 93301.
@article{0e796c1fbfe649fb8179fe8dce0212fe,
title = "Active MR-temperature feedback control of dynamic interstitial ultrasound therapy in brain: in vivo experiments and modeling in native and coagulated tissues",
abstract = "PURPOSE: The recent clinical emergence of minimally invasive image-guided therapy has demonstrated promise in the management of brain metastasis, although control over the spatial pattern of heating currently remains limited. Based on experience in other organs, the delivery of high-intensity contact ultrasound energy from minimally invasive applicators can enable accurate spatial control of energy deposition, large treatment volumes, and high treatment rate. In this acute study, the feasibility of active MR-Temperature feedback control of dynamic ultrasound heat deposition for interstitial thermal ablation in brain was evaluatedin vivo.METHODS: A four-element linear ultrasound transducer (f=8.2 MHz) originally developed for transurethral ultrasound therapy was used in a porcine model for generating thermal ablations in brain interstitially. First, the feasibility of treating and retreating preciselyin vivo brain tissues using stationary (non-rotating device) ultrasound exposures was studied in two pigs. Experimental results were compared to numerical simulations for maximum surface acoustic intensities ranging from 5 to 20 W cm(-2). Second, active MRT feedback-controlled ultrasound treatments were performed in three pigs with a rotating device to coagulate target volumes of various shapes. The acoustic power and rotation rate of the device were adjusted in real-time based on MR-thermometry feedback control to optimize heat deposition at the target boundary. Modeling of in vivo treatments were performed and compared to observed experimental results.RESULTS: Overall, the time-space evolution of the temperature profiles observedin vivo could be well estimated from numerical simulations for both stationary and dynamic interstitial ultrasound exposures. Dynamic exposures performed under closed-loop temperature control enabled accurate elevation of the brain tissues within the targeted region above the 55 °C threshold necessary for the creation of irreversible thermal damage. Treatment volumes ranging from 1 to 9 cm3 were completed within 8±3 min with a radial targeting error<2 mm on average (treatment rate: 0.7±0.5 cm3/min). Tissue changes were visible on T1-weighted contrast-enhanced (T1w-CE) images immediately after treatment. These changes were also evident on T2-weighted (T2w) images acquired 2 h after the 1st treatment and correlated well with the MR-thermometry measurements.CONCLUSIONS: These results support the feasibility of active MRT feedback control of dynamic interstitial ultrasound therapy ofin vivo brain tissues and confirm the feasibility of using simulations to predict spatial heating patterns in the brain.",
author = "N'Djin, {W. A.} and M. Burtnyk and N. Lipsman and M. Bronskill and W. Kucharczyk and Schwartz, {M. L.} and R. Chopra",
year = "2014",
month = "9",
day = "1",
doi = "10.1118/1.4892923",
language = "English (US)",
volume = "41",
pages = "93301",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "9",

}

TY - JOUR

T1 - Active MR-temperature feedback control of dynamic interstitial ultrasound therapy in brain

T2 - in vivo experiments and modeling in native and coagulated tissues

AU - N'Djin, W. A.

AU - Burtnyk, M.

AU - Lipsman, N.

AU - Bronskill, M.

AU - Kucharczyk, W.

AU - Schwartz, M. L.

AU - Chopra, R.

