Reproducibility of mandibular motion and muscle activity levels using a commercial computer recording system

Gaylord S. Throckmorton, Thomas J. Teenier, Edward Ellis

Research output: Contribution to journalArticle

29 Citations (Scopus)

Abstract

A commercial computer recording system (BioPak) was tested for its accuracy and reproducibility in recording mandibular motion and muscle activity levels (EMG). Accuracy of measurements of mandibular motion was tested using sliding calipers. Accuracy of muscle activity levels was checked with a sine-wave calibration signal of known amplitude and frequency. Reproducibility of motion and of EMG was made using five control subjects each tested on 3 separate days. Computer measurements of motion were found to require correction for nonlinearity at openings beyond 45 mm. Most maximum voluntary excursions were found to be more reproducible than motions during mastication. However, lateral deviation during opening was strongly affected by magnet orientation and varied greatly between trials. Electromyographic recordings were distorted by improper treatment of high-frequency components of the signal and also varied greatly between trials. This study suggests that the BioPak system may indicate changes in some measures of mandibular motion and EMG levels in patients for whom no changes have occurred.

Original languageEnglish (US)
Pages (from-to)348-354
Number of pages7
JournalThe Journal of Prosthetic Dentistry
Volume68
Issue number2
DOIs
StatePublished - 1992

Fingerprint

Computer Systems
Muscles
Magnets
Mastication
Calibration

ASJC Scopus subject areas

  • Oral Surgery
  • Dentistry(all)

Cite this

Reproducibility of mandibular motion and muscle activity levels using a commercial computer recording system. / Throckmorton, Gaylord S.; Teenier, Thomas J.; Ellis, Edward.

In: The Journal of Prosthetic Dentistry, Vol. 68, No. 2, 1992, p. 348-354.

Research output: Contribution to journalArticle

Throckmorton, Gaylord S. ; Teenier, Thomas J. ; Ellis, Edward. / Reproducibility of mandibular motion and muscle activity levels using a commercial computer recording system. In: The Journal of Prosthetic Dentistry. 1992 ; Vol. 68, No. 2. pp. 348-354.
@article{3248c530e07740dfaf57db18ee53a51f,
title = "Reproducibility of mandibular motion and muscle activity levels using a commercial computer recording system",
abstract = "A commercial computer recording system (BioPak) was tested for its accuracy and reproducibility in recording mandibular motion and muscle activity levels (EMG). Accuracy of measurements of mandibular motion was tested using sliding calipers. Accuracy of muscle activity levels was checked with a sine-wave calibration signal of known amplitude and frequency. Reproducibility of motion and of EMG was made using five control subjects each tested on 3 separate days. Computer measurements of motion were found to require correction for nonlinearity at openings beyond 45 mm. Most maximum voluntary excursions were found to be more reproducible than motions during mastication. However, lateral deviation during opening was strongly affected by magnet orientation and varied greatly between trials. Electromyographic recordings were distorted by improper treatment of high-frequency components of the signal and also varied greatly between trials. This study suggests that the BioPak system may indicate changes in some measures of mandibular motion and EMG levels in patients for whom no changes have occurred.",
author = "Throckmorton, {Gaylord S.} and Teenier, {Thomas J.} and Edward Ellis",
year = "1992",
doi = "10.1016/0022-3913(92)90344-A",
language = "English (US)",
volume = "68",
pages = "348--354",
journal = "Journal of Prosthetic Dentistry",
issn = "0022-3913",
publisher = "Mosby Inc.",
number = "2",

}

TY - JOUR

T1 - Reproducibility of mandibular motion and muscle activity levels using a commercial computer recording system

AU - Throckmorton, Gaylord S.

AU - Teenier, Thomas J.

AU - Ellis, Edward

PY - 1992

Y1 - 1992

N2 - A commercial computer recording system (BioPak) was tested for its accuracy and reproducibility in recording mandibular motion and muscle activity levels (EMG). Accuracy of measurements of mandibular motion was tested using sliding calipers. Accuracy of muscle activity levels was checked with a sine-wave calibration signal of known amplitude and frequency. Reproducibility of motion and of EMG was made using five control subjects each tested on 3 separate days. Computer measurements of motion were found to require correction for nonlinearity at openings beyond 45 mm. Most maximum voluntary excursions were found to be more reproducible than motions during mastication. However, lateral deviation during opening was strongly affected by magnet orientation and varied greatly between trials. Electromyographic recordings were distorted by improper treatment of high-frequency components of the signal and also varied greatly between trials. This study suggests that the BioPak system may indicate changes in some measures of mandibular motion and EMG levels in patients for whom no changes have occurred.

AB - A commercial computer recording system (BioPak) was tested for its accuracy and reproducibility in recording mandibular motion and muscle activity levels (EMG). Accuracy of measurements of mandibular motion was tested using sliding calipers. Accuracy of muscle activity levels was checked with a sine-wave calibration signal of known amplitude and frequency. Reproducibility of motion and of EMG was made using five control subjects each tested on 3 separate days. Computer measurements of motion were found to require correction for nonlinearity at openings beyond 45 mm. Most maximum voluntary excursions were found to be more reproducible than motions during mastication. However, lateral deviation during opening was strongly affected by magnet orientation and varied greatly between trials. Electromyographic recordings were distorted by improper treatment of high-frequency components of the signal and also varied greatly between trials. This study suggests that the BioPak system may indicate changes in some measures of mandibular motion and EMG levels in patients for whom no changes have occurred.

UR - http://www.scopus.com/inward/record.url?scp=0026901036&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0026901036&partnerID=8YFLogxK

U2 - 10.1016/0022-3913(92)90344-A

DO - 10.1016/0022-3913(92)90344-A

M3 - Article

C2 - 1501189

AN - SCOPUS:0026901036

VL - 68

SP - 348

EP - 354

JO - Journal of Prosthetic Dentistry

JF - Journal of Prosthetic Dentistry

SN - 0022-3913

IS - 2

ER -