Image-Guided Radiotherapy in Near Real Time With Intensity-Modulated Radiotherapy Megavoltage Treatment Beam Imaging

Weihua Mao, Annie Hsu, Nadeem Riaz, Louis Lee, Rodney Wiersma, Gary Luxton, Christopher King, Lei Xing, Timothy Solberg

Research output: Contribution to journalArticle

22 Citations (Scopus)

Abstract

Purpose: To utilize image-guided radiotherapy (IGRT) in near real time by obtaining and evaluating the online positions of implanted fiducials from continuous electronic portal imaging device (EPID) imaging of prostate intensity-modulated radiotherapy (IMRT) delivery. Methods and Materials: Upon initial setup using two orthogonal images, the three-dimensional (3D) positions of all implanted fiducial markers are obtained, and their expected two-dimensional (2D) locations in the beam's-eye-view (BEV) projection are calculated for each treatment field. During IMRT beam delivery, EPID images of the megavoltage treatment beam are acquired in cine mode and subsequently analyzed to locate 2D locations of fiducials in the BEV. Simultaneously, 3D positions are estimated according to the current EPID image, information from the setup portal images, and images acquired at other gantry angles (the completed treatment fields). The measured 2D and 3D positions of each fiducial are compared with their expected 2D and 3D setup positions, respectively. Any displacements larger than a predefined tolerance may cause the treatment system to suspend the beam delivery and direct the therapists to reposition the patient. Results: Phantom studies indicate that the accuracy of 2D BEV and 3D tracking are better than 1 mm and 1.4 mm, respectively. A total of 7330 images from prostate treatments were acquired and analyzed, showing a maximum 2D displacement of 6.7 mm and a maximum 3D displacement of 6.9 mm over 34 fractions. Conclusions: This EPID-based, real-time IGRT method can be implemented on any external beam machine with portal imaging capabilities without purchasing any additional equipment, and there is no extra dose delivered to the patient.

Original languageEnglish (US)
Pages (from-to)603-610
Number of pages8
JournalInternational Journal of Radiation Oncology Biology Physics
Volume75
Issue number2
DOIs
StatePublished - Oct 1 2009

Fingerprint

Image-Guided Radiotherapy
Intensity-Modulated Radiotherapy
radiation therapy
Equipment and Supplies
Fiducial Markers
Prostate
delivery
Therapeutics
electronics
Three-Dimensional Imaging
gantry cranes
markers
projection
dosage

Keywords

  • EPID
  • Fiducial marker
  • IGRT
  • Prostate
  • Real-time tracking

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging
  • Radiation
  • Cancer Research

Cite this

Image-Guided Radiotherapy in Near Real Time With Intensity-Modulated Radiotherapy Megavoltage Treatment Beam Imaging. / Mao, Weihua; Hsu, Annie; Riaz, Nadeem; Lee, Louis; Wiersma, Rodney; Luxton, Gary; King, Christopher; Xing, Lei; Solberg, Timothy.

In: International Journal of Radiation Oncology Biology Physics, Vol. 75, No. 2, 01.10.2009, p. 603-610.

Research output: Contribution to journalArticle

Mao, Weihua ; Hsu, Annie ; Riaz, Nadeem ; Lee, Louis ; Wiersma, Rodney ; Luxton, Gary ; King, Christopher ; Xing, Lei ; Solberg, Timothy. / Image-Guided Radiotherapy in Near Real Time With Intensity-Modulated Radiotherapy Megavoltage Treatment Beam Imaging. In: International Journal of Radiation Oncology Biology Physics. 2009 ; Vol. 75, No. 2. pp. 603-610.
@article{5f633f70b27f41f88b096cacc459c2c7,
title = "Image-Guided Radiotherapy in Near Real Time With Intensity-Modulated Radiotherapy Megavoltage Treatment Beam Imaging",
abstract = "Purpose: To utilize image-guided radiotherapy (IGRT) in near real time by obtaining and evaluating the online positions of implanted fiducials from continuous electronic portal imaging device (EPID) imaging of prostate intensity-modulated radiotherapy (IMRT) delivery. Methods and Materials: Upon initial setup using two orthogonal images, the three-dimensional (3D) positions of all implanted fiducial markers are obtained, and their expected two-dimensional (2D) locations in the beam's-eye-view (BEV) projection are calculated for each treatment field. During IMRT beam delivery, EPID images of the megavoltage treatment beam are acquired in cine mode and subsequently analyzed to locate 2D locations of fiducials in the BEV. Simultaneously, 3D positions are estimated according to the current EPID image, information from the setup portal images, and images acquired at other gantry angles (the completed treatment fields). The measured 2D and 3D positions of each fiducial are compared with their expected 2D and 3D setup positions, respectively. Any displacements larger than a predefined tolerance may cause the treatment system to suspend the beam delivery and direct the therapists to reposition the patient. Results: Phantom studies indicate that the accuracy of 2D BEV and 3D tracking are better than 1 mm and 1.4 mm, respectively. A total of 7330 images from prostate treatments were acquired and analyzed, showing a maximum 2D displacement of 6.7 mm and a maximum 3D displacement of 6.9 mm over 34 fractions. Conclusions: This EPID-based, real-time IGRT method can be implemented on any external beam machine with portal imaging capabilities without purchasing any additional equipment, and there is no extra dose delivered to the patient.",
keywords = "EPID, Fiducial marker, IGRT, Prostate, Real-time tracking",
author = "Weihua Mao and Annie Hsu and Nadeem Riaz and Louis Lee and Rodney Wiersma and Gary Luxton and Christopher King and Lei Xing and Timothy Solberg",
year = "2009",
month = "10",
day = "1",
doi = "10.1016/j.ijrobp.2009.04.068",
language = "English (US)",
volume = "75",
pages = "603--610",
journal = "International Journal of Radiation Oncology Biology Physics",
issn = "0360-3016",
publisher = "Elsevier Inc.",
number = "2",

