Enzymatic probe sonication as a tool for solid-liquid extraction for total selenium determination by electrothermal-atomic absorption spectrometry

G. Vale, S. Pereira, A. Mota, L. Fonseca, J. L. Capelo

Research output: Contribution to journalArticlepeer-review

27 Scopus citations

Abstract

A new fast and reproducible approach is described for the application of the enzymatic probe sonication (EPS) methodology [J.L. Capelo, P. Ximénez-Embún, Y. Madrid-Albarrán, C. Cámara, Anal. Chem. 76 (2004) 233-237] for total selenium determination by electrothermal atomic absorption spectrometry, ET-AAS. Ni(NO3)2 and Pd(NO3)2 were studied as matrix modifiers in conjunction with H2O2, being best results obtained with Pd(NO3)2 plus H2O2. The presence of H2O2 as matrix modifier increases up to 66% the time-life of the graphite tubes, by avoiding the building-up of carbonaceous residues. BCR-414 plankton and ERM-CE 278 mussel tissue reference materials were used for proof-of-the-methodology. Different enzymes, protease XIV, substilisin and trypsin were studied. The use of fresh enzyme was found critical. Good Se recoveries were obtained for oyster tissue, 111%; BCR-414 plankton, 106% and ERM-CE 278 mussel tissue, 93%, when protease XIV was used. Data regarding microwave digestion versus EPS methodology is also presented and discussed.

Original languageEnglish (US)
Pages (from-to)198-205
Number of pages8
JournalTalanta
Volume74
Issue number2
DOIs
StatePublished - Nov 30 2007

Keywords

  • EPS
  • ET-AAS
  • Enzymes
  • Focused ultrasound
  • Selenium
  • Solid-liquid extraction

ASJC Scopus subject areas

  • Analytical Chemistry

Fingerprint

Dive into the research topics of 'Enzymatic probe sonication as a tool for solid-liquid extraction for total selenium determination by electrothermal-atomic absorption spectrometry'. Together they form a unique fingerprint.

Cite this