PY - 2014/9/1

Y1 - 2014/9/1

N2 - PURPOSE: The recent clinical emergence of minimally invasive image-guided therapy has demonstrated promise in the management of brain metastasis, although control over the spatial pattern of heating currently remains limited. Based on experience in other organs, the delivery of high-intensity contact ultrasound energy from minimally invasive applicators can enable accurate spatial control of energy deposition, large treatment volumes, and high treatment rate. In this acute study, the feasibility of active MR-Temperature feedback control of dynamic ultrasound heat deposition for interstitial thermal ablation in brain was evaluatedin vivo.METHODS: A four-element linear ultrasound transducer (f=8.2 MHz) originally developed for transurethral ultrasound therapy was used in a porcine model for generating thermal ablations in brain interstitially. First, the feasibility of treating and retreating preciselyin vivo brain tissues using stationary (non-rotating device) ultrasound exposures was studied in two pigs. Experimental results were compared to numerical simulations for maximum surface acoustic intensities ranging from 5 to 20 W cm(-2). Second, active MRT feedback-controlled ultrasound treatments were performed in three pigs with a rotating device to coagulate target volumes of various shapes. The acoustic power and rotation rate of the device were adjusted in real-time based on MR-thermometry feedback control to optimize heat deposition at the target boundary. Modeling of in vivo treatments were performed and compared to observed experimental results.RESULTS: Overall, the time-space evolution of the temperature profiles observedin vivo could be well estimated from numerical simulations for both stationary and dynamic interstitial ultrasound exposures. Dynamic exposures performed under closed-loop temperature control enabled accurate elevation of the brain tissues within the targeted region above the 55 °C threshold necessary for the creation of irreversible thermal damage. Treatment volumes ranging from 1 to 9 cm3 were completed within 8±3 min with a radial targeting error<2 mm on average (treatment rate: 0.7±0.5 cm3/min). Tissue changes were visible on T1-weighted contrast-enhanced (T1w-CE) images immediately after treatment. These changes were also evident on T2-weighted (T2w) images acquired 2 h after the 1st treatment and correlated well with the MR-thermometry measurements.CONCLUSIONS: These results support the feasibility of active MRT feedback control of dynamic interstitial ultrasound therapy ofin vivo brain tissues and confirm the feasibility of using simulations to predict spatial heating patterns in the brain.

AB - PURPOSE: The recent clinical emergence of minimally invasive image-guided therapy has demonstrated promise in the management of brain metastasis, although control over the spatial pattern of heating currently remains limited. Based on experience in other organs, the delivery of high-intensity contact ultrasound energy from minimally invasive applicators can enable accurate spatial control of energy deposition, large treatment volumes, and high treatment rate. In this acute study, the feasibility of active MR-Temperature feedback control of dynamic ultrasound heat deposition for interstitial thermal ablation in brain was evaluatedin vivo.METHODS: A four-element linear ultrasound transducer (f=8.2 MHz) originally developed for transurethral ultrasound therapy was used in a porcine model for generating thermal ablations in brain interstitially. First, the feasibility of treating and retreating preciselyin vivo brain tissues using stationary (non-rotating device) ultrasound exposures was studied in two pigs. Experimental results were compared to numerical simulations for maximum surface acoustic intensities ranging from 5 to 20 W cm(-2). Second, active MRT feedback-controlled ultrasound treatments were performed in three pigs with a rotating device to coagulate target volumes of various shapes. The acoustic power and rotation rate of the device were adjusted in real-time based on MR-thermometry feedback control to optimize heat deposition at the target boundary. Modeling of in vivo treatments were performed and compared to observed experimental results.RESULTS: Overall, the time-space evolution of the temperature profiles observedin vivo could be well estimated from numerical simulations for both stationary and dynamic interstitial ultrasound exposures. Dynamic exposures performed under closed-loop temperature control enabled accurate elevation of the brain tissues within the targeted region above the 55 °C threshold necessary for the creation of irreversible thermal damage. Treatment volumes ranging from 1 to 9 cm3 were completed within 8±3 min with a radial targeting error<2 mm on average (treatment rate: 0.7±0.5 cm3/min). Tissue changes were visible on T1-weighted contrast-enhanced (T1w-CE) images immediately after treatment. These changes were also evident on T2-weighted (T2w) images acquired 2 h after the 1st treatment and correlated well with the MR-thermometry measurements.CONCLUSIONS: These results support the feasibility of active MRT feedback control of dynamic interstitial ultrasound therapy ofin vivo brain tissues and confirm the feasibility of using simulations to predict spatial heating patterns in the brain.

UR - http://www.scopus.com/inward/record.url?scp=84928189182&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84928189182&partnerID=8YFLogxK

U2 - 10.1118/1.4892923

DO - 10.1118/1.4892923

M3 - Article

C2 - 25186419

AN - SCOPUS:84928189182

VL - 41

SP - 93301

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 9

ER -