}

TY - JOUR

T1 - Image-Guided Radiotherapy in Near Real Time With Intensity-Modulated Radiotherapy Megavoltage Treatment Beam Imaging

AU - Mao, Weihua

AU - Hsu, Annie

AU - Riaz, Nadeem

AU - Lee, Louis

AU - Wiersma, Rodney

AU - Luxton, Gary

AU - King, Christopher

AU - Xing, Lei

AU - Solberg, Timothy

PY - 2009/10/1

Y1 - 2009/10/1

N2 - Purpose: To utilize image-guided radiotherapy (IGRT) in near real time by obtaining and evaluating the online positions of implanted fiducials from continuous electronic portal imaging device (EPID) imaging of prostate intensity-modulated radiotherapy (IMRT) delivery. Methods and Materials: Upon initial setup using two orthogonal images, the three-dimensional (3D) positions of all implanted fiducial markers are obtained, and their expected two-dimensional (2D) locations in the beam's-eye-view (BEV) projection are calculated for each treatment field. During IMRT beam delivery, EPID images of the megavoltage treatment beam are acquired in cine mode and subsequently analyzed to locate 2D locations of fiducials in the BEV. Simultaneously, 3D positions are estimated according to the current EPID image, information from the setup portal images, and images acquired at other gantry angles (the completed treatment fields). The measured 2D and 3D positions of each fiducial are compared with their expected 2D and 3D setup positions, respectively. Any displacements larger than a predefined tolerance may cause the treatment system to suspend the beam delivery and direct the therapists to reposition the patient. Results: Phantom studies indicate that the accuracy of 2D BEV and 3D tracking are better than 1 mm and 1.4 mm, respectively. A total of 7330 images from prostate treatments were acquired and analyzed, showing a maximum 2D displacement of 6.7 mm and a maximum 3D displacement of 6.9 mm over 34 fractions. Conclusions: This EPID-based, real-time IGRT method can be implemented on any external beam machine with portal imaging capabilities without purchasing any additional equipment, and there is no extra dose delivered to the patient.

AB - Purpose: To utilize image-guided radiotherapy (IGRT) in near real time by obtaining and evaluating the online positions of implanted fiducials from continuous electronic portal imaging device (EPID) imaging of prostate intensity-modulated radiotherapy (IMRT) delivery. Methods and Materials: Upon initial setup using two orthogonal images, the three-dimensional (3D) positions of all implanted fiducial markers are obtained, and their expected two-dimensional (2D) locations in the beam's-eye-view (BEV) projection are calculated for each treatment field. During IMRT beam delivery, EPID images of the megavoltage treatment beam are acquired in cine mode and subsequently analyzed to locate 2D locations of fiducials in the BEV. Simultaneously, 3D positions are estimated according to the current EPID image, information from the setup portal images, and images acquired at other gantry angles (the completed treatment fields). The measured 2D and 3D positions of each fiducial are compared with their expected 2D and 3D setup positions, respectively. Any displacements larger than a predefined tolerance may cause the treatment system to suspend the beam delivery and direct the therapists to reposition the patient. Results: Phantom studies indicate that the accuracy of 2D BEV and 3D tracking are better than 1 mm and 1.4 mm, respectively. A total of 7330 images from prostate treatments were acquired and analyzed, showing a maximum 2D displacement of 6.7 mm and a maximum 3D displacement of 6.9 mm over 34 fractions. Conclusions: This EPID-based, real-time IGRT method can be implemented on any external beam machine with portal imaging capabilities without purchasing any additional equipment, and there is no extra dose delivered to the patient.

KW - EPID

KW - Fiducial marker

KW - IGRT

KW - Prostate

KW - Real-time tracking

UR - http://www.scopus.com/inward/record.url?scp=69549128103&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=69549128103&partnerID=8YFLogxK

U2 - 10.1016/j.ijrobp.2009.04.068

DO - 10.1016/j.ijrobp.2009.04.068

M3 - Article

C2 - 19735886

AN - SCOPUS:69549128103

VL - 75

SP - 603

EP - 610

JO - International Journal of Radiation Oncology Biology Physics

JF - International Journal of Radiation Oncology Biology Physics

SN - 0360-3016

IS - 2

ER -