TY - JOUR
T1 - Cardiac perforation after device closure of atrial septal defects with the Amplatzer septal occluder
AU - Divekar, Abhay
AU - Gaamangwe, Tidimogo
AU - Shaikh, Nasir
AU - Raabe, Michael
AU - Ducas, John
PY - 2005/4/19
Y1 - 2005/4/19
N2 - Objectives: Amplatzer septal occluder (ASO)-associated cardiac perforation (CP) at our institution prompted this retrospective review. Background: Cardiac perforation is a rare complication after transcatheter atrial septal defect (ASD) closure. Methods: To identify CP after transcatheter ASD closure with ASO, cardiac events (CE) describing definite CP, hemopericardium, pericardial effusion, cardiovascular collapse, or sudden death were analyzed. Cardiac events were identified from published literature (MEDLINE), medical device regulating agencies in North America and the European Commission, and AGA Medical Corporation (Golden Valley, Minnesota). Institutional cases were reviewed. Cardiac events were defined as early (pre-discharge) or late (post-discharge). Results: Twenty-nine CEs were identified. Five were excluded because findings were inconclusive for device-related CP. Ten patients were <18 years of age. Late CEs occurred in 66.6%; 25% presented weeks later (longest, three years). Three deaths were reported. Cardiac perforation occurred predominantly in the anterosuperior atrial walls and/or adjacent aorta. Conclusions: Amplatzer septal occluder-associated CP uniquely involves the anterosuperior atrial walls and adjacent aorta. Pathophysiology remains poorly understood.
AB - Objectives: Amplatzer septal occluder (ASO)-associated cardiac perforation (CP) at our institution prompted this retrospective review. Background: Cardiac perforation is a rare complication after transcatheter atrial septal defect (ASD) closure. Methods: To identify CP after transcatheter ASD closure with ASO, cardiac events (CE) describing definite CP, hemopericardium, pericardial effusion, cardiovascular collapse, or sudden death were analyzed. Cardiac events were identified from published literature (MEDLINE), medical device regulating agencies in North America and the European Commission, and AGA Medical Corporation (Golden Valley, Minnesota). Institutional cases were reviewed. Cardiac events were defined as early (pre-discharge) or late (post-discharge). Results: Twenty-nine CEs were identified. Five were excluded because findings were inconclusive for device-related CP. Ten patients were <18 years of age. Late CEs occurred in 66.6%; 25% presented weeks later (longest, three years). Three deaths were reported. Cardiac perforation occurred predominantly in the anterosuperior atrial walls and/or adjacent aorta. Conclusions: Amplatzer septal occluder-associated CP uniquely involves the anterosuperior atrial walls and adjacent aorta. Pathophysiology remains poorly understood.
UR - http://www.scopus.com/inward/record.url?scp=17144396704&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=17144396704&partnerID=8YFLogxK
U2 - 10.1016/j.jacc.2004.12.072
DO - 10.1016/j.jacc.2004.12.072
M3 - Article
C2 - 15837251
AN - SCOPUS:17144396704
SN - 0735-1097
VL - 45
SP - 1213
EP - 1218
JO - Journal of the American College of Cardiology
JF - Journal of the American College of Cardiology
IS - 8
ER -